AVANOS 93816Recalls & reported problems
Avanos Medical, Inc. · 1500 Flex Heat and Moisture Exchanger · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Avanos heat and moisture exchanger warms and moistens air for people using breathing support. The 24 reports mostly describe the device coming apart or loosening from its connection, with alarms sounding and staff replacing the unit, with no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00609038938165), received between Mar 3, 2020 and Apr 10, 2023 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Disconnection” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Disconnection21
- Loose or Intermittent Connection3
Patient problems
- No Consequences Or Impact To Patient21
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
ADDITIONAL INFORMATION RECEIVED 10MAY2023 REPORTED, THE RESPIRATORY THERAPIST (RT) RESPONDED TO THE ALARM WITHIN SECONDS AND CHANGED THE VYAIRE HEAT AND MOISTURE EXCHANGERS (HME) WHICH RESOLVED THE ISSUE AND THERE WAS NO ADVERSE OUTCOME TO THE RESIDENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16707005 ↗MalfunctionApr 10, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
ADDITIONAL INFORMATION RECEIVED 10MAY2023 REPORTED, THE RESPIRATORY THERAPIST (RT) RESPONDED TO THE ALARM WITHIN SECONDS AND CHANGED THE VYAIRE HEAT AND MOISTURE EXCHANGERS (HME) WHICH RESOLVED THE ISSUE AND THERE WAS NO ADVERSE OUTCOME TO THE RESIDENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16707003 ↗MalfunctionApr 10, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
ADDITIONAL INFORMATION RECEIVED 10MAY2023 REPORTED, THE RESPIRATORY THERAPIST (RT) RESPONDED TO THE ALARM WITHIN SECONDS AND CHANGED THE VYAIRE HEAT AND MOISTURE EXCHANGERS (HME) WHICH RESOLVED THE ISSUE AND THERE WAS NO ADVERSE OUTCOME TO THE RESIDENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16707001 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE FIFTEENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102166 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE THIRTEENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102163 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE SEVENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT . REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTAC
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102159 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE NINTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102134 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE FOURTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102132 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE SECOND OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102128 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE FIFTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102123 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
ADDITIONAL INFORMATION RECEIVED FROM THE USER FACILITY ON 01-JUN-2020 CONFIRMED THAT THIS REPORT IS A DUPLICATE OF AVANOS MEDWATCH 3011270181-2020-00041.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102111 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE TWELFTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102110 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
ADDITIONAL INFORMATION RECEIVED FROM THE USER FACILITY ON 01-JUN-2020 CONFIRMED THAT THIS REPORT IS A DUPLICATE OF AVANOS MEDWATCH 3011270181-2020-00039.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102105 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE ELEVENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102096 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE THIRD OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102095 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE SIXTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102092 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
ADDITIONAL INFORMATION RECEIVED FROM THE USER FACILITY ON 01-JUN-2020 CONFIRMED THAT THIS REPORT IS A DUPLICATE OF MEDWATCH 3011270181-2020-00040.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102086 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE FOURTEENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102081 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE TENTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102077 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE EIGHTH OF 18 REPORTS. REFER TO 3011270181-2020-00064 FOR THE FIRST REPORT. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102073 ↗MalfunctionMay 29, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 18 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 18 DIFFERENT PATIENTS. THIS IS THE FIRST OF 18 REPORTS. REFER TO 3011270181-2020-00065 FOR THE SECOND REPORT. REFER TO 3011270181-2020-00066 FOR THE THIRD REPORT. REFER TO 3011270181-2020-00067 FOR THE FOURTH REPORT. REFER TO 3011270181-2020-00068 FOR THE FIFTH REPORT. REFER TO 3011270181-2020-00069 FOR THE SIXTH REPORT. REFER TO 3011270181-2020-00070 FOR THE SEVENTH REPORT. REFER TO 3011270181-2020-00071 FOR THE EIGHTH REPORT. REFER TO 3011270181-2020-00072 FOR THE NINTH REPORT. REFER TO 3011270181-2020-00073 FOR THE TENTH REPORT. REFER TO 3011270181-2020-00074 FOR THE ELEVENTH REPORT. REFER TO 3011270181-2020-00075 FOR THE TWELFTH REPORT. REFER TO 3011270181-2020-00076 FOR THE THIRTEENTH REPORT. REFER TO 3011270181-2020-00077 FOR THE FOURTEENTH REPORT. REFER TO 3011270181-2020-00078 FOR THE FIFTEENTH REPORT. REFER TO 3011270181-2020-00079 FOR THE SIXTEENTH REPORT. REFER TO 3011270181-2020-00080 FOR THE SEVENTEENTH REPORT. REFER TO 3011270181-2020-00081 FOR THE EIGHTEENTH REPORT. IT WAS REPORTED THAT THE HEAT AND MOISTURE EXCHANGER (HME) ATTACH
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10102070 ↗MalfunctionMar 3, 2020Disconnection; Disconnection
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 3011270181-2020-00040 FOR THE SECOND EVENT. REFER TO 3011270181-2020-00041 FOR THE THIRD EVENT. IT WAS REPORTED THE HEAT AND MOISTURE EXCHANGER ATTACHED TO THE END OF THE ENDOTRACHEAL TUBE (ETT) AND TO THE VENTILATOR CIRCUIT DISCONNECTED DURING VENTILATION. THERE WAS NO REPORTED INJURY AND THE DEVICE WAS REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9779426 ↗MalfunctionMar 3, 2020Disconnection; Disconnection
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3011270181-2020-00039 FOR THE FIRST EVENT. REFER TO 3011270181-2020-00041 FOR THE THIRD EVENT. IT WAS REPORTED THE HEAT AND MOISTURE EXCHANGER ATTACHED TO THE END OF THE ENDOTRACHEAL TUBE (ETT) AND TO THE VENTILATOR CIRCUIT DISCONNECTED DURING VENTILATION. THERE WAS NO REPORTED INJURY AND THE DEVICE WAS REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9779425 ↗MalfunctionMar 3, 2020Disconnection; Disconnection
AVANOS MEDICAL RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT EVENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 3011270181-2020-00039 FOR THE FIRST EVENT. REFER TO 3011270181-2020-00040 FOR THE SECOND EVENT. IT WAS REPORTED THE HEAT AND MOISTURE EXCHANGER ATTACHED TO THE END OF THE ENDOTRACHEAL TUBE (ETT)AND TO THE VENTILATOR CIRCUIT DISCONNECTED DURING VENTILATION. THERE WAS NO REPORTED INJURY AND THE DEVICE WAS REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9779424 ↗
Other CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices
- HUDSON RCI IPN042780Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall3 reports
- DAR 353S19046Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- HUDSON RCI IPN042781Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- DAR 354/5876Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)25 reports
- HUDSON RCI IPN040673Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)11 reports
- Portex C45031682D-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
- Portex 450952-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
- HUDSON RCI IPN044386Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record