Adapta™ ADDR03 (UDI 00613994222060)Recalls & reported problems
Medtronic, Inc. · IPG, ADDR03, ADAPTA, DR, 3.2 MM, US · Class 3 · Implant
All models of Adapta (27)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Adapta ADDR03 is an implantable pacemaker pulse generator about the size of a small matchbox, placed under the skin to keep the heart beating at a steady rate. The 26 reports mostly describe infections at the implant site with fever and swelling, unit that moved out of position, leads stuck in the device, and units that could not be read or checked by the programmer. Reports mention 1 death and 15 injuries, with 10 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 1
- Injuries reported
- 15
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00613994222060), received between Feb 9, 2017 and Feb 3, 2026 — classified as 1 death, 15 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Pacing Problem” (3 reports).
16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Pacing Problem3
- Appropriate Device Problem Term/Code Not Available2
- Communication or Transmission Problem2
- Device Sensing Problem2
- Failure to Interrogate2
- Mechanical Problem2
- Premature Discharge of Battery2
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient6
- Unspecified Infection6
- Palpitations2
- Atrial Fibrillation1
- Bradycardia1
- Death1
- Device Overstimulation of Tissue1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
InjuryFeb 3, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION OF UNKNOWN ORIGIN. IT WAS ALSO NOTED THAT THE PATIENT EXPERIENCED A FEVER. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Fever; Unspecified Infection
MAUDE report 24247104 ↗MalfunctionSep 10, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) COULD NOT BE INTERROGATED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23014362 ↗InjuryJan 3, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED DUE TO UNSPECIFIED MALFUNCTION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21066002 ↗InjurySep 29, 2023Migration or Expulsion of Device; Misconnection; Migration or Expulsion of Device; Misconnection
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) MIGRATED. IT WAS ALSO REPORTED THAT BOTH THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WERE STUCK IN THE HEADER. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17841899 ↗InjuryApr 19, 2023Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NOTED TO BE PACING BELOW THE LOWER RATE. THE IPG WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 16768895 ↗InjuryJan 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 16214647 ↗InjuryDec 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT AN INFECTION OCCURRED. THE IMPLANTABLE PULSE GENERATOR SYSTEM WAS EXPLANTED AND REPLACED. NO FUR THER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. (B)(6) 2022 IT WAS FURTHER REPORTED BY THE PATIENT THAT THEY EXPERIENCED A WEEPING SWOLLEN WOUND POST EXPLANT AND HOSPITALIZATION FOR 9 DAYS WITH ANTIBIOTICS PRIOR TO IMPLANT OF REPLACEMENT IPG.
Patient problems: Unspecified Infection
MAUDE report 16012497 ↗MalfunctionDec 19, 2022High impedance; Impedance Problem; High impedance; Impedance Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A FALSE OUTLIER LEAD IMPEDANCE MEASUREMENT FOR THE RIGHT VENTRICULAR (RV) LEAD AND RIGHT ATRIAL (RA) LEAD. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16003818 ↗MalfunctionAug 19, 2022Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NO TELEMETRY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15256763 ↗MalfunctionJul 19, 2022Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A RACING HEART/FLUTTER. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS UNDERREPORTING ATRIAL FIBRILLATION EPISODES DUE TO POST MODE SWITCH OVERDRIVE PACING (PMOP) BEING PROGRAMMED ON AND COLLECTION DELAY BEING SET TO A NON-ZERO. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Atrial Fibrillation
MAUDE report 15050513 ↗InjuryApr 10, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT STATED THAT THEIR DEVICE " WAS NO LONGER WORKING" . THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14069167 ↗InjuryDec 14, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12998510 ↗InjuryDec 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND NOT REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 12954389 ↗InjuryNov 25, 2020Failure to Capture; Failure to Capture
IT WAS FURTHER REPORTED THAT NO MALFUNCTION WAS NOTED ON THE IPG.
Patient problems: Syncope/Fainting
MAUDE report 10905660 ↗MalfunctionFeb 12, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED BY THE PATIENT THAT THEY COULD NOT ESTABLISH A TELEMETRY SESSION FOR THE IMPLANTABLE PULSE GENERATOR (IPG) WITH THEIR MONITOR. THE PATIENT WAS REFERRED TO THEIR CLINIC. THE DEVICE AND REMOTE MONITOR REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9702688 ↗MalfunctionFeb 11, 2020Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) COULD NOT BE INTERROGATED. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9691654 ↗MalfunctionAug 9, 2019Signal Artifact/Noise; Electromagnetic Interference; Signal Artifact/Noise; Electromagnetic Interference
THE PATIENT REPORTED THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) RESET WHEN THEY WERE NEAR THE METAL DETECTOR IN A STORE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8877804 ↗MalfunctionFeb 18, 2019Pacing Problem; Pacing Problem
THE PATIENT REPORTED THEY "PUT OFF GETTING THEIR DEVICE CHECKED BECAUSE THE PAIN IS UNBEARABLE, I HAVE A GREAT DEAL OF DISCOMFORT EVERY TIME I HAVE MY DEVICE CHECK, I HAVE RESIDUAL DISCOMFORT FOR HOURS AFTER; IT FEELS BRUISED. THIS IS SOME OF THE WORST SENSATIONS I HAVE FELT IN MY LIFE". FOLLOW UP WITH THE PATIENT'S PHYSICIAN INDICATED THAT THE PATIENT HAS HIGH OUTPUT PACING WHEN A MAGNET IS APPLIED. THE PATIENT'S IMPLANTABLE PULSE GENERATOR REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Device Overstimulation of Tissue; Discomfort; Fatigue; Muscle Stimulation; Pain
MAUDE report 8348890 ↗InjuryJan 9, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED AN INFECTION OCCURRED. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 8231766 ↗InjuryJul 27, 2018Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED EROSION OCCURRED. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND TEMPORARY PACING WAS UTILIZED. A REPLACEMENT SYSTEM WAS LATER IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erosion; No Code Available
MAUDE report 7726249 ↗InjuryFeb 8, 2018Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED BY THE PATIENT'S FAMILY MEMBER THAT THE PATIENT'S LEAD HAD DETERIORATED AND THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) EXPERIENCED PREMATURE BATTERY DEPLETION. ADDITIONAL INFORMATION OBTAINED FOUND THAT THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED A STEADY RISE OF IMPEDANCE AND THRESHOLD TO HIGH LEVELS, RESULTING IN A LEAD POLARITY SWITCH. THE LEAD LATER EXPERIENCED INTERMITTENT CAPTURE. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). THE RIGHT VENTRICULAR (RV) LEAD WAS CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7254160 ↗InjuryJun 8, 2017Mechanical Problem; Capturing Problem; Device Sensing Problem; Mechanical Problem; Capturing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A PACING AND SENSING PROBLEM DUE TO A HEADER ISSUE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6627315 ↗InjuryJun 8, 2017Pacing Problem; Pacing Problem
IT WAS REPORTED THAT WITHIN ONE DAY POST-AV NODAL ABLATION, THE PATIENT EXPERIENCED PALPITATIONS AND A RACING HEART RATE. DURING A DEVICE CHECK, IT WAS NOTED THAT THE DEVICE WAS PACING ABOVE THE LOWER RATE EVEN THOUGH THE PATIENT WAS NOT ACTIVE. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Palpitations; Tachycardia
MAUDE report 6624598 ↗DeathApr 21, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT'S SPOUSE THAT THE PATIENT HAD PASSED AWAY. THE SPOUSE REPORTED THAT AFTER THE IMPLANTABLE PULSE GENERATOR (IPG) WAS IMPLANTED THE PATIENT "HAD AN INFECTION." THE SPOUSE ALSO MENTIONED THAT THE PATIENT HAD MANY OTHER PROBLEMS. FOLLOW UP WITH THE PATIENT'S CLINIC INDICATED THAT THEY HAD NO RECORDS REGARDING THE CIRCUMSTANCES OF THE PATIENT'S DEATH. THE CAUSE OF DEATH WAS NOT NOTED IN THEIR RECORDS. NO FURTHER INFORMATION WAS ABLE TO BE OBTAINED.
Patient problems: Death; Unspecified Infection
MAUDE report 6512708 ↗MalfunctionApr 6, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT REPORTED THAT THEY EXPERIENCED PALPITATIONS AND DID NOT THINK THEIR IMPLANTABLE PULSE GENERATOR (IPG) WAS "OKAY". THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Palpitations
MAUDE report 6470186 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
…and 1 more.
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page