Sure-Vue 4581415021Recalls & reported problems
Fisher Scientific Company L.L.C. · SURE-VUE HCGDIPSTICK URIN25/PK · Class 2
All models of Sure-Vue (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00613647000588), received between Nov 14, 2023 and Jul 2, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 20, 2026Class IIOpen, Classified
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Z-3273-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- False Positive Result6
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING FALSE POSITIVE HCG RESULTS WHILE USING FISHER SURE-VUE HCG URINE DIPSTICKS ON AN UNKNOWN NUMBER OF PATIENTS WITH TWO LOTS. EACH PATIENT PRESENTED FOR A PREGNANCY CHECK PRIOR TO PERFORMING AN UNSPECIFIED NON-EMERGENT DIAGNOSTIC PROCEDURE. ALL OF THE PATIENTS PROVIDED A FRESH URINE SAMPLE THAT WAS TESTED IMMEDIATELY. THE RESULTS WERE READ AT 4 MINUTES AND POSITIVE HCG RESULTS WERE OBTAINED. AS A RESULT, THE DIAGNOSTIC PROCEDURES WERE DELAYED 24 HOURS AS CONFIRMATORY BLOOD TESTING WAS PERFORMED. NO ADVERSE EVENTS WERE REPORTED. NO FURTHER INFORMATION WAS PROVIDED. PER THE PACKAGE INSERT, "ALLOW THE TEST STRIP, URINE SPECIMEN AND/OR CONTROLS TO EQUILIBRATE TO ROOM TEMPERATURE (15-30°C) PRIOR TO TESTING." SEE MDR 2027969-2024-00088 FOR THE EVENT ON LOT NUMBER 0000695546.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19662199 ↗MalfunctionJul 2, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING FALSE POSITIVE HCG RESULTS WHILE USING FISHER SURE-VUE HCG URINE DIPSTICKS ON AN UNKNOWN NUMBER OF PATIENTS WITH TWO LOTS. EACH PATIENT PRESENTED FOR A PREGNANCY CHECK PRIOR TO PERFORMING AN UNSPECIFIED NON-EMERGENT DIAGNOSTIC PROCEDURE. ALL OF THE PATIENTS PROVIDED A FRESH URINE SAMPLE THAT WAS TESTED IMMEDIATELY. THE RESULTS WERE READ AT 4 MINUTES AND POSITIVE HCG RESULTS WERE OBTAINED. AS A RESULT, THE DIAGNOSTIC PROCEDURES WERE DELAYED 24 HOURS AS CONFIRMATORY BLOOD TESTING WAS PERFORMED. NO ADVERSE EVENTS WERE REPORTED. NO FURTHER INFORMATION WAS PROVIDED. PER THE PACKAGE INSERT, "ALLOW THE TEST STRIP, URINE SPECIMEN AND/OR CONTROLS TO EQUILIBRATE TO ROOM TEMPERATURE (15-30°C) PRIOR TO TESTING." SEE MDR 2027969-2024-00089 FOR THE EVENT ON LOT NUMBER 0000695335.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19661311 ↗MalfunctionNov 14, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING A TOTAL OF FOUR FALSE POSITIVE HCG RESULTS ON THE SURE-VUE HCG URINE TEST FOR FOUR PATIENTS. THE FOUR PATIENTS WERE TESTED IN AN INFERTILITY CLINIC AND THE CUSTOMER REPORTED NOTED 3 FALSE POSITIVE RESULTS ON LOT HCG2042082 AND 1 ON LOT HCG2042074. A FRESH URINE SAMPLE WAS PROVIDED IN EACH INSTANCE AND THE SAMPLES APPEARED "NORMAL". THE CUSTOMER INDICATED THE TEST WAS PERFORMED ACCORDING TO THE PACKAGE INSERT AND THE RESULTS WERE READ AT 4 MINUTES. REPEAT TESTS WITH THE SAME KITS WERE PERFORMED FOR ALL FOUR PATIENTS AND NEGATIVE RESULTS WERE OBTAINED. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2023-00107 AND 2027969-2023-00109 FOR ADDITIONAL FALSE POSITIVE RESULTS ON LOT HCG2042082 AND MDR 2027969-2023-00108 FOR THE REPORTED FALSE POSITIVE ON LOT HCG2042074.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18136219 ↗MalfunctionNov 14, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING A TOTAL OF FOUR FALSE POSITIVE HCG RESULTS ON THE SURE-VUE HCG URINE TEST FOR FOUR PATIENTS. THE FOUR PATIENTS WERE TESTED IN AN INFERTILITY CLINIC AND THE CUSTOMER REPORTED NOTED 3 FALSE POSITIVE RESULTS ON LOT HCG2042082 AND 1 ON LOT HCG2042074. A FRESH URINE SAMPLE WAS PROVIDED IN EACH INSTANCE AND THE SAMPLES APPEARED "NORMAL". THE CUSTOMER INDICATED THE TEST WAS PERFORMED ACCORDING TO THE PACKAGE INSERT AND THE RESULTS WERE READ AT 4 MINUTES. REPEAT TESTS WITH THE SAME KITS WERE PERFORMED FOR ALL FOUR PATIENTS AND NEGATIVE RESULTS WERE OBTAINED. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2023-00107 AND 2027969-2023-00110 FOR ADDITIONAL FALSE POSITIVE RESULTS ON LOT HCG2042082 AND MDR 2027969-2023-00108 FOR THE REPORTED FALSE POSITIVE ON LOT HCG2042074.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18136193 ↗MalfunctionNov 14, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING A TOTAL OF FOUR FALSE POSITIVE HCG RESULTS ON THE SURE-VUE HCG URINE TEST FOR FOUR PATIENTS. THE FOUR PATIENTS WERE TESTED IN AN INFERTILITY CLINIC AND THE CUSTOMER REPORTED NOTED 3 FALSE POSITIVE RESULTS ON LOT HCG2042082 AND 1 ON LOT HCG2042074. A FRESH URINE SAMPLE WAS PROVIDED IN EACH INSTANCE AND THE SAMPLES APPEARED "NORMAL". THE CUSTOMER INDICATED THE TEST WAS PERFORMED ACCORDING TO THE PACKAGE INSERT AND THE RESULTS WERE READ AT 4 MINUTES. REPEAT TESTS WITH THE SAME KITS WERE PERFORMED FOR ALL FOUR PATIENTS AND NEGATIVE RESULTS WERE OBTAINED. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2023-00107, 2027969-2023-00110 AND 2027969-2023-00109 FOR ADDITIONAL FALSE POSITIVE RESULTS ON LOT HCG2042082.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18134400 ↗MalfunctionNov 14, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED RECEIVING A TOTAL OF FOUR FALSE POSITIVE HCG RESULTS ON THE SURE-VUE HCG URINE TEST FOR FOUR PATIENTS. THE FOUR PATIENTS WERE TESTED IN AN INFERTILITY CLINIC AND THE CUSTOMER REPORTED NOTED 3 FALSE POSITIVE RESULTS ON LOT HCG2042082 AND 1 ON LOT HCG2042074. A FRESH URINE SAMPLE WAS PROVIDED IN EACH INSTANCE AND THE SAMPLES APPEARED "NORMAL". THE CUSTOMER INDICATED THE TEST WAS PERFORMED ACCORDING TO THE PACKAGE INSERT AND THE RESULTS WERE READ AT 4 MINUTES. REPEAT TESTS WITH THE SAME KITS WERE PERFORMED FOR ALL FOUR PATIENTS AND NEGATIVE RESULTS WERE OBTAINED. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2023-00108 AND 2027969-2023-00109 FOR ADDITIONAL FALSE POSITIVE RESULTS ON LOT HCG2042082 AND MDR 2027969-2023-00110 FOR THE REPORTED FALSE POSITIVE ON LOT HCG2042074.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18134237 ↗
Other Visual, pregnancy hCG, prescription use devices
- McKesson 5001Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls165 reports
- CARDINAL HEALTH B1077-23Cardinal Health 200, LLC · Visual, pregnancy hCG, prescription use1 recall149 reports
- McKesson 5000Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls90 reports
- McKesson 5002Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls33 reports
- Sure-Vue 4581405021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall83 reports
- Henry Schein 9004072Henry Schein, Inc. · Visual, pregnancy hCG, prescription use1 recall79 reports
- AlereTM 92210Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls23 reports
- MEDLINE MPH12025HCGMedline Industries, Inc. · Visual, pregnancy hCG, prescription use1 recall69 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page