NA 2067 (UDI 00613994362896)Recalls & reported problems
Medtronic, Inc. · RF Head 2067 10L 9FT Global · Class 3
Quick summary
This Medtronic RF head is a non-implanted component used with pacemaker, ICD, and CRT systems. The 36 reports mostly describe short circuits or out-of-specification performance that can shut down the programmer; there were no deaths or injuries. One recall, closed in January 2019, involved an error in calculating and displaying the programmer’s remaining device longevity. Reports fell from 6 in the previous 12 months to 2 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00613994362896), received between Apr 10, 2018 and Aug 30, 2026 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Unexpected Shutdown” (10 reports).
1 FDA recall record name this device's identifier.
10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 17, 2019Terminated
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
Z-0795-2019 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- Unexpected Shutdown10
- Material Integrity Problem7
- Communication or Transmission Problem5
- Electrical /Electronic Property Problem5
- Electrical Shorting5
- Corroded3
- Break2
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Patient Involvement11
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2026Electrical Shorting
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50513034 ↗MalfunctionApr 13, 2026Unexpected Shutdown
THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24854350 ↗MalfunctionAug 11, 2025Unexpected Shutdown; Unexpected Shutdown
THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22769019 ↗MalfunctionMay 20, 2025Electrical Shorting; Electrical Shorting
IT WAS REPORTED THAT THE RADIOFREQUENCY(RF) HEAD HAD A SHORT CIRCUIT THAT CAUSED THE PROGRAMMER TO SHUT DOWN. IT WAS NOTED THAT THE PROGRAMMER FUNCTIONED NORMALLY WHEN RF HEAD IS DISCONNECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22050661 ↗MalfunctionMay 7, 2025Connection Problem; Material Integrity Problem; Connection Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PROGRAMMER SHUT DOWN SPONTANEOUSLY AFTER CONNECTING THE RADIOFREQUENCY (RF) HEAD. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21974861 ↗MalfunctionMar 3, 2025Unexpected Shutdown; Unexpected Shutdown
THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21501989 ↗MalfunctionDec 2, 2024Electrical Shorting; Material Integrity Problem; Electrical Shorting; Material Integrity Problem
IT WAS REPORTED THAT WHENEVER THE RADIOFREQUENCY (RF) HEAD WAS SLIGHTLY MOVED, IT CAUSED A SHORT-CIRCUIT TO THE PROGRAMMER, RESULTED IN SUDDEN POWER OFF/SHUTDOWN. THE RF HEAD WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20818737 ↗MalfunctionNov 13, 2024Unexpected Shutdown; Unexpected Shutdown
IT WAS FURTHER REPORTED THAT THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20672562 ↗MalfunctionAug 14, 2024Unexpected Shutdown; Unexpected Shutdown
THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19979214 ↗MalfunctionApr 8, 2024Communication or Transmission Problem; Unexpected Shutdown; Communication or Transmission Problem; Unexpected Shutdown
IT WAS REPORTED THAT THE RADIOFREQUENCY (RF) HEAD WAS UNABLE TO INTERROGATE. THE RF HEAD HAS BEEN RETURNED FOR EVALUATION AND SUBSEQ UENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19058142 ↗MalfunctionMay 8, 2023Corroded; Use of Device Problem; Corroded; Use of Device Problem
IT WAS REPORTED THAT THE RADIOFREQUENCY (RF) HEAD WAS NOT WORKING. THE RF HEAD WAS RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16884411 ↗MalfunctionNov 24, 2022Electrical /Electronic Property Problem; Communication or Transmission Problem; Electrical /Electronic Property Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE RADIOFREQUENCY (RF) HEAD HAD AN INTERROGATION ISSUE. IT WAS FURTHER REPORTED THAT THE RF HEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15859950 ↗MalfunctionNov 15, 2022Break; Material Integrity Problem; Unexpected Shutdown; Break; Material Integrity Problem; Unexpected Shutdown
IT WAS REPORTED THAT THE CABLE INSULATION OF THE RADIOFREQUENCY (RF) HEAD WAS BROKEN. THE RF HEAD WAS RETURNED TO SERVICE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15796341 ↗MalfunctionSep 19, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE RADIOFREQUENCY HEAD WHICH WAS RETURNED AS AN ASSOCIATE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ASS ESSMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15444727 ↗MalfunctionJul 7, 2022Corroded; Corroded
IT WAS FURTHER REPORTED THAT THE RADIOFREQUENCY HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14943037 ↗MalfunctionMar 1, 2021Use of Device Problem; Communication or Transmission Problem; Use of Device Problem; Communication or Transmission Problem
IT WAS FURTHER REPORTED THROUGH FOLLOW UP THAT THE RADIOFREQUENCY PROGRAMMING HEAD WAS USED TO INTERROGATE A DEVICE PRIOR TO IMPLANT AND WOULD NOT ESTABLISH CONSISTENT COMMUNICATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391493 ↗MalfunctionFeb 2, 2021Electrical Shorting; Material Integrity Problem; Electrical Shorting; Material Integrity Problem
IT WAS REPORTED THAT THE RADIO FREQUENCY (RF) PROGRAMMER HEAD WIRES WERE BEGINNING TO SHORT. THE RF HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11264874 ↗MalfunctionJan 12, 2021Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT WHEN THE PROGRAMMER IS POWERED ON, THE SCREEN DOES NOT TURN ON. THE PROGRAMMER WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT. THE PROGRAMMER SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THE RADIOFREQUENCY PROGRAMMER HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11154402 ↗MalfunctionJan 8, 2021Unexpected Shutdown; Unexpected Shutdown
IT WAS REPORTED THAT WHEN THE PROGRAMMER HEAD CONNECTOR WAS MOVED, THE PROGRAMMER TURNED OFF. TROUBLESHOOTING STEPS WERE TAKEN TO INCLUDING CARRYING OUT SEVERAL TESTS WHICH GAVE THE SAME RESULTS. THERE WAS NO PATIENT INVOLVEMENT. IT WAS FURTHER REPORTED THAT THE RADIOFREQUENCY PROGRAMMER HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11138528 ↗MalfunctionDec 22, 2020Moisture Damage; Moisture Damage
IT WAS REPORTED THAT THE RADIOFREQUENCY PROGRAMMER HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11054203 ↗MalfunctionOct 23, 2020Material Integrity Problem; Material Integrity Problem
THE RADIOFREQUENCY HEAD (RF HEAD) RETURNED WITH THE PROGRAMMER AS A ASSOCIATED DEVICE TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10725916 ↗MalfunctionOct 23, 2020Material Integrity Problem; Material Integrity Problem
THE RADIO FREQUENCY (RF) PROGRAMMER HEAD WAS RETURNED AS AN ASSOCIATED DEVICE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURERS ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10725251 ↗MalfunctionOct 13, 2020Computer Operating System Problem; Computer Operating System Problem
IT WAS FURTHER REPORTED THAT THE RADIOFREQUENCY HEAD ORIGINALLY RETURNED AS AN ASSOCIATED DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10670160 ↗MalfunctionOct 8, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE PROGRAMMER BLACKED OUT, AND SHUT DOWN DURING A SOFTWARE UPDATE ALONG WITH A CRACKLING NOISE RESULTING IN THE PROGRAMMER BEING IMPOSSIBLE TO UPDATE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10648361 ↗MalfunctionSep 22, 2020Electrical Shorting; Electrical Shorting
IT WAS REPORTED THAT AFTER THE "PROGRAM CONTROL HEAD" (RADIOFREQUENCY HEAD) WAS PLUGGED IN, THE "PROGRAM CONTROLLER" (PROGRAMMER) AUTOMATICALLY POWERED OFF. IT WAS NOTED THAT THE ISSUE OCCURRED DURING SET UP. THE RADIOFREQUENCY HEAD HAS BEEN RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10567100 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-0399-2018 · Jul 14, 2017 · TerminatedMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Pacemaker/icd/crt non-implanted components devices
- Gallant™ CDDRA500QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components2 recalls33 deaths3,028 reports
- NA 229047Medtronic, Inc. · Pacemaker/icd/crt non-implanted components1 recall662 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls559 reports
- CareLink Encore™ 29901AMedtronic, Inc. · Pacemaker/icd/crt non-implanted components2 recalls474 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls422 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls151 reports
- Medtronic CareLink® Programmer 2090AA4Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls82 reports
- Entrant™ CDDRA300QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components1 recall2 deaths158 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.