Kii® Optical Dual Pack CTR12Recalls & reported problems
Applied Medical Resources Corporation · Laparoscopic Trocar Sleeve/Single Use · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915123505), received between Mar 17, 2020 and May 8, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: DURING A DIFFICULT LAPAROSCOPIC CHOLECYSTECTOMY, AMIDST MULTIPLE INSTRUMENT EXCHANGE, ONE OF THE VALVES DETACHED FROM THE PORT AND FELL INTO THE ABDOMINAL CAVITY, IT WAS THE TRANSPARENT PLASTIC PART THAT SITS ON TOP OF THE BLACK DOUBLE DUCKBILL. THE SURGEON IN QUESTION WAS A [NAME REDACTED] WHO COMPLETED THE PROCEDURE WITH ANOTHER TROCAR. THE THEATRE CO-ORDINATOR, [NAME REDACTED] HAS KEPT THE PORT IN QUESTION AND IT IS AVAILABLE TO BE SENT TO YOU. INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL 5MAY23: SURGERY COMPLETED AND PATIENT TRANSFERRED TO RECOVERY AREA. ALL PARTS WERE RETRIEVED FROM THE PATIENT. CLEAR SEAL VALVE CAME OFF NOT THE ENTIRE COMPONENT OF THE PORT. ONE SIDE OF THE VALVE APPEARS TORN OFF. AT TIME OF THE ACCIDENT, LAPAROSCOPIC FENESTRATED GRASPER WAS BEING USED. PRIOR TO THE ACCIDENT, FENESTRATED GRASPER WAS USED. INTERNAL SEAL COMPONENTS WERE NOT REMOVED OR CUT IN ORDER TO INSPECT THE DAMAGED COMPONENT. SURGEON COMPLETED THE PROCEDURE WITH ANOTHER 5 MM PORT, CTF03. TYPE OF INTERVENTION: PROCEDURE COMPLETED WITH OTHER TROCAR (CTF03). PATIENT STATUS: NO PATIENT INJURY - SURGERY C
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16892416 ↗MalfunctionMar 9, 2022Fracture; Fracture
PROCEDURE PERFORMED: LAP APPENDECTOMY. EVENT DESCRIPTION: ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 7-MAR-2022: THE PATIENT HAD 2 PREVIOUS C-SECTIONS AS THEIR ONLY PREVIOUS SURGERY. DURING THE PROCEDURE THERE WAS LOTS OF TWISTING AND PUSHING. AFTER ABOUT 15 MINUTES OF TRYING THE TROCAR ENTERED THE ABDOMEN WITH THE TIP HANGING OFF SLIGHTLY. THE REGISTRAR USED A GRASPER TO REMOVE THE TIP BEFORE IT COULD DROP INTO THE SPACE. THE TROCAR WAS THEN REMOVED AND REPLACED WITH ANOTHER WITHOUT ISSUES THE PROCEDURE WAS A LAP APPENDECTOMY. THIS WAS THE THIRD TROCAR PLACED, 2 OTHERS WERE ALREADY IN PLACE. THE TROCAR WAS INSERTED UNDER VISION. OTHER THAN THE OTHER TROCARS, THERE WERE NO INSTRUMENTS IN THE OPERATING FIELD. THE TROCAR WAS INSERTED BELOW THE UMBILICUS, LEFT ABDOMINAL. AS FAR AS THE CUSTOMER IS AWARE, THE TIP OF THE OBTURATOR DID NOT COME INTO CONTACT WITH HARD/DENSE SURFACES. IN SURGERY, THE STAFF WAS HAPPY WITH THE RETRIEVAL, HOWEVER CONFIRMATION WILL FOLLOW IF ALL PIECES WERE RETRIEVED. THE DEVICE HAS BEEN DISPOSED OF. INTERVENTION: CHANGE OF DEVICE. PATIENT STATUS: UNAWARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13715789 ↗MalfunctionMar 17, 2020Break; Break
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY EVENT DESCRIPTION: ACCORDING TO THE STOCK CONTROLLER, THE TROCAR BROKE OFF IN THE PATIENT. ADDITIONAL INFORMATION RECEIVED FROM DISTRIBUTOR VIA EMAIL 20: THE SLEEVE AND OBTURATOR BROKE. NO PIECES FELL INTO THE PATIENT. DOCTOR PUSHED INTO THE PATIENT AND BROKE WHILE DOING IT. THE TROCAR WAS CLEANLY BROKEN. A NEW TROCAR WAS OPENED. PATIENT STATUS: NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9845614 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page