Sprint Quattro Secure S™ 693575Recalls & reported problems
Medtronic, Inc. · LEAD 693575 US EN · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00613994446367), received between Aug 21, 2018 and Dec 10, 2025 — classified as 6 injuries, 5 malfunctions.
The most frequently reported specific device problem is “High impedance” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High impedance6
- Over-Sensing4
- Fracture3
- Signal Artifact/Noise3
- Under-Sensing3
- Impedance Problem2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Capture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Dyspnea1
- Shock from Patient Lead(s)1
- Unspecified Infection1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING DURING A TREATED VENTRICULAR FIBRILLATION (VF) EPISODE T HOUGH THERE WAS NO OBSERVED IMPACT TO OVERALL DEVICE FUNCTION OR PERFORMANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771840 ↗MalfunctionDec 10, 2025High impedance; Low impedance; High impedance; Low impedance
IT WAS REPORTED THAT AN ALERT TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD FOR ZERO OHM IMPEDANCE IN ONE CONFIGURATION WHILE ALERTING FOR HIGH UNDEFINED IMPEDANCE ON ANOTHER. THE RIGHT ATRIAL (RA) LEAD ALSO ALERTED FOR ZERO OHM IMPEDANCE. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771772 ↗MalfunctionJul 17, 2023High impedance; Impedance Problem; High impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH DEFIBRILLATION IMPEDANCE. IT WAS NOTED THAT THERE HAD BEEN A SLOW INCREASE IN RV LEAD DEFIBRILLATION IMPEDANCE OVER TIME WITH THE PATIENT'S CHANGING CLINICAL CONDITION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17331964 ↗InjuryMar 30, 2023Signal Artifact/Noise; High impedance; Over-Sensing; Signal Artifact/Noise; High impedance; Over-Sensing
IT WAS REPORTED THAT AN AUDIBLE ALERT WAS TRIGGERED. AN INTERROGATION SHOWED THAT A LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD FOR OVERSENSING OF NOISE ON THE RV CHANNEL AND NOISE DURING HIGH-RATE EVENTS, AND AN OUT OF RANGE (OOR) PACING IMPEDANCE MEASUREMENT WAS OBSERVED THAT WAS LATER CLARIFIED TO BE HIGH/UNDEFINED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16644614 ↗InjurySep 7, 2022Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING (TWOS) POST VENTRICULAR SENSE (VS) ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD RE MAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15371397 ↗InjuryNov 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. THE SYSTEM WAS LATER REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10809605 ↗MalfunctionOct 16, 2020Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Impedance Problem; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Impedance Problem
IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXTRACTED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10690202 ↗InjuryJun 30, 2020Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE PATIENT¿S RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR UNDEFINED HIGH RV COIL IMPEDANCE. A FRACTURE OF THE RV COIL WAS CONFIRMED. THE RV LEAD WAS CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10211876 ↗InjuryMar 19, 2020Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD UNDERSENSED ONE R-WAVE IN THE NON-SUSTAINED TACHYCARDIA EPISODE. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9853255 ↗MalfunctionFeb 6, 2019Under-Sensing; Under-Sensing
IT WAS REPORTED THAT RIGHT VENTRICULAR UNDERSENSING WAS NOTED ON STORED ELECTROGRAMS. THE RIGHT VENTRICULAR LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8312774 ↗InjuryAug 21, 2018Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold; Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold
IT WAS REPORTED THAT THE PATIENT WAS INAPPROPRIATELY SHOCKED BY THE RIGHT VENTRICULAR (RV) LEAD WHILE ASYMPTOMATIC AND WENT TO THE EMERGENCY ROOM (ER). THE RV LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO OVERSENSING OF NOISE WITH HIGH RATE NON-SUSTAINED (HR-NS) AND SENSING INTEGRITY COUNTER (SIC) EPISODES. THE LEAD HAD HIGH IMPEDANCE, HIGH THRESHOLDS WITH LOSS OF CAPTURE, AND A CONFIRMED FRACTURE. A MAGNET WAS PLACED OVER THE DEVICE TO PREVENT ANY FURTHER SHOCKS AND THE LEAD WAS PROGRAMMED OFF. THE LEAD REMAINS IN THE PATIENT OUT OF SERVICE. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAS HIGH THRESHOLDS. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea; Shock from Patient Lead(s)
MAUDE report 7801940 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page