Skip to content

ROHO QUADTRO SELECT HIGH PROFILE Cushion QS1211CRecalls & reported problems

Roho, Inc. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613732061210), received between Jul 3, 2025 and Jul 3, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Pressure Sore/Ulcer1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 3, 2025Use of Device Problem; Use of Device Problem

    MR. (B)(6) RECEIVED A NEW ROHO CUSHION. HE HAS BEEN A ROHO USER FOR SEVERAL YEARS. UPON THE FIRST DAY OF USE, ONE QUADRANT OF THE CUSHION HAD DEFLATED WHEN HE WENT TO GET OFF THE CUSHION AT THE END OF THE DAY. REDNESS WAS IDENTIFIED FROM SITTING ON THE DEFLATED SECTION. WITHIN TWO DAYS, THE REDNESS HAD WORSENED TO A PRESSURE SORE INJURY, AND MR. (B)(6) HAD TO SEEK TREATMENT FROM THE WOUND CARE CLINIC. A NEW CUSHION WAS PROVIDED UNDER THE WARRANTY REPLACEMENT PROGRAM.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 22398782 ↗

Other Cushion, Flotation devices

All Cushion, Flotation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page