Intuitive 5mm Cannula Seal 400161Recalls & reported problems
Applied Medical Resources Corporation · Trocar Seal · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00607915119775), received between Dec 11, 2017 and Jan 17, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Component Falling” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Component Falling1
- Detachment Of Device Component1
- Material Fragmentation1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2018Material Fragmentation; Material Fragmentation
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: THIS CER WAS CREATED FOR THE MEDWATCH RECEIVED 18 DEC 2017 VIA MAIL UNDER (B)(4). THE EVENT DATE WAS (B)(6) 2017. "A PIECE TORE OFF AND FELL INTO THE ABDOMEN. IT WAS REMOVED AND THE ENTIRE ITEM WAS REMOVED FROM THE FIELD" "THE DEVICE IS A 5 MM CANNULA SEAL FOR A DA VINCI CASE, THE SIZE OF THE PIECE THAT FELL INTO PATIENT IS APPROXIMATELY THE SIZE OF A PENCIL LED." "DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING)" THIS WAS NOT A LABORATORY DEVICE OR TEST. THE PATIENT IS NOT (B)(6). ADDITIONAL INFORMATION RECEIVED VIA TELEPHONE ON 20DEC2017 AT 9:05 AM FROM CLINICAL ENGINEER AT (B)(6) MED CTR: THE DEVICE IS AVAILABLE FOR RETURN. TYPE OF INTERVENTION: "IT WAS REMOVED AND THE ENTIRE ITEM WAS REMOVED FROM THE FIELD" PATIENT STATUS: NI.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7201115 ↗MalfunctionDec 11, 2017Detachment Of Device Component; Component Falling
"A PIECE TORE OFF AND FELL INTO THE ABDOMEN. IT WAS REMOVED AND THE ENTIRE ITEM WAS REMOVED FROM THE FIELD" THE DEVICE IS A 5MM CANNULA SEAL FOR A DA VINCI CASE, THE SIZE OF THE PIECE THAT FELL INTO PATIENT IS APPROXIMATELY THE SIZE OF A PENCIL LED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7103971 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page