Skip to content

Kii Optical Access System CFR01Recalls & reported problems

Applied Medical Resources Corporation · Trocar · Class 2

All models of Kii Optical Access System (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915122294), received between Aug 9, 2024 and Oct 17, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2024Material Fragmentation; Material Fragmentation

    PROCEDURE PERFORMED: UNKNOWN EVENT DESCRIPTION: AFTER INCISION THE TROCAR WAS BY ROTATING MOVEMENTS AND APPROPRIATE PRESSURE PLACED IN THE ABDOMINAL WALL. BY PASSING THE TISSUE, A PEACE OF THE SLEEVE WAS BROKEN OFF. THE SURGEON REMOVED THE PEACE OUT OF THE PATIENT BY USING A GRASPER. THE SURGEON FIRST REMOVED THE TROCAR, THAN HE USED A GRASPER TO REMOVE THE PEACE OUT OF THE PATIENTS ABDOMEN. THAN HE OPEN A NEW TROCAR AND PLACED IT WITHOUT PROBLEMS. NO CLINICAL IMPACT. TYPE OF INTEVENTION: OPEND A NEW TROCAR AND PLACED IT WITHOUT PROBLEMS. PATIENT STATUS: NO INJURY OR ILLNESS OCCURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20477220 ↗
  • MalfunctionSep 4, 2024Fracture; Fracture

    PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: (B)(4), COMPLAINT 1 OF 2. (MDR REPORT ID: 2027111-2024-00782). (B)(4), COMPLAINT 2 OF 2. (MDR REPORT ID: 2027111-2024-00819). IN (B)(6) 2024, TWO DIFFERENT, DR. [NAME] ENCOUNTERED ISSUES WITH A SPECIFIC TROCAR (CFR01) DURING LAPAROSCOPIC CHOLECYSTECTOMY (LAB CHOLE) PROCEDURES. SURGEON REPORTED THAT THE TROCAR BROKE DURING THE SURGICAL CASES. A RESULT OF THESE INCIDENTS, THE SURGEON STOPPED USING THE CFR01 TROCAR CATALOG AND HAVE A SUBSTANTIAL REMAINING INVENTORY OF THIS PRODUCT. THEY WOULD LIKE TO UNDERSTAND WHY THE TROCAR BROKE DURING THE PROCEDURES SO THEY CAN DETERMINE IF THERE ARE ANY ISSUES WITH THE DESIGN OR MANUFACTURING OF THIS PARTICULAR DEVICE. THERE IS NO IMPACT CLINICAL IMPACT ON THE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM THE FIELD TEAM ON SEPTEMBER 4, 2024 THROUGH EMAIL: YES, THERE WERE TWO INCIDENTS INVOLVING CFR01 WITH TWO DIFFERENT SURGEONS. ADDITIONAL INFORMATION RECEIVED FROM THE FIELD TEAM ON SEPTEMBER 23: THE SURGEON EXPERIENCED THE SAME TYPE OF TROCAR BREAKAGE. TYPE OF INTERVENTION: THEY REPLACED THE DEVICE WITH OTHER TROCAR FROM DIFFERENT MANUFACTURE. PATIENT STATUS: PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20138232 ↗
  • MalfunctionAug 9, 2024Fracture; Fracture

    PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: IN MAY 2024, TWO DIFFERENT SURGEONS DR. [NAME] AND DR. [NAME], ENCOUNTERED ISSUES WITH A SPECIFIC TROCAR (CFR01) DURING SEPARATE LAPAROSCOPIC CHOLECYSTECTOMY (LAB CHOLE) PROCEDURES. BOTH SURGEONS REPORTED THAT THE TROCAR BROKE DURING THE SURGICAL CASES. AS A RESULT OF THESE INCIDENTS, THE SURGEONS STOPPED USING THE CFR01 TROCAR CATALOG AND HAVE SUBSTANTIAL REMAINING INVENTORY OF THIS PRODUCT. THEY WOULD LIKE TO UNDERSTAND WHY THE TROCAR BROKE DURING THE PROCEDURES SO THEY CAN DETERMINE IF THERE ARE ISSUES WITH THE DESIGN OR MANUFACTURING OF THIS PARTICULAR DEVICE. TYPE OF INTERVENTION: THEY REPLACED THE DEVICE WITH ANOTHER TROCAR FROM DIFFERENT MANUFACTURER. PATIENT STATUS: PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19955908 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →