Skip to content

Portex C45031682D-NLRecalls & reported problems

Icu Medical, Inc. · Class 1

All models of Portex (409)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 15019315089381), received between Feb 13, 2025 and Apr 16, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak6
  • Material Puncture/Hole5
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THE DEVICE EXHIBITED A HOLE IN THE CIRCUIT. THE DEVICE WAS OPERATED UNDER ANESTHESIA. THERE WAS NO PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897000 ↗
  • MalfunctionOct 17, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT LEAKAGE/HOLES WERE FOUND IN THE TUBING INTRAOPERATION WHILE THE ANESTHESIA CIRCUIT WAS CONNECTED TO AN INTUBATED PATIENT. THE CIRCUIT WAS REPLACED TO ADDRESS THIS ISSUE AND A PREGNANT TECHNICIAN IN THE OPERATING ROOM WAS EXPOSED TO ANESTHESIA GASSES. NO PATIENT HARM/ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23327734 ↗
  • MalfunctionMay 12, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THE ANESTHESIA REPEATEDLY FAILED THE LEAK CHECK. A PIN SIZED HOLE WAS FOUND IN THE MACHINE CIRCUIT WHICH MIGHT NEGATIVELY IMPACTED THE PATIENT AND IMPAIRED VENTILATION. THIS HAPPENED AGAIN AFTER EXCHANGE FOR A NEW CIRCUIT. THE EVENT WAS FOUND BEFORE STARTING THE THERAPY. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22002881 ↗
  • MalfunctionApr 1, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THE ANESTHESIA REPEATEDLY FAILED THE LEAK CHECK. A PIN SIZED HOLE WAS FOUND IN THE MACHINE CIRCUIT WHICH MIGHT NEGATIVELY IMPACTED THE PATIENT AND IMPAIRED VENTILATION. THIS HAPPENED AGAIN AFTER EXCHANGE FOR A NEW CIRCUIT. THE EVENT WAS FOUND BEFORE STARTING THE THERAPY. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21741397 ↗
  • MalfunctionApr 1, 2025Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak

    IT WAS REPORTED THE ANESTHESIA REPEATEDLY FAILED THE LEAK CHECK. A PIN SIZED HOLE WAS FOUND IN THE MACHINE CIRCUIT WHICH MIGHT NEGATIVELY IMPACTED THE PATIENT AND IMPAIRED VENTILATION. THIS HAPPENED AGAIN AFTER EXCHANGE FOR A NEW CIRCUIT. THE EVENT WAS FOUND BEFORE STARTING THE THERAPY. THERE WAS NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21741227 ↗
  • MalfunctionMar 13, 2025Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole

    IT WAS REPORTED THAT THREE ANESTHESIA CIRCUITS DID NOT PASS THE DAILY LEAKAGE TESTS BECAUSE THE TUBING HAD HOLES IN THEM. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21596619 ↗
  • MalfunctionFeb 13, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE PRODUCTS HAVE BEEN LEAKING. DURING THE LEAK TEST, THE PRODUCTS FAILED WITH GREATER THAN 450ML OF LOSS ON MULTIPLE ANESTHESIA MACHINES. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21379743 ↗
  • MalfunctionFeb 13, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE PRODUCTS HAVE BEEN LEAKING. DURING THE LEAK TEST, THE PRODUCTS FAILED WITH GREATER THAN 450ML OF LOSS ON MULTIPLE ANESTHESIA MACHINES. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21379708 ↗

Other CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices

All CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page