Kii® Blunt Tip Access System C0R85Recalls & reported problems
Applied Medical Resources Corporation · Trocars · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00607915118549), received between Feb 2, 2016 and Oct 4, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment Of Device Component2
- Break1
- Material Separation1
Patient problems
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2022Material Separation; Material Separation
PROCEDURE PERFORMED: LAP HERNIA. EVENT DESCRIPTION: "TURTLENECK" WOULD NOT LOCK. UPON REVIEW, STAFF NOTICED THAT IT WAS MISSING AN "O" RING. DR. [NAME] WAS THE SURGEON. PRODUCT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED VIA EMAIL FOR MEDWATCH # (B)(4) ON 23SEP2022: HUSSAN TROCAR IN PATIENT FOR PROCEDURE. DOCTOR SPONGE PULLED OUT THROUGH HUSSAN TROCAR. TROCAR BECAME DISLODGED. TROCAR REINTRODUCED INTO PATIENT AND LOCKING MECHANISM WOULD NOT LOCK. TROCAR REMOVED FROM FIELD AND INSPECTED. TROCAR FOUND TO BE MISSING A PLASTIC O RING FROM INSIDE THE TURTLENECK PIECE OF TROCAR. CAVITY SEARCHED BY MD. OPERATIONG ROOM SEARCHED BY STAFF. MISSING PIECE NOT LOCATED. ADDITIONAL INFORMATION RECEIVED VIA EMAIL FROM USER FACILITY ON 28SEP2022: UNABLE TO ANSWER ACCURATELY IF THE ENTIRE SEAL HOUSING OR CAP DETACHED. UNSURE OF WHAT TURTLENECK IS. THIS MISSING PIECE WAS AN INNER PIECE. NO PICTURES OF DEVICE ARE AVAILABLE. PATIENT STATUS: DISCHARGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15538100 ↗MalfunctionOct 12, 2016Detachment Of Device Component; Detachment Of Device Component
MEDWATCH REPORT. "TOP PIECE FELL OFF REPEATEDLY. WOULD NOT STAY CLICKED IN. THE PROCEDURE WAS COMPLETED WITH AN IDENTICAL ITEM." WHAT WAS THE ORIGINAL INTENDED PROCEDURE? NOT KNOWN. DEVICE USAGE PROBLEM: DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING)" ADDITIONAL INFORMATION RECEIVED VIA E-MAIL FROM THE CUSTOMER ON SEPTEMBER 14, 2016: THE STAFF DID NOT INDICATE HOW THE SEAL HOUSING FELL OFF OR IF IT WAS EASILY REMOVED. NO INDICATION WAS MENTIONED FROM THE STAFF REGARDING WHETHER OR NOT THE TABS WERE BEING PINCHED AT THE TIME OF BREAKAGE. THE DEVICE IS AVAILABLE FOR RETURN FOR EVALUATION AND THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6024839 ↗MalfunctionMay 18, 2016Break; Break
LAP. CHOLE. - "AT THE END OF PROCEDURE, SURGEON, TO TAKE OFF A GAUZE USED DURING PROCEDURE, CAUGHT IT WITH JOHANN GRASPER AND TOOK IT OFF THROUGH OUR C0R85 TROCAR. AFTER TROCAR EXTRACTION HE REALIZED THAT PART OF TIP TROCAR (FRAGMENT OF ABOUT 5MM) WAS BROKEN. HE COULD NOT FIND ANYMORE PLASTIC PIECE AND HE DOESN'T KNOW IF IT REMAINED IN PATIENT ABDOMEN. EVENT WAS REPORTED TO ITALIAN MINISTRY OF HEALTH AS "INCIDENT." EVENT SAMPLE IS AVAILABLE BUT WE NEED TO WAIT LEGAL TIME BEFORE ORGANIZE ITS PICK UP (AND IF MINISTRY DECIDE NOT TO CHECK IT BY HIMSELF)." PATIENT STATUS - UNKNOWN.
Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient
MAUDE report 5665537 ↗MalfunctionFeb 2, 2016Detachment Of Device Component; Detachment Of Device Component
LAP CHOLE - "TROCAR WENT WITHOUT ANY ISSUE, THEN THE TOP PORTION BROKE (OUTSIDE THE BODY) AS THE CAMERA WAS BEING INSERTED. DR (B)(6) PERFORMED THIS PROCEDURE." PATIENT STATUS - PATIENT IS FINE. MEDWATCH REPORT: LAP CHOLECYSTECTOMY ON (B)(6) 2016 - "TROCAR WAS INSERTED WITHOUT DIFFICULTY. WHEN THE SCOPE WAS PLACED INTO THE TROCAR THE TOP OF THE TROCAR FELL APART. NO HARM TO THE PATIENT, NO FRAGMENTS OR FOREIGN BODIES NEEDED TO BE RETRIEVED FROM INSIDE THE PATIENT."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5405764 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page