Endo Clip 176625Recalls & reported problems

Covidien LP · Clip Applier · Class 2 · Implant

Quick summary

This Covidien Endo Clip surgical clip applier has 766 reports, including 1 death and 52 injuries. Many reports describe clips that fail to fire, fire late, or stop working partway through a procedure, and some mention a knob falling off. An injury report describes a clip that did not deploy during surgery, which led to a cut artery and heavy bleeding. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
768
Deaths reported
1
Injuries reported
52
Recalls naming it
0

Summary

The FDA's MAUDE database contains 768 adverse event reports that list this exact device (DI 10884521057845), received between Apr 12, 2016 and Sep 22, 2026 — classified as 1 death, 52 injuries, 715 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (402 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 15 reportsDec 2023: 9 reportsJan 2024: 7 reportsFeb 2024: 6 reportsMar 2024: 6 reportsApr 2024: 7 reportsMay 2024: 7 reportsJun 2024: 5 reportsJul 2024: 7 reportsAug 2024: 4 reportsSep 2024: 13 reportsOct 2024: 6 reportsNov 2024: 12 reportsDec 2024: 3 reportsJan 2025: 11 reportsFeb 2025: 6 reportsMar 2025: 6 reportsApr 2025: 3 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 5 reportsAug 2025: 5 reportsSep 2025: 7 reportsOct 2025: 3 reportsNov 2025: 6 reportsDec 2025: 2 reportsJan 2026: 7 reportsFeb 2026: 12 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 5 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanics Altered402
  • Mechanical Problem222
  • Material Deformation101
  • Detachment of Device or Device Component63
  • Difficult to Open or Close53
  • Device Slipped35
  • Break32
  • Adverse Event Without Identified Device or Use Problem27

Patient problems

  • No Clinical Signs, Symptoms or Conditions401
  • No Known Impact Or Consequence To Patient137
  • No Consequences Or Impact To Patient130
  • Insufficient Information51
  • No Code Available22
  • No Patient Involvement18
  • Tissue Breakdown18
  • Unspecified Tissue Injury13

Recent adverse event reports

The 25 most recent of 768 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Mechanics Altered

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860261 ↗
  • MalfunctionSep 21, 2026Mechanical Problem

    It was reported that during a laparoscopic procedure, while being used to apply clips to the cystic duct, the clips became deformed during attempted clipping despite no excessive force reportedly being applied by the surgeon, and the issue recurred during severalattempts. The surgeon was able to squeeze the handle. A different product was used to complete the procedure. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841071 ↗
  • MalfunctionAug 28, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498843 ↗
  • MalfunctionAug 27, 2026Detachment of Device or Device Component

    According to the reporter, the knob fell off. There was no patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50486708 ↗
  • MalfunctionAug 27, 2026Mechanics Altered

    It was reported that the device experienced delayed deployment.

    Patient problems: Insufficient Information

    MAUDE report 50470230 ↗
  • MalfunctionJul 28, 2026Mechanics Altered

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50061270 ↗
  • MalfunctionJul 20, 2026Mechanics Altered

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25874972 ↗
  • InjuryJul 16, 2026Mechanics Altered; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING CLIPPING OF THE CYSTIC ARTERY, THE HANDLE OF THE CLIP APPLIER WAS NOT SQUEEZED, AND THE INITIAL FIRING FAILED TO DEPLOY A CLIP. UPON REPEAT ACTIVATION, THE FIRING MECHANISM ADVANCED WITHOUT DEPLOYING A CLIP, RESULTING IN TRANSECTION OF THE CYSTIC ARTERY AND SIGNIFICANT BLEEDING. ATTEMPTS TO FIRE THE CLIP APPLIER OUTSIDE THE PATIENT ALSO FAILED TO DEPLOY ANY CLIPS. THE MALFUNCTION CAUSED SIGNIFICANT BLOOD LOSS AND MASSIVE HEMORRHAGE, PROMPTED A CODE PROTOCOL, AND RESULTED IN THE PATIENT BEING ADMITTED TO THE ICU. IT WAS CONFIRMED THAT THE PATIENT IS DOING WELL WITH NO FURTHER COMPLICATIONS IN HOSPITAL AND HAS BEEN DISCHARGED HOME.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25850113 ↗
  • MalfunctionJul 15, 2026Mechanics Altered

    ACCORDING TO THE REPORTER, THE DEFECTIVE CLIP APPLIER MISFIRED.

    Patient problems: Insufficient Information

    MAUDE report 25827898 ↗
  • MalfunctionJul 6, 2026Mechanics Altered

    ACCORDING TO THE REPORTER DURING A PROCEDURE, AFTER REPEATED FIRING OF THE CLIPS, THE DEVICE COULD NO LONGER DEPLOY THE CLIPS IN THE LATER STAGE OF USE. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 25727446 ↗
  • MalfunctionJun 15, 2026Mechanical Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTROTOMY PROCEDURE, THE DEVICE HAD AN UNKNOWN ISSUE. A NEW PRODUCT WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE JAWS WOULD NOT CLOSE AND THE LOADING TRIGGER COULD NOT BE ACTUATED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25513691 ↗
  • MalfunctionMay 29, 2026Mechanical Problem

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A GALLBLADDER PROCEDURE, THE CLIPS WERE CLOSING AT A DIAGONAL ANGLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25328040 ↗
  • MalfunctionMay 29, 2026Mechanical Problem

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A GALLBLADDER PROCEDURE, THE CLIPS WERE CLOSING AT A DIAGONAL ANGLE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25327955 ↗
  • MalfunctionApr 14, 2026Mechanics Altered

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, ON THE STAGE OF EXTRACTING THE GALLBLADDER, THE SURGEON WAS ABLE TO SQUEEZE THE HANDLE AND USAGE WAS DISCONTINUED BECAUSE THE CLIP WAS MISALIGNED DURING FIRING. THE MISALIGNED CLIP WAS COLLECTED AND THERE WAS NO PROBLEMS WITH THE NEW CLIP. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872238 ↗
  • MalfunctionApr 3, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE DID NOT CLOSE PROPERLY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781749 ↗
  • MalfunctionMar 26, 2026Mechanical Jam

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE CLIP APPLIER WAS INITIALLY WORKING, AND THEN IT STARTED TO JAM AND NOT BE USABLE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24708230 ↗
  • MalfunctionMar 19, 2026Mechanical Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SURGEON WAS ABLE TO SQUEEZE THE HANDLE AND THE JAWS WERE ABLE TO CLOSE, BUT THE CLIPS DID NOT FORM CORRECTLY. ANOTHER CLIP APPLIER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24645772 ↗
  • MalfunctionFeb 27, 2026Detachment of Device or Device Component; Mechanics Altered

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A LAPAROSCOPIC PROCEDURE, THE SURGEON WAS ABLE TO SQUEEZE THE HANDLE, BUT THE DEVICE MISFIRED. THE RING FROM THE CONNECTION OF THE HANDLE¿S BODY AND THE SHAFT ALSO DETACHED AND SHOT ACROSS THE ROOM. NO DEVICE COMPONENT FELL INTO THE PATIENT¿S CAVITY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24470615 ↗
  • MalfunctionFeb 25, 2026Break; Mechanics Altered

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, AFTER SQUEEZING THE HANDLE, THE CLIPS WERE NOT LOADED WELL, AND THE LAUNCH PART WAS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 24444350 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THERE WERE NO INDICATIONS THAT THE CLIPS WERE MISPLACED OR MALFORMED. IT WAS STANDARDIZED TO DO A LEAKAGE CONTROL AND INSPECT THE CLIPS' PLACEMENT. THERE WAS NOTHING ABNORMAL DURING THE INITIAL SURGERY. DUE TO PROLONGED REHABILITATION POSTOPERATIVELY, THE PATIENT UNDERWENT EXTRA CONTROLS AND WAS DIAGNOSED WITH BILE LEAKAGE. THE EXTRA INTERVENTIONS CONSISTED OF PLACEMENT OF DRAINAGE, INTRAVENOUS ANTIBIOTICS, AND ENDOSCOPIC RETROGRADE CHOLANGIOGRAPHY (ERC). A PROLONGED HOSPITAL STAY WAS NEEDED.

    Patient problems: Biliary Leak; Tissue Breakdown

    MAUDE report 24398434 ↗
  • InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THERE WERE NO INDICATIONS THAT THE CLIPS WERE MISPLACED OR MALFORMED. IT WAS STANDARDIZED TO DO A LEAKAGE CONTROL AND INSPECT THE CLIPS' PLACEMENT. THERE WAS NOTHING ABNORMAL DURING THE INITIAL SURGERY. DUE TO PROLONGED REHABILITATION POSTOPERATIVELY, THE PATIENT UNDERWENT EXTRA CONTROLS AND WAS DIAGNOSED WITH BILE LEAKAGE. THE EXTRA INTERVENTIONS CONSISTED OF PLACEMENT OF DRAINAGE, INTRAVENOUS ANTIBIOTICS, AND ENDOSCOPIC RETROGRADE CHOLANGIOGRAPHY (ERC). A PROLONGED HOSPITAL STAY WAS NEEDED.

    Patient problems: Biliary Leak; Tissue Breakdown

    MAUDE report 24394123 ↗
  • MalfunctionFeb 19, 2026Material Deformation

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIPS WERE SCISSORED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385878 ↗
  • MalfunctionFeb 19, 2026Material Deformation

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A BARIATRIC PROCEDURE, WHILE IN THE STOMACH, THE CLIPS WERE SCISSORED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385871 ↗
  • MalfunctionFeb 19, 2026Material Deformation

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIPS WERE SCISSORED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385862 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THERE WERE NO INDICATIONS THAT THE CLIPS WERE MISPLACED OR MALFORMED. IT WAS STANDARDIZED TO DO A LEAKAGE CONTROL AND INSPECT THE CLIPS' PLACEMENT. THERE WAS NOTHING ABNORMAL DURING THE INITIAL SURGERY. DUE TO PROLONGED REHABILITATION POSTOPERATIVELY, THE PATIENT UNDERWENT EXTRA CONTROLS AND WAS DIAGNOSED WITH BILE LEAKAGE. THE EXTRA INTERVENTIONS CONSISTED OF PLACEMENT OF DRAINAGE, INTRAVENOUS ANTIBIOTICS, AND ENDOSCOPIC RETROGRADE CHOLANGIOGRAPHY (ERC). A PROLONGED HOSPITAL STAY WAS NEEDED.

    Patient problems: Biliary Leak; Tissue Breakdown

    MAUDE report 24376343 ↗

Other CLIP, IMPLANTABLE devices