Adapta™ ADSR01 (UDI 00613994222152)Recalls & reported problems
Medtronic, Inc. · IPG, ADSR01, ADAPTA, SR, IS-1, US · Class 3 · Implant
All models of Adapta (27)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Adapta pacemaker is an implanted device that sends electrical pulses to the heart. The 237 reports mostly describe batteries running down early, pacing problems and failures to read the device during checks, and some describe the device being replaced. The reports include 149 injuries and 8 deaths, and some mention slow heart rates or infections in patients. No recalls name this device, and reports fell to 1 in the last 12 months from 9 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 237
- Deaths reported
- 8
- Injuries reported
- 149
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 237 adverse event reports that list this exact device (DI 00613994222152), received between Feb 9, 2017 and Dec 22, 2025 — classified as 8 deaths, 149 injuries, 80 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (59 reports).
16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Discharge of Battery59
- Adverse Event Without Identified Device or Use Problem41
- Pacing Problem36
- Appropriate Device Problem Term/Code Not Available32
- Failure to Interrogate18
- Battery Problem17
- Communication or Transmission Problem15
- Failure to Capture14
Patient problems
- No Clinical Signs, Symptoms or Conditions74
- No Known Impact Or Consequence To Patient51
- Unspecified Infection34
- Bradycardia22
- No Consequences Or Impact To Patient9
- Death7
- Syncope/Fainting6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
Recent adverse event reports
The 25 most recent of 237 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED END OF LIFE (EOL) AND PACEMAKER MALFUNCTION. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23864469 ↗MalfunctionAug 18, 2025Pacing Problem; Unable to Obtain Readings; Battery Problem; Pacing Problem; Unable to Obtain Readings; Battery Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA WITH A HEART RATE IN THE FORTIES AND FIFTIES. THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED POSSIBLE PACING BELOW THE LOWER RATE AND INVALID DATA AS BATTERY AND LEAD MEASUREMENTS NOT AVAILABLE. THE DEVICE HAS REACHED RECOMMENDED REPLACEMENT TIME (RRT). IT WAS ALSO NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING BASED ON MARKERS ON NON-MAGNET ELECTROGRAMS (EGM) AND UNDER SENSED VENTRICULAR BEATS NOTED ON EPISODES. THE IPG AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 22833816 ↗MalfunctionAug 12, 2025Electromagnetic Interference; Impedance Problem; Electromagnetic Interference; Impedance Problem
IT WAS REPORTED THAT THERE WAS A LEAD WARNING FOR RIGHT VENTRICULAR (RV) LEAD IMPEDANCE WHICH WAS DURING CAUTERY TO THE PATIENT'S LEFT FOREARM. THE LEAD WAS FUNCTIONING AS EXPECTED WITH ALL MEASUREMENTS WITHIN RANGE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22780654 ↗InjuryAug 1, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED HEAT COMING FROM THEIR DEVICE POCKET AND THAT IT WAS ''WARM TO TOUCH'' AFTER BEING OUTSIDE ALL DAY. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22684675 ↗InjuryJul 28, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT HAD NO KNOWN PACEMAKER FOLLOW UP SINCE 2023, WHEN THE BATTERY ESTIMATE WAS NINETEEN MONTHS. THEY WERE NOW EXPERIENCING SYNCOPE AND NEAR SYNCOPE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Presyncope; Syncope/Fainting
MAUDE report 22628549 ↗InjuryMay 14, 2025Battery Problem; Battery Problem
IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22016518 ↗InjuryMar 18, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED EXCESSIVE BATTERY DEPLETION TRIGGERING ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO HIGH PACING OUTPUTS. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21637028 ↗MalfunctionFeb 4, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED RAPID BATTERY DEPLETION AS A RESULT OF CAPTURE MANAGEMENT UNEXPECTEDLY INCREASING THE PACING OUTPUTS. THE IPG STILL REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21303716 ↗MalfunctionJan 28, 2025Premature Discharge of Battery; Pacing Problem; Communication or Transmission Problem; Premature Discharge of Battery; Pacing Problem; Communication or Transmission Problem
IT WAS REPORTED THAT UPON ATTEMPTING TO CONNECT THE DEVICE TO THE PROGRAMMER THERE WAS CONCERN THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED BATTERY DEPLETION. PACING SPIKES WERE ALSO NOTED. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21252112 ↗MalfunctionDec 13, 2024Pacing Problem; Reset Problem; Pacing Problem; Reset Problem
IT WAS REPORTED THAT A PATIENT WITH A PACEMAKER EXPERIENCED HEART RATES IN THE THIRTIES AT HOME ACCOMPANIED BY LIGHTHEADEDNESS. AN EXTERNAL CARDIAC MONITOR INDICATED PAUSES, AND THERE WAS A POTENTIAL ISSUE OF OVERSENSING WITH THE DEVICE. THE DEVICE UNDERWENT AN ELECTRICAL RESET, AND A LEAD WARNING WAS NOTED WITH IMPEDANCE MEASURING LESS THAN ONE HUNDRED OHMS. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED AND RIGHT VENTRICULAR (RV) LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness; Non specific EKG/ECG Changes
MAUDE report 20924597 ↗InjuryOct 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A SYSTEMIC INFECTION. THE SOURCE OF THE INFECTION IS REPORTED TO BE THE LEG, WHICH WAS AMPUTATED. VEGETATION WAS ALSO NOTED ON THE TIP OF THE RIGHT VENTRICULAR (RV) LEAD. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20520420 ↗InjuryOct 4, 2024Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH PACING IMPEDANCE, SPIKE PACING IMPEDANCE, INTERMITTENT CAPTURE AND NOISE. RV LEAD FRACTURE WAS SUSPECTED. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG), WHICH WAS SCHEDULED FOR A CHANGE OUT DUE TO RECOMMENDED REPLACEMENT TIME, WAS LOCATED ABOVE THE CLAVICLE AND WAS NOT CAPTURING. IT WAS NOTED THAT THE PHYSICIAN WAS ABLE TO PALPATE AND FREELY MOVE THE IPG UNDER THE SKIN WHICH RESULTED IN INTERMITTENT CAPTURING. THE RV LEAD WAS CAPPED AND REPLACED. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20366943 ↗InjurySep 26, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED POTENTIAL PERFORMANCE ISSUES. THE IPG WAS EXPLANTED AND REPLACED. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20311495 ↗InjurySep 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) AND RIGHT VENTRICULAR (RV) LEAD WERE REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED, AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20117732 ↗InjuryAug 15, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA AND FATIGUE. IT WAS ALSO NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER LIMIT. THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Fatigue
MAUDE report 19989591 ↗InjuryAug 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED ENDOCARDITIS DUE TO AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND AN RV LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Endocarditis; Unspecified Infection
MAUDE report 19949275 ↗InjuryJun 24, 2024False Alarm; False Alarm
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A FALSE LOW IMPEDANCE WARNING. THE IPG WAS EXPLANTED AND REPLACED. THE RIGHT ATRIAL (RA) LEAD WAS CAPPED FOR AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19594306 ↗InjuryMay 1, 2024Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS UNABLE TO INTERROGATE. THE DEVICE WAS AT END OF LIFE (EOL). THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19219802 ↗MalfunctionApr 17, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) ''HAS BEEN CAUSING PROBLEMS''. THE IPG SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19123709 ↗InjuryApr 3, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. IT WAS ALSO NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NO PACING AT THE LOWER RATE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 19030680 ↗MalfunctionMar 12, 2024Failure to Capture; Over-Sensing; Failure to Capture; Over-Sensing
IT WAS REPORTED WHEN THE PHYSICIAN WAS OPENING THE DEVICE POCKET DURING A GENERATOR CHANGE AND THE PLASMA BLADE WAS APPLIED, THERE WAS OBSERVED LOSS OF PACING ON THE TELEMETRY. ONCE PLASMA BLADE WAS STOPPED, NORMAL PACING OCCURRED. TELEMETRY STRIPS WERE REVIEWED AND THERE WAS SOME ARTIFACT ON THE RECORDINGS BUT IT DID NOT APPEAR THERE WERE PACING SPIKES OR QRS EVENTS WHEN PLASMA BLADE WAS APPLIED. THE PATIENT WAS NOT SYMPTOMATIC AND THE GENERATOR CHANGE PROCEDURE WAS COMPLETED WITHOUT FURTHER COMPLICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18886604 ↗InjuryJan 19, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING THE GENERATOR CHANGE OUT THE PATIENT EXPERIENCED VENTRICULAR FIBRILLATION (VF) AND THE PATIENT WAS DEFIB RILLATED. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Ventricular Fibrillation
MAUDE report 18542895 ↗InjuryDec 13, 2023Failure to Capture; Failure to Interrogate; Reset Problem; Failure to Capture; Failure to Interrogate; Reset Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT (ED) WITH UNSPECIFIED SYMPTOMS. IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS POSSIBLY AT END OF SERVICE, AND EXHIBITED LOSS OF CAPTURE, AND A POWER-ON RESET HAD PREVIOUSLY OCCURRED. IT WAS ALSO REPORTED THAT THE IPG WAS UNABLE TO BE INTERROGATED IN THE ED USING THE PROGRAMMER. THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Fall; Insufficient Information
MAUDE report 18318077 ↗InjuryDec 6, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18274616 ↗InjuryNov 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED ITCHING FOR A YEAR AND HALF AT THE IMPLANTABLE PULSE GENERATOR (IPG) AFTER RECEIVING THE COVID VACCINE. THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Itching Sensation
MAUDE report 18200928 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
…and 1 more.
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page