Skip to content

Kii® Blunt Tip Access System C0R95Recalls & reported problems

Applied Medical Resources Corporation · Trocars · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00607915127602), received between Mar 2, 2017 and Sep 11, 2023 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Fragmentation2
  • Break1
  • Detachment of Device or Device Component1
  • Physical Resistance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2023Material Fragmentation; Material Fragmentation

    PROCEDURE PERFORMED: NA (BEFORE USE). EVENT DESCRIPTION: PLASTIC ATTACHMENT PIECE CRACKED OFF DISPOSABLE LAPAROSCOPIC TROCAR DURING SET-UP. NO CONTACT WITH PATIENT, ITEM REMOVED FROM SURGICAL FIELD. 1 ITEM IMPACTED, NOT A REPEAT ISSUE. TYPE OF INCIDENT/PROBLEM: DEFECT (DETECTED BEFORE USE) LEVEL OF HARM: NO EFFECT - DID NOT REACH PATIENT - DETECTED BEFORE USE OR DOES NOT TOUCH PERSON DURING USE OPTIONAL REPORT TO CMDSNET (HEALTH CANADA): YES, HOSPITAL SUBMITTED MDIP (B)(4) COMPLAINT SAMPLE IS RETAINED FOR RETURN AND IS CLEAN/NOT USED PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL ON 26APR2023 FROM DISTRIBUTOR: THE EVENT DATE SHOULD HAVE BEEN (B)(6) 2023. PATIENT STATUS: NA (BEFORE USE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17725690 ↗
  • MalfunctionAug 3, 2021Break; Break

    PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: NURSE UNIT MANAGER SAID THAT THE TROCAR SNAPPED UPON INSERTION. DEVICE HAS BEEN RETAINED TO SHIP BACK FOR ANALYSIS. ADDITIONAL INFORMATION RECEIVED VIA EMAIL FROM [APPLIED MEDICAL REPRESENTATIVE] ON (B)(6) 2021: I WAS TOLD BY THE NURSE UNIT MANAGER AT THE TIME OF NOTIFICATION THAT THE PATIENT WAS FINE AND NO HARM DONE. PATIENT STATUS: THE PATIENT WAS FINE AND NO HARM DONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12268959 ↗
  • MalfunctionMar 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    PROCEDURE PERFORMED: UNKNOWN. EVENT DESCRIPTION: THE INTERIOR VALVE OF THE C0R95 BECAME DISLODGED FROM THE PORT AND WENT INTO THE ABDOMEN DURING THE PROCEDURE. THE PART WAS OBTAINED AND THERE WAS NO HARM TO THE PATIENT. PATIENT STATUS: PATIENT IS FINE, AND COMPLETELY UNAFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11445773 ↗
  • MalfunctionApr 14, 2017Material Fragmentation; Material Fragmentation

    PROCEDURE PERFORMED: LAP HERNIA. INCIDENT DESCRIPTION: "FEEDING MESH DOWN THROUGH SEAL WITH A [COMPETITOR] GRASPER, THE RUBBER ON THE SEAL DISLODGED, THE SURGEON IDENTIFIED THE PIECE SITTING IN THE INCISION AND RETRIEVED IT SAFELY." TYPE OF INTERVENTION: "THE SURGEON IDENTIFIED THE PIECE SITTING IN THE INCISION AND RETRIEVED IT SAFELY." PATIENT STATUS: UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6494218 ↗
  • MalfunctionMar 2, 2017Physical Resistance; Physical Resistance

    LAP PROSTATE- "SURGEON COMPLAINING OF CAMERA DRAG THROUGH C0R95 PORT. DRAG NOTED WHEN MOVING SCOPE INTO AND OUT OF THE PORT. CAMERA IS OLYMPUS ENDOEYE 10MM. SCOPE CLEANING SOLUTION USED ALSO." PATIENT STATUS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6372807 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page