HUDSON RCI IPN040673Recalls & reported problems
Teleflex Incorporated · HUMID VENT MINI STERILE · Class 1
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04026704347810), received between Jun 3, 2021 and Feb 10, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Moisture or Humidity Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Moisture or Humidity Problem10
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
Patient problems
- Unspecified Respiratory Problem10
- Ventilator Dependent1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342295 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342280 ↗MalfunctionFeb 10, 2025Fluid/Blood Leak; Moisture or Humidity Problem; Fluid/Blood Leak; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342257 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00046, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342238 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342228 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342218 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342175 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342119 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21342108 ↗MalfunctionFeb 10, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT: "CHILDREN TRY TO TAKE IT OFF, WHICH IS A RISK. CUSTOMER THINKS THIS ITEM GETS WET TOO EASILY AND THAT IS WHY CHILDREN TRY TO TAKE THEM OFF BY THEMSELVES." ASSOCIATED COMPLAINTS 8040412-2025-00031, 8040412-2025-00040, 8040412-2025-00041, 80404 12-2025-00042, 8040412-2025-00043, 8040412-2025-00044, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00045, 8040412-2025-00046, 8040412-2025-00047 AND 8040412-2025-00048.
Patient problems: Unspecified Respiratory Problem
MAUDE report 21341970 ↗MalfunctionJun 3, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT " WE HAVE HAD A FAMILY ON CASELOAD WHO HAS HAD ISSUES WITH THE HUMID-VENT, THEY HAVE BEEN USING THEM FOR A WHILE. THEY HAD RANG TO HAVE MORE STOCK AS THE WHITE FILTER WAS FALLING OUT SO WAS INEFFECTIVE". NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: Ventilator Dependent
MAUDE report 11934976 ↗
Other CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE) devices
- HUDSON RCI IPN042780Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall3 reports
- DAR 353S19046Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- HUDSON RCI IPN042781Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)1 recall1 report
- DAR 354/5876Covidien LP · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)25 reports
- AVANOS 93816Avanos Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)24 reports
- Portex C45031682D-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
- Portex 450952-NLIcu Medical, Inc. · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)8 reports
- HUDSON RCI IPN044386Teleflex Incorporated · CONDENSER, HEAT AND MOISTURE (ARTIFICIAL NOSE)7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page