Kii Low Profile Bladed Dual Pack CTB23Recalls & reported problems
Applied Medical Resources Corporation · Trocar · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915126582), received between Dec 18, 2025 and Dec 18, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 11, 2025Class IIOpen, Classified
There is the potential for extended length of the blade obturator after assembly into the cannula and seal housing.
Z-1010-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Component3
- Nonstandard Device3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 18, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: BEFORE USING THE 5 MM DIAMETER TROCAR, THE SURGEON NOTICED THAT ITS MANDREL PROTRUDED MUCH MORE THAN USUAL FROM THE SLEEVE. THE SURGEON REFUSED TO USE IT, CONSIDERING IT DANGEROUS. ALERT FROM THE SURGEON REGARDING AN ANOMALY IN THE LENGTH OF THE LAPAROSCOPIC TROCAR MANDREL. ABNORMALLY LONG (A0201). A DIFFERENCE OF NEARLY 1 CM WAS OBSERVED COMPARED TO TWO OTHER BATCHES WITH THE SAME REFERENCE NUMBER. ADDITIONAL INFORMATION RECEIVED FROM AN APPLIED MEDICAL REPRESENTATIVE VIA PCF ON 01DEC25: HE WASN¿T SURE WHETHER IT WAS THE SHIRT THAT WAS TOO SHORT OR THE PYRAMIDAL-POINTEDMANDREL THAT WAS TOO LONG. AFTER CHECKING, THEY FOUND THAT ONLY THE LOT NUMBER INDICATED HAD THIS ANOMALY. ADDITIONAL INFORMATION RECEIVED FROM AN APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 02DEC25: THE NURSE INITIALLY HANDED THE SURGEON A TROCAR WHICH, UPON INSPECTION BEFORE USE, WAS FOUND TO BE NON-COMPLIANT WITH THE TYPE NORMALLY USED. THE NURSE THEN PROVIDED A SECOND TROCAR, WHICH TURNED OUT TO BE IDENTICAL TO THE FIRST ONE. CONSEQUENTLY, THE TEAM SWITCHED TO A DIFFERENT BOX AND LOT NUMBER TO OBTAIN A COMPLIANT TROCAR. THE TWO NON-COMPLIANT TROCARS, ALONG WITH THE BOX
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23844322 ↗MalfunctionDec 18, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: A PROBLEM HAS BEEN REPORTED CONCERNING A 5 MM LAPAROSCOPY TROCAR COMPOSED OF TWO PARTS: THE SHARP METAL MANDREL WAS TOO LONG IN RELATION TO THE OUTLET PORT OF THE PLASTIC CANNULA, THUS MAKING THE BATCH DANGEROUS. ADDITIONAL INFORMATION RECEIVED BY AN APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 10DEC25: THE PRODUCT HAS NOT BEEN USED ON THE PATIENT. NO PATIENT INJURY OCCURED AS A RESULT OF THIS EVENT. HOW DID THE USER RESOLVE THE SITUATION? TAKE A NEW PRODUCT. TIMING OF THE EVENT: BEFORE THE BEGINING. QUESTIONS ABOUT BLADE EXPOSED NOT APPLICABLE FOR THIS REFERENCE. DID ANYONE (PATIENT OR STAFF) GOT INJURED? NO. PATIENT STATUS: NO PATIENT INJURY OR ILLNESS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23844280 ↗MalfunctionDec 18, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
PROCEDURE PERFORMED: NI. EVENT DESCRIPTION: BEFORE USING THE 5 MM DIAMETER TROCAR, THE SURGEON NOTICED THAT ITS MANDREL PROTRUDED MUCH MORE THAN USUAL FROM THE SLEEVE. THE SURGEON REFUSED TO USE IT, CONSIDERING IT DANGEROUS. ALERT FROM THE SURGEON REGARDING AN ANOMALY IN THE LENGTH OF THE LAPAROSCOPIC TROCAR MANDREL. ABNORMALLY LONG (A0201). A DIFFERENCE OF NEARLY 1 CM WAS OBSERVED COMPARED TO TWO OTHER BATCHES WITH THE SAME REFERENCE NUMBER. ADDITIONAL INFORMATION RECEIVED FROM AN APPLIED MEDICAL REPRESENTATIVE VIA PCF ON 01DEC2025: HE WASN¿T SURE WHETHER IT WAS THE SHIRT THAT WAS TOO SHORT OR THE PYRAMIDAL-POINTEDMANDREL THAT WAS TOO LONG. AFTER CHECKING, THEY FOUND THAT ONLY THE LOT NUMBER INDICATED HAD THIS ANOMALY. ADDITIONAL INFORMATION RECEIVED FROM AN APPLIED MEDICAL REPRESENTATIVE VIA EMAIL ON 02DEC2025: THE NURSE INITIALLY HANDED THE SURGEON A TROCAR WHICH, UPON INSPECTION BEFORE USE, WAS FOUND TO BE NON-COMPLIANT WITH THE TYPE NORMALLY USED. THE NURSE THEN PROVIDED A SECOND TROCAR, WHICH TURNED OUT TO BE IDENTICAL TO THE FIRST ONE. CONSEQUENTLY, THE TEAM SWITCHED TO A DIFFERENT BOX AND LOT NUMBER TO OBTAIN A COMPLIANT TROCAR. THE TWO NON-COMPLIANT TROCARS, ALONG WITH THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23844237 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page