Epix® Universal Clip Applier CA500Recalls & reported problems
Applied Medical Resources Corporation · Disposable Laparoscopic Clip Applier with 20 Clips. · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Applied Medical Epix Universal Clip Applier is a disposable laparoscopic device that places surgical clips during surgery. Many of its 1,185 reports describe clips not firing, jamming or misaligning, and some mention bleeding and lacerations. The reports include 1 death and 23 injuries. One recall is open, from January 2024, over the applier possibly not loading a clip after the trigger is pressed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,176
- Deaths reported
- 1
- Injuries reported
- 24
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,176 adverse event reports that list this exact device (DI 00607915125318), received between Jan 25, 2016 and Sep 30, 2026 — classified as 1 death, 24 injuries, 1,151 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (619 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jan 30, 2024Class IIOpen, Classified
Their is the potential that Universal Clip applier may not load clip after the trigger is actuated.
Z-1388-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Fire619
- Misfire129
- Mechanical Jam120
- Failure To Adhere Or Bond86
- Failure to Form Staple82
- Loss of or Failure to Bond70
- Failure to Align66
- Fluid/Blood Leak20
Patient problems
- No Clinical Signs, Symptoms or Conditions694
- No Consequences Or Impact To Patient297
- No Known Impact Or Consequence To Patient132
- Hemorrhage/Blood Loss/Bleeding12
- Laceration(s)11
- Perforation of Vessels8
- Injury6
- Insufficient Information5
Recent adverse event reports
The 25 most recent of 1,176 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Fluid/Blood Leak
Name of Procedure: Lap Chole.\\Event Description:\\\"[Hospital]\" called us today to report 2 incidents both with CA500. \\Case 9/17. Lap Chole. Dr. \"[NAME]\". Patient admitted to hospital after procedure over the weekend due to clip failure and leaking. \\Additional information provided by an Applied Medical representative via email on 23SEP2026:\\Patient went to Emergency Room (ER) complaining of pain, was admitted and brought into surgery to repair leaking duct. Dr. \"[NAME]\" did not perform the follow up case. As far as we are aware, patient is okay. The timing of the event occurred post-procedure- Dr. \"[NAME]\" applied 3 clips Patient-side for security and patient experienced post-op complications with a leaking duct. Dr. \"[NAME]\" does not know if clips fell off.\\The surgeon was ligating the ducts in a lap. chole procedure. The duct tissue was leaking. It is unknown/not reported by surgeon if the distal end of the jaws visible around the tissue once the clip was placed around the tissue. It is unknown/not reported by the surgeon if any tissue was trapped in the jaws. It is unknown/not reported by the surgeon if the bleeding/leaking occurred due to the clip not occluding
Patient problems: Biliary Leak
MAUDE report 50947549 ↗MalfunctionSep 10, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50724273 ↗MalfunctionSep 3, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50594307 ↗MalfunctionAug 31, 2026Failure to Align; Failure to Form Staple
Clip applier misaligning clips after 2-3 firings.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50525861 ↗MalfunctionAug 31, 2026Mechanical Jam
clip applier jammed during application.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50525649 ↗MalfunctionAug 28, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50508284 ↗MalfunctionJun 9, 2026Failure to Fire
PROCEDURE PERFORMED: NI EVENT DESCRIPTION: THE CUSTOMER CALLED AND WANTED TO FILE A COMPLAINT REGARDING ITEM CA500 LOT# 1557343 THAT MISSED FIRED ON 5/8/25. CONTACT [NAME] [EMAIL] [PHONE NUMBER] ADDITIONAL INFORMATION PROVIDED BY [CONTACT NAME] ON 09JUN2026: IT WAS 05/08/26. I HAVE SPOKEN TO THE STAFF AND I/VE GOTTEN THE RUN AROUND ABOUT WHO WAS IN THE ROOM. I¿VE SENT THEM YOUR EMAIL ASKING THEM TO COMPLETE THE QUESTIONS AND NO RESPONSE. SO I THINK IT¿S FORGOTTEN ABOUT. INTERVENTION: NI PATIENT STATUS: NO PATIENT INJURY INDICATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25446298 ↗MalfunctionJun 2, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: [HOSPITAL] "DURING USE, THERE WAS A PROBLEM WITH SHOOTING." PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON (B)(6) 2026 FROM [NAME], TERRITORY MANAGER: "THE ¿LOADING THE CLIP¿ ACTION WAS BEING PERFORMED WHEN THE EXPERIENCE OCCURRED. THE INSTRUMENT ¿MISFIRED¿ IN THAT IT WAS STRUCK WHEN THE TRIGGER WAS PULLED. A TOTAL OF 3¿4 CLIPS WERE DISPENSED. THE INCIDENT OCCURRED WITH 1¿2 CLIPS. WHEN IT STRUCK, THE SURGEON DID NOT TRY TO FIX IT ANYMORE BECAUSE IT COULD MISFIRE WHEN THE CLIP APPLIER IS INSIDE PERITONEUM." INTERVENTION: THE CASE WAS COMPLETED WITH THE NEW DEVICE. PATIENT STATUS: THERE WAS NO PROBLEM WITH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25367695 ↗MalfunctionMay 7, 2026Failure to Fire
PROCEDURE PERFORMED: UNKNOWN. EVENT DESCRIPTION: AT THE BEGINNING OF THE SURGERY, EVERY OTHER CLIP FAILED TO LOAD (DID NOT LOAD), AFTER USING 5-10 CLIPS THE DEVICE WORKED NORMALLY. NO CLINICAL IMPACT TO THE PATIENT AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION RECEIVED BY APPLIED MEDICAL REP VIA EMAIL ON (B)(6) 2026: THIS HAS HAPPENED ALREADY ON (B)(6). INTERVENTION: UNKNOWN. PATIENT STATUS: PATIENT IS OK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25116159 ↗MalfunctionApr 30, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: [HOSPITAL]. (THIS EVENT WAS DIRECTLY REPORTED TO MFDS BY [HOSPITAL], BUT THE EVENT OCCURRED AT [HOSPITAL]. SINCE TWO HOSPITALS ARE PARTNER HOSPITAL (BOTH AFFILIATED WITH [UNIVERSITY]), IT WAS REPORTED BY [HOSPITAL] THAT HAS ADVERSE EVENT REPORTING CENTER.) "THE CLIP WAS NOT LOADED." INTERVENTION: NI. PATIENT STATUS: NO PATIENT INJURY INDICATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25050193 ↗MalfunctionApr 22, 2026Failure to Fire
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY EVENT DESCRIPTION: THE EVENT DATE IS UNKNOWN. THE CUSTOMER REPORTED THAT ABOUT FIVE OR SIX DIFFERENT CLIP APPLIERS (LIKELY FROM THE SAME LOT) DID NOT DEPLOY THE FIRST (OR FIRST AND SECOND CLIP) CORRECTLY. THERE WAS EITHER NO CLIP OR A DAMAGED CLIP. AFTER SOME "PLASTIC TO PLASTIC" USAGES THE FOLLOWING CLIPS WORKED JUST FINE. THE PHYSICIANS ARE TRAINED TO USE THE CLIP APPLIER CORRECTLY AND THEY ARE NOT PRELOADING THE CLIPAPPLIER BEFORE ENTERING THE TROCAR. I TALKED TO THE CHIEF PHYSICIAN ABOUT THOSE INCIDENTS AND HE ASSURED ME THAT HE HAS BEEN USING THE CLIP APPLIERS THE SAME WAY HE ALWAYS HAD. NO CLINICAL IMPACT TO THE PATIENT AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 13APR26: THE CUSTOMER PURCHASED IT AS A KIT. EVENT DATES ARE DIFFERENT, BASICALLY IN THE 3 WEEKS PRIOR TO THE EVENT DATE. I CAN NOT CONFIRM IF ALL THE INCIDENT DEVICES BELONGED TO LOT 1559037. BUT THE CUSTOMER TOLD ME THAT THEY NEVER HAD ISSUES BEFORE OR AFTER. THEY THINK THEY ALL BELONGED TO THE SAME LOT. I CAN NOT CLARIFY IT THOUGH. 4 OF THE CLIP APPLIERS DID NOT APPLY ANY CLIPS FOR THE FIRST FEW "PLASTI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24960653 ↗MalfunctionApr 22, 2026Failure to Fire
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: THE EVENT DATE IS UNKNOWN. THE CUSTOMER REPORTED THAT ABOUT FIVE OR SIX DIFFERENT CLIP APPLIERS (LIKELY FROM THE SAME LOT) DID NOT DEPLOY THE FIRST (OR FIRST AND SECOND CLIP) CORRECTLY. THERE WAS EITHER NO CLIP OR A DAMAGED CLIP. AFTER SOME "PLASTIC TO PLASTIC" USAGES THE FOLLOWING CLIPS WORKED JUST FINE. THE PHYSICIANS ARE TRAINED TO USE THE CLIPAPPLIER CORRECTLY AND THEY ARE NOT PRELOADING THE CLIPAPPLIER BEFORE ENTERING THE TROCAR. I TALKED TO THE CHIEF PHYSICIAN ABOUT THOSE INCIDENTS AND HE ASSURED ME THAT HE HAS BEEN USING THE CLIP APPLIERS THE SAME WAY HE ALWAYS HAD. NO CLINICAL IMPACT TO THE PATIENT AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 13-APR-2026: THE CUSTOMER PURCHASED IT AS A KIT. EVENT DATES ARE DIFFERENT, BASICALLY IN THE 3 WEEKS PRIOR TO THE EVENT DATE. I CAN NOT CONFIRM IF ALL THE INCIDENT DEVICES BELONGED TO LOT 1559037. BUT THE CUSTOMER TOLD ME THAT THEY NEVER HAD ISSUES BEFORE OR AFTER. THEY THINK THEY ALL BELONGED TO THE SAME LOT. I CAN NOT CLARIFY IT THOUGH. 4 OF THE CLIP APPLIERS DID NOT APPLY ANY CLIPS FOR THE FIRST FEW "PL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24960623 ↗MalfunctionApr 22, 2026Failure to Fire
PROCEDURE PERFORMED: CHOLECYSTECTOMY. EVENT DESCRIPTION: THE EVENT DATE IS UNKNOWN. ADDITIONAL INFORMATION RECEIVED BY AN APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 5-JAN-2026: NO PATIENT INJURY OCCURRED AS A RESULT OF THIS EVENT. PATIENT STATUS OK. RESOLVE THE SITUATION: USE ANOTHER ONE. ADDITIONAL INFORMATION RECEIVED BY AN APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 17-MAR-2026: TM CONFIRMS THE LOT AS 1542695. INTERVENTION: USE ANOTHER ONE. PATIENT STATUS: OK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24958506 ↗MalfunctionApr 16, 2026Fluid/Blood Leak
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY EVENT DESCRIPTION: WHEN PLACING THE CLIPS TO CLAMP THE CYSTIC DUCT: THE CLIPS REMAINED OPEN AND FAILED. CONSEQUENTLY, THE CYSTIC DUCT OPENED, BILE LEAKED INTO THE ABDOMINAL CAVITY, THE DURATION OF THE PROCEDURE WAS PROLONGED, AND THERE WAS AN INCREASED RISK OF A POST-OPERATIVE ABSCESS. THE STAPLER WAS CHANGED (TO A NON-AUTOMATIC STAPLER AND CLIP CARTRIDGES) TO CONTINUE THE PROCEDURE. INTERVENTION: THE STAPLER WAS CHANGED (TO A NON-AUTOMATIC STAPLER AND CLIP CARTRIDGES) TO CONTINUE THE PROCEDURE. PATIENT STATUS: PATIENT IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24901687 ↗MalfunctionApr 1, 2026Failure to Fire
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY EVENT DESCRIPTION: THE DEVICE WAS BEING USED TO CLIP A VESSEL AND IT WOULD NOT RELOAD THE CLIP. THERE WAS NO CLINICAL IMPACT TO THE PATIENT AS A RESULT OF THE EVENT. NO HARM DONE TO THE PATIENT. THE USER RESOLVED THE SITUATION BY OPENING ANOTHER CLIP APPLIER. THE OTHER INSTRUMENTS USED AT THE TIME WAS STANDARD KIT FOR LAP CHOLE. ADDITIONAL INFORMATION WAS PROVIDED BY [TERRITORY MANAGER] VIA EMAIL ON 18-MAR-2026: "THE HOSPITAL DOES NOT USE APPLIED MEDICAL LAP CHOLE KITS, THEY MAKE UP THEIR OWN." INTERVENTION: THEY RESOLVED THE SITUATION BY OPENING ANOTHER APPLIER. PATIENT STATUS: NO HARM DONE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757186 ↗MalfunctionMar 30, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: WHEN THEY TRIED THE CLIP APPLIER, THERE WAS INSIDE ONLY ONE CLIP. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA TEAMS CALL ON 10MAR26: ALL THE EVENTS TOOK PLACE ON MARCH 8TH BEFORE USE. (B)(4) DEVICES FROM LOT 1560548 HAD THE SAME PROBLEM (ONLY ONE CLIP WAS INSIDE THE DEVICE) AND THEY WERE DISCARDED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 17MAR26: I¿VE JUST CHECKED AND THE CORRECT LOT NUMBER IS 1567140. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 19MAR26: THE USER RESOLVED THE SITUATION CHANGING DEVICE. ONLY ONE CLIP LOADED INTO THE JAWS.12 MM TROCAR WAS USED. THE CLIP WAS LOADED AND VISIBLE BETWEEN THE JAWS OF THE DEVICE, AND IT WAS PROPERLY SEATED. NO (REGARDING CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR) BECAUSE THEY TRIED IT BEFORE PUTTING INSIDE THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP WAS MANUALLY REMOVED FROM THE JAWS. THEY TRIED AGAIN, BUT IT WAS EMPTY. NO OTHER CLIP INSIDE. LIKE ALWAY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24721642 ↗MalfunctionMar 30, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: WHEN THEY TRIED THE CLIP APPLIER, THERE WAS INSIDE ONLY ONE CLIP. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA TEAMS CALL ON 10MAR26: ALL THE EVENTS TOOK PLACE ON MARCH 8TH BEFORE USE. 4 DEVICES FROM LOT 1560548 HAD THE SAME PROBLEM (ONLY ONE CLIP WAS INSIDE THE DEVICE) AND THEY WERE DISCARDED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 17MAR26: I¿VE JUST CHECKED AND THE CORRECT LOT NUMBER IS 1567140. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 19MAR26: THE USER RESOLVED THE SITUATION CHANGING DEVICE. ONLY ONE CLIP LOADED INTO THE JAWS.12 MM TROCAR WAS USED. THE CLIP WAS LOADED AND VISIBLE BETWEEN THE JAWS OF THE DEVICE, AND IT WAS PROPERLY SEATED. NO (REGARDING CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR) BECAUSE THEY TRIED IT BEFORE PUTTING INSIDE THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP WAS MANUALLY REMOVED FROM THE JAWS. THEY TRIED AGAIN, BUT IT WAS EMPTY. NO OTHER CLIP INSIDE. LIKE ALWAYS, TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24721313 ↗MalfunctionMar 30, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: WHEN THEY TRIED THE CLIP APPLIER, THERE WAS INSIDE ONLY ONE CLIP. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA TEAMS CALL ON 10MAR26: ALL THE EVENTS TOOK PLACE ON MARCH 8TH BEFORE USE. 4 DEVICES FROM LOT 1560548 HAD THE SAME PROBLEM (ONLY ONE CLIP WAS INSIDE THE DEVICE) AND THEY WERE DISCARDED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 17MAR26: I¿VE JUST CHECKED AND THE CORRECT LOT NUMBER IS 1567140. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 19MAR26: THE USER RESOLVED THE SITUATION CHANGING DEVICE. ONLY ONE CLIP LOADED INTO THE JAWS.12 MM TROCAR WAS USED. THE CLIP WAS LOADED AND VISIBLE BETWEEN THE JAWS OF THE DEVICE, AND IT WAS PROPERLY SEATED. NO (REGARDING CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR) BECAUSE THEY TRIED IT BEFORE PUTTING INSIDE THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP WAS MANUALLY REMOVED FROM THE JAWS. THEY TRIED AGAIN, BUT IT WAS EMPTY. NO OTHER CLIP INSIDE. LIKE ALWAYS, TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24721294 ↗MalfunctionMar 26, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: WHEN THEY TRIED THE CLIP APPLIER, THERE WAS ONLY ONE CLIP INSIDE. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA TEAMS CALL ON 10MAR26: ALL THE EVENTS TOOK PLACE ON MARCH 8TH BEFORE USE. (B)(4) DEVICES FROM LOT 1560548 HAD THE SAME PROBLEM (ONLY ONE CLIP WAS INSIDE THE DEVICE) AND THEY WERE DISCARDED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 17MAR26: I¿VE JUST CHECKED AND THE CORRECT LOT NUMBER IS 1567140. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 19MAR26: THE USER RESOLVED THE SITUATION CHANGING DEVICE. ONLY ONE CLIP LOADED INTO THE JAWS.12 MM TROCAR WAS USED. THE CLIP WAS LOADED AND VISIBLE BETWEEN THE JAWS OF THE DEVICE, AND IT WAS PROPERLY SEATED. NO (REGARDING CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR) BECAUSE THEY TRIED IT BEFORE PUTTING INSIDE THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP WAS MANUALLY REMOVED FROM THE JAWS. THEY TRIED AGAIN, BUT IT WAS EMPTY. NO OTHER CLIP INSIDE. LIKE ALWAY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24701022 ↗MalfunctionMar 19, 2026Failure to Fire
PROCEDURE PERFORMED: CHOLECYSTECTOMY. EVENT DESCRIPTION: AFTER ENGAGING THE CLIP BEFORE INSERTION INTO THE TROCAR, NO CLIP IS POSITIONED WHEN CLIPPING THE CYSTIC DUCT/ARTERY AND VICE VERSA , CLIP APPEARS WHEN NOT RELEASED. IT WAS RESOLVED BY USING [COMPETITOR DEVICE]. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 18MAR26: NO PICTURE NO VIDEO AVAILABLE. CLIP FALL OUT OF THE JAW. CLIP WAS RETRIEVED FROM THE PATIENT BY USING ATRAUMATIC FORCEPS. NO CLIP PRELOADED BEFORE THE JAWS WERE LOCATED COMPLETELY AROUND THE VESSEL OR STRUCTURE TO BE LIGATED. THE INCIDENT INITIALLY OBSERVED WHEN THE FIRST CLIP WAS LOADED. IT OCCURRED 2 ¿ 3 TIMES. THE CLIP WAS LOADED AND VISIBLE BETWEEN THE JAWS OF THE DEVICE, AND IT WAS PROPERLY SEATED. A CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION THROUGH THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC. NO LOADED CLIP MANUALLY REMOVED FROM THE JAWS. THE DEVICE WAS NOT ACTUATED AND RESET. CLIPS FALL OUT OF THE JAW AND RETAINED BY ATRAUMATIC FORCEPS. THE TRIGGER BE EASILY AND COMPLETELY ACTUATED. NO RESISTANCE IN TRIGGER ACTUATION. INTERVENTION: USED [COMPETITOR DEVICE]. PAT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24644273 ↗MalfunctionMar 19, 2026Failure to Fire
NAME OF PROCEDURE: LAPAROSCOPIC CHOLECYSTECTOMY CASE. EVENT DESCRIPTION: WHILE THE SURGEON WAS USING THE CLIP APPLIER, A CLIP BECAME STUCK INSIDE THE CATHETER. THE CLIP APPLIER WAS THEN SET ASIDE AND NOT USED FURTHER, AND THE SURGEON PROCEEDED WITH A DIFFERENT CLIP APPLIER INSTEAD SURGEON TRIED TO SQUEEZE THE CLIP APPLIER HANDLE PLASTIC TO PLASTIC SEVERAL TIMES BUT THE CLIP STILL STUCK INSIDE CATHETER. NO CLINICAL IMPACT. 20.02.2026 - ADDITIONAL INFORMATION RECEIVED FROM [NAME] (TERRITORY MANAGER) VIA EMAIL: IT WAS A LAPAROSCOPIC CHOLECYSTECTOMY CASE. ADDITIONALLY, ON ASKING "WHERE YOU MENTIONED CATHETER ON THE PCF, DID YOU MEAN CLIP APPLIER?" [NAME] ADVISED IT WAS A CLIP APPLIER, CA500. INTERVENTION: USE A NEW ONE TO CONTINUE THE PROCEDURE. PATIENT STATUS: PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24639366 ↗MalfunctionMar 13, 2026Failure to Fire
PROCEDURE PERFORMED: CHOLECYSTECTOMY EVENT DESCRIPTION: EVENT DATE IS UNKNOWN. I HAVE BEEN CONTACTED BY THE HOSPITAL JUST KNOW ABOUT ANOTHER CA500 DEFECT. THEY WILL BE KEEPING THE SAMPLE TO BE PICKED UP. AND THE ANSWERS ARE THE SAME AS LAST TIME (NO CLIP LOADED INTO THE JAWS. INCIDENT INITIALLY OBSERVED WHEN THE SUBSEQUENT CLIP WAS LOADED. IT OCCURRED AGAIN AND HE OPENED ANOTHER CLIP APPLIED AND IT HAD THE SAME PROBLEM. IT HAPPENED TWICE. NO LOADED CLIP WAS VISIBLE BETWEEN THE JAWS OF THE DEVICE; IT WAS PROPERLY SEATED. NO CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP MANUALLY REMOVED FROM THE JAWS; IT FELL OUT OF THE JAWS. THE TRIGGER COULD BE EASILY AND COMPLETELY ACTUATED. NO RESISTANCE IN TRIGGER ACTUATION, IS JUST THAT THE CLIP WAS NOT GETTING IN PLACE IN THE JAWS), SAME PROCEDURE SAME DOCTOR OPERATING SO IT WAS THE SAME COURSE OF ACTION. APPARENTLY, THIS PROBLEM ONLY STARTED WITH THE SAME LOT NUMBER. AND THE HOSPITAL JUST INFORMED ME THAT THEY WOULD LIKE TO NO LONGER RECEIVE THIS LOT NUMBER. ADDITIONAL INFO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24595708 ↗MalfunctionMar 13, 2026Failure to Fire
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: DURING THE CLOSURE OF THE CYSTIC ARTERIAL, THE SURGEON REPORTS THAT DURING USE, OCCASIONALLY AFTER APPLYING A CLIP, CLOSING THE HANDLE DID NOT IMMEDIATELY RELEASE THE OTHER CLIP, TO RELEASE IT, IT WAS NECESSARY TO CLOSE THE HANDLE AGAIN, THIS EVENT WAS REPEATED AT LEAST TWICE. NO CLINICAL IMPACT. COMPLETED THE PROCEDURE USING THE TOOL BY CLOSING THE HANDLE FOR THE SECOND TIME WHEN THE CLIP DID NOT COME OUT WHEN THE HANDLE WAS CLOSED FOR THE FIRST TIME. INTERVENTION: COMPLETED THE PROCEDURE USING THE TOOL BY CLOSING THE HANDLE FOR THE SECOND TIME WHEN THE CLIP DID NOT COME OUT WHEN THE HANDLE WAS CLOSED FOR THE FIRST TIME. PATIENT STATUS: NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591965 ↗MalfunctionMar 6, 2026Failure to Fire
NAME OF PROCEDURE: CHOLECYSTECTOMY. EVENT DESCRIPTION: WHEN USING THE CLIP APPLIER THERE WAS NO CLIPS COMING OUT OF THE SHAFT AFTER MANY TRIALS. THEY TOOK ANOTHER ONE. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 12FEB26: I HAVE BEEN CONTACTED BY THE HOSPITAL JUST KNOW ABOUT ANOTHER CA500 DEFECT. THEY WILL BE KEEPING THE SAMPLE TO BE PICKED UP. AND THE ANSWERS ARE THE SAME AS LAST TIME (NO CLIP LOADED INTO THE JAWS. INCIDENT INITIALLY OBSERVED WHEN THE SUBSEQUENT CLIP WAS LOADED. IT OCCURRED AGAIN AND HE OPENED ANOTHER CLIP APPLIED AND IT HAD THE SAME PROBLEM. IT HAPPENED TWICE. NO LOADED CLIP WAS VISIBLE BETWEEN THE JAWS OF THE DEVICE, IT WAS PROPERLY SEATED. NO CLIP LOADED INTO THE JAWS PRIOR TO THE INSTRUMENT¿S INSERTION/REMOVAL THROUGH THE TROCAR. THE ACTUATION COMPLETED BY FULLY SQUEEZING THE TRIGGER ¿PLASTIC TO PLASTIC¿ AND ALLOWING THE INSTRUMENT TO REST. NO LOADED CLIP MANUALLY REMOVED FROM THE JAWS, IT FELL OUT OF THE JAWS. THE TRIGGER COULD BE EASILY AND COMPLETELY ACTUATED. NO RESISTANCE IN TRIGGER ACTUATION, IS JUST THAT THE CLIP WAS NOT GETTING IN PLACE IN THE JAWS), SAME PROCEDURE SAME DOCTOR OPERATING SO IT WAS THE SAME COURSE OF A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532649 ↗MalfunctionMar 6, 2026Failure to Fire
NAME OF PROCEDURE: NI EVENT DESCRIPTION: AI TRANSLATED: THE STAPLER COULD NOT BE RELEASED INTRAOPERATIVELY. INTERVENTION: NI PATIENT STATUS: NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532276 ↗
FDA clearance and classification
The FDA regulates product code FZP (clip, implantable) under 21 CFR 878.4300 as a Class II device. 21 CFR 878.4300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Epix® Universal Clip Applier CA500
Has the Epix® Universal Clip Applier CA500 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1388-2024 (Class II), began on Jan 30, 2024: Their is the potential that Universal Clip applier may not load clip after the trigger is actuated. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Epix® Universal Clip Applier CA500?
The FDA's MAUDE database lists 1,176 adverse event reports for this device, received between Jan 25, 2016 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Fire” (619), “Misfire” (129) and “Mechanical Jam” (120); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (12), “Laceration(s)” (11) and “Perforation of Vessels” (8). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Epix® Universal Clip Applier CA500?
1 report is classified as a death, 24 as injuries and 1,151 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Epix® Universal Clip Applier CA500 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Epix® Universal Clip Applier CA500, and is it still sold?
The FDA's device database lists Applied Medical Resources Corporation as the company responsible for it (brand Epix® Universal Clip Applier, model CA500). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Clip, Implantable devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,769 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,943 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,176 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,098 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE779 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death767 reports
- LIGACLIP ER420Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE652 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page