Skip to content

MOSAIC® 305C223 (UDI 00613994479112)Recalls & reported problems

Medtronic, Inc. · VALVE 305C223 MOSAIC AOR CINCHII US · Class 3 · Implant

All models of MOSAIC (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00613994479112), received between Jul 29, 2019 and Dec 5, 2025 — classified as 9 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information6
  • Device Stenosis1
  • Gradient Increase1
  • Material Fragmentation1
  • Off-Label Use1
  • Patient Device Interaction Problem1
  • Perivalvular Leak1

Patient problems

  • Aortic Valve Stenosis5
  • Aortic Insufficiency2
  • Insufficient Information2
  • No Information2
  • Aortic Valve Insufficiency/ Regurgitation1
  • Dyspnea1
  • Endocarditis1
  • Pulmonary Valve Stenosis1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Device Stenosis; Insufficient Device Problem Information; Patient Device Interaction Problem; Device Stenosis

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS SEVERE OBSTRUCTIVE PULMONARY VALVE STENOSIS DUE TO ACTIVE PULMONARY VALVE ENDOCARDITIS. THE ORGANISM CULTURED WAS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Endocarditis; Insufficient Information; Pulmonary Valve Stenosis

    MAUDE report 23725234 ↗
  • InjuryJun 20, 2023Gradient Increase; Gradient Increase

    MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 1 MONTH POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 3 DAYS LATER, A 23MM TRANSCATHETER VALVE OF A DIFFERENT MANUFACTURER WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) ALSO DISCOVERED A PEAK GRADIENT OF 102MMHG, A MEAN GRADIENT OF 58MMHG, AN EJECTION FRACTION OF 55-60%, A PEAK VELOCITY OF 5.1M/S AND AN AORTIC VALVE AREA (AVA) OF 1.25CM2. THE PATIENT ALSO HAD SYMPTOMS OF SHORTNESS OF BREATH AND EXERTIONAL SYNCOPE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Dyspnea; Syncope/Fainting

    MAUDE report 17167152 ↗
  • InjuryJan 5, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 10 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. SUBSEQUENTLY 6 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 13171415 ↗
  • InjuryOct 27, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 1 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 12707401 ↗
  • InjuryMay 14, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 1 MONTH POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A 26MM TRANSCATHETER AORTIC VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 11825801 ↗
  • InjuryNov 6, 2019Material Fragmentation; Insufficient Device Problem Information; Material Fragmentation; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Insufficiency; No Information; Stroke/CVA; Thromboembolism

    MAUDE report 9287727 ↗
  • InjurySep 20, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM BIOPROSTHETIC AORTIC VALVE, IT WAS REPLACED VALVE-IN-VALVE WITH A 23MM TRANSCATHETER VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 9098057 ↗
  • InjurySep 12, 2019Off-Label Use; Off-Label Use

    MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 3 MONTHS POST IMPLANT OF THIS 23MM BIOPROSTHETIC VALVE IN THE PULMONARY POSITION, IT WAS REPLACED VALVE-IN-VALVE WITH A TRANSCATHETER BIOPROSTHETIC PULMONARY VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 9008472 ↗
  • InjuryJul 29, 2019Perivalvular Leak; Perivalvular Leak

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 4 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A 23MM TRANSCATHETER AORTIC VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASONS FOR REPLACEMENT WERE REPORTED AS AORTIC STENOSIS, MODERATE AORTIC INSUFFICIENCY, AND POSSIBLE PARAVALVULAR LEAK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Insufficiency; Aortic Valve Stenosis

    MAUDE report 8839453 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page