Sprint Quattro Secure® 694758Recalls & reported problems

Medtronic, Inc. · LEAD 694758 US FPA QUAD · Class 3 · Implant

Quick summary

This Medtronic Sprint Quattro Secure is a wire lead that connects a pacemaker or defibrillator to the heart. Many reports describe the lead reading electrical signals incorrectly, high electrical resistance readings, and breaks in the lead, and some mention infection or shocks. The records include 1,932 reports, 1,249 injuries and 25 deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,937
Deaths reported
25
Injuries reported
1,252
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,937 adverse event reports that list this exact device (DI 00613994249876), received between Feb 11, 2013 and Sep 30, 2026 — classified as 25 deaths, 1,252 injuries, 660 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (707 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 26 reportsDec 2023: 18 reportsJan 2024: 14 reportsFeb 2024: 19 reportsMar 2024: 17 reportsApr 2024: 11 reportsMay 2024: 14 reportsJun 2024: 13 reportsJul 2024: 11 reportsAug 2024: 11 reportsSep 2024: 9 reportsOct 2024: 14 reportsNov 2024: 12 reportsDec 2024: 9 reportsJan 2025: 19 reportsFeb 2025: 10 reportsMar 2025: 7 reportsApr 2025: 5 reportsMay 2025: 12 reportsJun 2025: 17 reportsJul 2025: 15 reportsAug 2025: 7 reportsSep 2025: 14 reportsOct 2025: 11 reportsNov 2025: 11 reportsDec 2025: 13 reportsJan 2026: 12 reportsFeb 2026: 11 reportsMar 2026: 10 reportsApr 2026: 9 reportsMay 2026: 12 reportsJun 2026: 9 reportsJul 2026: 6 reportsAug 2026: 7 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202626
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing707
  • High impedance684
  • Fracture469
  • Signal Artifact/Noise389
  • Impedance Problem375
  • High Capture Threshold247
  • Adverse Event Without Identified Device or Use Problem179
  • Under-Sensing136

Patient problems

  • No Clinical Signs, Symptoms or Conditions806
  • No Known Impact Or Consequence To Patient671
  • Unspecified Infection141
  • Shock from Patient Lead(s)108
  • No Consequences Or Impact To Patient83
  • Syncope/Fainting21
  • Dizziness16
  • Erosion16

Recent adverse event reports

The 25 most recent of 1,937 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947271 ↗
  • InjurySep 28, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924096 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Erythema; Medical device site infection; Swelling/ Edema

    MAUDE report 50903247 ↗
  • InjurySep 18, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829460 ↗
  • MalfunctionSep 15, 2026Impedance Problem

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) stored episodes resulting in termination of AT/AF detected events in progress. An abrupt drop of P wave was noted during this session. It was further reported that there was an RV tip to RV coil impedance warning. The RA lead and the RV lead remain in use. No patient complications have been reported as a result of this event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766627 ↗
  • MalfunctionSep 14, 2026Over-Sensing; Decreased Sensitivity; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50762407 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed an infection. The Implantable Cardioverter Defibrillator (ICD) system was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50756625 ↗
  • InjurySep 1, 2026Fracture

    It was reported that the Right Ventricular (RV) lead had a suspected high voltage RV coil fracture. The RV lead was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50550180 ↗
  • InjuryAug 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to pocket infection. No further patient complications have been reported as a result of this event..

    Patient problems: Medical device site infection

    MAUDE report 50487566 ↗
  • MalfunctionAug 17, 2026Low impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50279215 ↗
  • MalfunctionAug 16, 2026High impedance; Impedance Problem

    It was reported that the Right Ventricular (RV) lead triggered an alert for elevated pacing impedance, which appeared to be a physiologic interaction with the RVR electrode as the impedance rise was slow and steady. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50265861 ↗
  • InjuryAug 13, 2026Decreased Sensitivity

    It was reported that the Right Ventricular (RV) lead exhibited diminished sensing. Reprogramming occurred and the RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50236654 ↗
  • MalfunctionAug 12, 2026High impedance; Capturing Problem; Impedance Problem

    Patient problems: Tachycardia

    MAUDE report 50226763 ↗
  • MalfunctionAug 10, 2026Impedance Problem

    It was reported that the Right Ventricular (RV) lead exhibited RV and Superior Vena Cava (SVC) defibrillation intermittent impedance drops. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50196366 ↗
  • InjuryAug 4, 2026Signal Artifact/Noise; Over-Sensing; High Capture Threshold

    It was reported that a Lead Integrity Alert (LIA) triggered on the Right Ventricular (RV) lead due to High Rate Non-Sustained episodes and short ventricular intervals. Capture thresholds were noted to be high and the lead was oversensing. Another LIA triggered the following day and noise was observed. A new spike in Non-Sustained Ventricular Tachycardia episodes was noted. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50126589 ↗
  • InjuryJul 31, 2026Capturing Problem

    It was reported that a patient with cardiomyopathy attributed to an apical infarct experienced rising right ventricular (RV) lead th resholds, and the patient was hospitalized. The patient reached Elective Replacement Indicator (ERI) at a normal interval, and the decision was made to perform a lead revision at that time. The RV lead was explanted and replaced. The patient outcome was reported as alive with no injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50099848 ↗
  • MalfunctionJul 29, 2026Over-Sensing; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50070674 ↗
  • MalfunctionJul 10, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25785589 ↗
  • InjuryJul 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN INFECTION WITH BACTEREMIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 25766212 ↗
  • MalfunctionJul 7, 2026Low impedance; Decreased Sensitivity; Capturing Problem; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OUT OF RANGE (OOR) LOW PACING IMPEDANCE, RISING THRESHOLD TREND AND HIGH RV THRESHOLD. THERE WAS ALSO A DECREASE IN SENSING OVER TIME. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25733997 ↗
  • InjuryJul 2, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. GROWTH WAS NOTED ON THE SUPERIOR VENA CAVA (SVC) COIL OF THE RIGHT VENTRICULAR (RV) LEAD. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection

    MAUDE report 25695206 ↗
  • MalfunctionJun 26, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25640977 ↗
  • MalfunctionJun 25, 2026High impedance; Unstable Impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A WARNING FOR HIGH OUT OF RANGE (OOR) PACING IMPEDANCE AND EXHIBITED VARYING DEFIBRILLATION AND SUPERIOR VENA CAVA (SVC) IMPEDANCE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25625948 ↗
  • MalfunctionJun 25, 2026High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A WARNING FOR RVTIP TO RVCOIL HIGH/UNDEFINED PACING IMPEDANCE. IT WAS ALSO REPORTED THAT THE RV LEAD EXHIBITED SPIKES IN BOTH RVCOIL AND SVCCOIL. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25625940 ↗
  • InjuryJun 24, 2026Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CROSS TALK WITH CROSS CHAMBER OVERSENSING AND OCCASIONAL PACING INHIBITION. THE PATIENT EXPERIENCED PRE-SYNCOPE, SYNCOPE, AND ASYSTOLE. THE RV LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Asystole; Presyncope; Syncope/Fainting

    MAUDE report 25611777 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →