Direct Drive® Clip Applier CA090Recalls & reported problems

Applied Medical Resources Corporation · Disposable Laparoscopic Clip Applier with 20 Clips · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Applied Medical Direct Drive clip applier has 806 adverse event reports, including 1 death and no injuries. Many describe clips failing to fire, jamming, or not forming properly, and most mention no effect on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
806
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 806 adverse event reports that list this exact device (DI 00607915117306), received between Jan 25, 2016 and Mar 21, 2017 — classified as 1 death, 805 malfunctions.

The most frequently reported specific device problem is “Misfire” (393 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Misfire393
  • Failure To Adhere Or Bond239
  • Mechanical Jam170
  • Failure to Form Staple124
  • Unintended Ejection91
  • Physical Resistance10
  • Expulsion6
  • Chemical Problem3

Patient problems

  • No Consequences Or Impact To Patient433
  • No Known Impact Or Consequence To Patient381
  • Extravasation12
  • Blood Loss3
  • Injury2
  • Brain Injury1
  • Death1
  • Fever1

Recent adverse event reports

The 25 most recent of 806 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 21, 2017Mechanical Jam; Mechanical Jam

    PROCEDURE PERFORMED UNKNOWN - "AFTER OPENING THE PACKAGING, DEVICE WAS FOUND UNUSABLE. JAW WAS BLOCKED." PATIENT STATUS - "NA."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6423158 ↗
  • MalfunctionMar 20, 2017Failure To Adhere Or Bond; Misfire; Failure To Adhere Or Bond; Misfire

    PATIENT PROCEDURE UNKNOWN- "USER REPORTED THAT ONE CLIP IS FIRED EVERY 3 TO 4 ACTUATION. DEVICE NEEDS TO BE ACTUATED 2 TO 3 TIMES IN ORDER TO CORRECTLY FIRE CLIP." PATIENT STATUS: OK.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 6419307 ↗
  • MalfunctionMar 10, 2017Failure To Adhere Or Bond; Misfire; Failure To Adhere Or Bond; Misfire

    LAP CHOLE- "4 EVENTS REPORTED WITH 4 DIFFERENT CLIP APPL, 2 DIFFERENT LOTS. NOT REPORTED SPECIFIC MALFUNCTION ASSOCIATED TO PROD. LOTS REPORTED ARE 1213391 AND 1213264 ( PLEASE REFER TO CER PORRETTA 3,4). THE SURG REPORTED NO FIRING OF CLIP AND NO COMPLETE CLOSURE/ WRONG CLOSURE OF THE CLIP AFTER FIRING. INTERVENTION WAS COMPLETED WITH A DIFFERENT BLAND CLIP APPLIER. NOT POSSIBLE TO RETIRE THE SAMPLES FOR INSPECTION BECAUSE THE EVENT WAS REPORTED TO ITALIAN MINISTRY OF HEALTH, SO PRODUCTS HAVE TO REMAIN AVAILABLE FOR MINISTRY EXAMINATION." PATIENT STATUS: GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6395597 ↗
  • MalfunctionMar 9, 2017Misfire; Misfire

    PROCEDURE PERFORMED UNKNOWN. "NO CLIPS WERE OBSERVED DURING USE OF THE DEVICE. A SECOND DEVICE NEEDED TO BE USED." PATIENT STATUS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6391470 ↗
  • MalfunctionMar 9, 2017Misfire; Misfire

    LAPAROSCOPIC CHOLECYSTECTOMY - "THE CLIP APPLIER ONLY FIRED THREE CLIPS AND NO CLIPS WOULD LOAD OR FIRE AFTER THE THIRD CLIP." PATIENT STATUS - "INCIDENT HAD NO AFFECT ON PATIENT. "

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6391443 ↗
  • MalfunctionMar 9, 2017Misfire; Misfire

    LAPAROSCOPIC CHOLECYSTECTOMY - "ONLY 4 CLIPS CAME OUT OF THE CLIP APPLIER, AFTER WHICH THE CLIP APPLIER SEEMS TO BE EMPTY OF CLIPS." INTERVENTION - "X." PATIENT STATUS - "OK."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6391441 ↗
  • MalfunctionMar 9, 2017Misfire; Misfire

    PROCEDURE PERFORMED UNKNOWN - "DURING USE OF THE DEVICE IT WAS OBSERVED THAT IT CONTAINED 2 CLIPS INSTEAD OF 20. A SECOND DEVICE WAS USED." PATIENT STATUS - "OK."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6391440 ↗
  • MalfunctionMar 9, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    LAP CHOLE - "THE STAFF OPENED A CLIP APPLIER AND ATTEMPTED TO USE IT IN THE PATIENT, IT ONLY HAD THREE CLIPS IN IT WHICH WERE ATTEMPTED TO BE USED IN THE PATIENT. UPON QUERY TO THE STAFF NURSE MICHAEL INFORMS THAT THEY NEEDED TO OPEN ANOTHER DEVICE IN ORDER TO FINISH THE PROCEDURE." PATIENT STATUS - UNKNOWN, PATIENT STATUS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6391436 ↗
  • MalfunctionMar 9, 2017Failure To Adhere Or Bond; Failure To Adhere Or Bond

    LAP CHOLE - "INST WERE CRACKED ON SIDE, IT SEEMED LIKE INST SPREAD APART AND DID NOT STAPLE CORRECTLY. I HAVE I INST HERE HAVE NOT SENT BACK YET." PATIENT STATUS - UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6391429 ↗
  • MalfunctionMar 8, 2017Misfire; Misfire

    PROCEDURE PERFORMED UNKNOWN. "STAPLES DOES NOT LOAD PROPERLY." PATIENT STATUS- FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6390204 ↗
  • MalfunctionMar 8, 2017Mechanical Jam; Mechanical Jam

    N/A- "THIS CLIP WAS NOT USED ON CASE. PRIOR TO CASE SURGICAL TECH TRIED TO FIRE THE CLIP APPLIER. IT JAMMED SEVERAL CLIPS IN THE JAW OF THE INSTRUMENT." PATIENT STATUS - UNKNOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6390195 ↗
  • MalfunctionMar 8, 2017Failure to Form Staple; Mechanical Jam; Failure to Form Staple; Mechanical Jam

    "HESITATION UPON FIRING, CLIP DID NOT FIRE." PATIENT STATUS: UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6390187 ↗
  • MalfunctionMar 8, 2017Misfire; Misfire

    LAP CHOLE - "ORIGINAL CUSTOMER DESCRIPTION: "TIP GOES IN WRONG DIRECTION" I REVIEWED THE CLIP APPLIER WHEN I PICKED IT UP ON TUESDAY JULY 8 AND A CLIP APPEARED AS THOUGHT IT FED SIDEWAYS, BUT AFTER FIRING IT ONCE TO SEE WHAT WAS MEANT BY "TIP GOES IN WRONG DIRECTION" THE CLIP FIRED NORMALLY. I WORKED TO GET ADDITIONAL INFORMATION FROM THE CUSTOMER ON THE DESCRIPTION AND WAS UNSUCCESSFUL." PATIENT STATUS - "OK."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6390182 ↗
  • MalfunctionMar 8, 2017Mechanical Jam; Mechanical Jam

    LAP ROUXEN-Y- "THE CLIP APPLIER HAD NO BLUE CLIPS AT THE END OF THE CLIP APPLIER AND IT NEVER LOCKED UP. THE AMOUNT OF CLIPS THAT WERE USED IS UNKNOWN." PATIENT STATUS - "NORMAL".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6390176 ↗
  • MalfunctionMar 7, 2017Misfire; Misfire

    LAP CHOLE- "FIRED FIRST AND SECOND CLIP THEY WERE FINE. THE THIRD CLIP FELL OUT OF THE CLIP APPLIER. THE 4TH 5TH AND 6TH CLIP ALSO FELL OUT OF THE CLIP APPLIER. THE NEXT CLIP APPLIER WAS FINE. CASE #2 AND 3, THE CLIPS WERE BOTH FINE. THE FOURTH CASE, THE THIRD CLIP FIRED FELL OUT OF THE CLIP APPLIER ALSO. ALL OF THE CLIP APPLIERS WERE DISCARDED AND UNRECOVERABLE." PATIENT STATUS: UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6385902 ↗
  • MalfunctionMar 7, 2017Misfire; Misfire

    LAP CHOLE- "FIRED FIRST AND SECOND CLIP THEY WERE FINE. THE THIRD CLIP FELL OUT OF THE CLIP APPLIER. THE FOURTH, FIFTH AND SIXTH CLIP ALSO FELL OUT OF THE CLIP APPLIER. THE NEXT CLIP APPLIER WAS FINE. CASE #2 AND 3, THE CLIPS WERE BOTH FINE. THE FOURTH CASE, THE THIRD CLIP FIRED FELL OUT OF THE CLIP APPLIER ALSO. ALL OF THE CLIP APPLIERS WERE DISCARDED AND UNRECOVERABLE." PATIENT STATUS: UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6385901 ↗
  • MalfunctionMar 2, 2017Misfire; Misfire

    LAPAROSCOPIC CHOLECYSTECTOMY - "LOADED APPLIER PRIOR TO USE, MISFIRED, SECOND ATTEMPT 2 STAPLE FIRED RELOADED MISFIRED AGAIN X5 REMOVED FROM FIELD, REPLACED WITH NEW ONE. DR. (B)(6) WAS PHYSICIAN. NURSE REPORTED, (B)(6)." PATIENT STATUS: POSSIBLE BLEEDING.

    Patient problems: No Code Available

    MAUDE report 6375126 ↗
  • MalfunctionMar 2, 2017Failure to Form Staple; Failure to Form Staple

    PROCEDURE PERFORMED UNKNOWN- "DURING USE OF THE DEVICE CLIPS GOT CROSSED- RISK OF BLEEDING." PATIENT STATUS: UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6375125 ↗
  • MalfunctionMar 2, 2017Failure To Adhere Or Bond; Failure To Adhere Or Bond

    LAPAROSCOPIC GASTRIC BANDING- "WOULD NOT CLOSE COMPLETELY ON 5 CLIPS." PATIENT STATUS - UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6374955 ↗
  • MalfunctionMar 2, 2017Unintended Ejection; Unintended Ejection

    LAP CHOLE - "THE DEVICE SPIT SEVERAL CLIPS WITHOUT PROPERLY FORMING ANY." PATIENT STATUS - "PATIENT DID WELL."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6373617 ↗
  • MalfunctionMar 2, 2017Misfire; Misfire

    "THIS IS A COMPLAINT FROM THE MARKET. (B)(4). THE DEFECT UNIT WITHOUT CLIPS WILL BE RETURNED TO AMR. PLEASE REFER TO THE COMPLAINT SHEET FOR INVESTIGATION." COMPLAINT SHEET- "THE THIRD CLIP WOULD NOT BE LOADED, CAUSING THE CLIP APPLIER TO BE UNUSABLE. THE PHYSICIAN, WHO IS SKILLFUL WITH THE DEVICE, COULD NOT PROCEED WITH THE PROCEDURE AND WOULD LIKE TO KNOW WHY THE INCIDENT HAPPENED. THE FIRST AND SECOND CLIP HAD NO TROUBLE WITH CLIPPING. THE EVEN UNIT WITHOUT CLIPS WAS RETURNED TO OLYMPUS AND VISUALLY INSPECTED : NO VISIBLE DAMAGE WAS FOUND WITH THE UNIT. THE UNIT WILL BE RETURNED TO AMR FOR FURTHER EVALUATION. (B)(4)." PATIENT STATUS: NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6373616 ↗
  • MalfunctionMar 2, 2017Mechanical Jam; Mechanical Jam

    LAP CHOLE- "ONLY 3 CLIPS WERE AVAILABLE TO USE BEFORE IT LOCKED UP." PATIENT STATUS- "OK."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6373614 ↗
  • MalfunctionMar 2, 2017Failure to Form Staple; Failure to Form Staple

    LAPAROSCOPIC: "SURGEON WENT TO USE CLIP APPLIER BUT SAYS IT FAILED, NO OTHER DESCRIPTION GIVEN AT THE TIME." PATIENT STATUS: OK.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6373613 ↗
  • MalfunctionMar 2, 2017Failure to Form Staple; Failure to Form Staple

    "SURGEON WENT TO USE CLIP APPLIER BUT SAYS IT FAILED, NO OTHER DESCRIPTION GIVEN AT THE TIME." PATIENT STATUS - UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6373612 ↗
  • MalfunctionMar 2, 2017Unintended Ejection; Misfire; Unintended Ejection; Misfire

    LAPAROSCOPIC CHOLE CYSTECTOMY- "WHEN PHYSICIAN TRIGGER TO APPLY CLIP, MULTIPLE CLIPS CAME OUT. PHYSICIAN STATES APPLIER "EXPLODED" WHEN MULTIPLE CLIPS CAME OUT AT ONCE." PATIENT STATUS- "PATIENT IS OK, NO HARM DONE".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6373610 ↗

FDA clearance and classification

The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Direct Drive® Clip Applier CA090

Has the Direct Drive® Clip Applier CA090 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00607915117306), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Direct Drive® Clip Applier CA090?

The FDA's MAUDE database lists 806 adverse event reports for this device, received between Jan 25, 2016 and Mar 21, 2017. In those reports, the most common specific device problems are “Misfire” (393), “Failure To Adhere Or Bond” (239) and “Mechanical Jam” (170); the most common patient problems are “Extravasation” (12), “Blood Loss” (3) and “Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Direct Drive® Clip Applier CA090?

1 report is classified as a death, 0 as injuries and 805 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Direct Drive® Clip Applier CA090 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Direct Drive® Clip Applier CA090, and is it still sold?

The FDA's device database lists Applied Medical Resources Corporation as the company responsible for it (brand Direct Drive® Clip Applier, model CA090). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Clip, Implantable devices

All Clip, Implantable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page