Capsure® EPI 4968-60Recalls & reported problems
Medtronic, Inc. · LEAD 4968-60 US BIPOLAR EPICARDIAL 60CM · Class 3 · Implant
All models of Capsure EPI (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Capsure EPI is an epicardial pacemaker lead, a wire attached to the outside of the heart. Many of its 454 reports describe the lead failing to send or read heartbeat signals, such as needing higher electrical settings, while others describe infections and fever that led to the lead being removed. The reports list 12 deaths, 234 injuries and 208 malfunctions, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 459
- Deaths reported
- 12
- Injuries reported
- 234
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 459 adverse event reports that list this exact device (DI 00613994492340), received between Dec 9, 2016 and Sep 17, 2026 — classified as 12 deaths, 234 injuries, 213 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (121 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Capture Threshold121
- Adverse Event Without Identified Device or Use Problem110
- Over-Sensing90
- Failure to Capture70
- Under-Sensing52
- High impedance50
- Capturing Problem29
- Fracture22
Patient problems
- No Clinical Signs, Symptoms or Conditions218
- No Known Impact Or Consequence To Patient70
- Unspecified Infection60
- No Consequences Or Impact To Patient18
- Bradycardia11
- Dyspnea11
- Discomfort8
- Dizziness8
Recent adverse event reports
The 25 most recent of 459 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Failure to Capture
It was reported that the Right Atrial (RA) leads exhibited loss of capture. Both RA leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803483 ↗MalfunctionSep 17, 2026Failure to Capture
It was reported that the Right Atrial (RA) leads exhibited loss of capture. Both RA leads remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803481 ↗MalfunctionSep 14, 2026High impedance
It was reported that the Right Ventricular (RV) lead exhibited high impedance during a shoulder procedure. The RV lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50751276 ↗MalfunctionSep 7, 2026Over-Sensing; High Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50646950 ↗MalfunctionSep 3, 2026Under-Sensing
It was reported that the Right Ventricular (RV) lead exhibited under sensing during ventricular arrhythmia. The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50586123 ↗MalfunctionAug 28, 2026Failure to Capture; High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50499899 ↗InjuryAug 27, 2026High Capture Threshold
Patient problems: Bradycardia
MAUDE report 50474576 ↗InjuryAug 27, 2026High Capture Threshold
It was reported that both the Right Ventricular (RV) epicardial leads exhibited high thresholds and the patient was bradycardic. Both leads were inactivated and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Bradycardia
MAUDE report 50473850 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced an infection and fever. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) and the Right Atrial (RA) lead were explanted. The epicardial Left Ventricular (LV) lead the epicardial Right Ventricular (RV) lead were partially explanted. No further patient complications have been reported as a result of this event.
Patient problems: Fever; Unspecified Infection
MAUDE report 50436183 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced an infection and fever. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) and the Right Atrial (RA) lead were explanted. The epicardial Left Ventricular (LV) lead the epicardial Right Ventricular (RV) lead were partially explanted. No further patient complications have been reported as a result of this event.
Patient problems: Fever; Unspecified Infection
MAUDE report 50436099 ↗MalfunctionAug 14, 2026Over-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50249953 ↗InjuryAug 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported by the patient that they experienced ongoing concerns related to Cardiac Resynchronization Therapy Defibrillator (CRT-D) follow-up and cardiac evaluation. The patient reported shortness of breath, dizziness, left-sided pain, leg edema, and difficulty walking, and stated that they had been told they had an obstruction and needed additional workup, including blood testing and an echocardiogram with contrast. The patient also reported that their CareLink monitor was not online and that there were delays and difficulties with follow-up care and referral coordination. The patient expressed frustration that the recommended evaluation and testing had not yet been completed. The CRT-D system remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Ambulation or Postural Difficulties; Dizziness; Dyspnea; Obstruction/Occlusion; Pain; Peripheral Edema
MAUDE report 50195224 ↗InjuryJul 30, 2026Failure to Capture; High impedance; Decreased Sensitivity; Capturing Problem; High Capture Threshold
It was reported that the patient experienced dizziness and bradycardia. It was also reported that the Right Ventricular (RV) lead exhibited loss of capture, rising thresholds, high thresholds, diminished sensing and triggered an alert for Out of Range (OOR) high impedance. The lead was capped and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Bradycardia; Dizziness
MAUDE report 50079752 ↗InjuryJul 29, 2026Fracture
It was reported that during implant the Right Ventricular (RV) lead pin was fractured. The lead was attempted not used and replaced. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50066943 ↗MalfunctionJul 9, 2026Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING AND OVERSENSING AND THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND HAD AN IMPEDANCE WARNING. THE RA AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25765693 ↗MalfunctionJun 29, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND AN ATRIAL LEAD POSITION CHECK FAILURE. IT WAS ALSO REP ORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CROSS CHAMBER OVERSENSING AND A HIGH NUMBER OF SHORT VENTRICLE TO VENTRICLE (V-V) INTERVALS. THE RA LEAD AND THE RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25649252 ↗MalfunctionJun 29, 2026Electrical /Electronic Property Problem; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND AN ATRIAL LEAD POSITION CHECK FAILURE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CROSS CHAMBER OVERSENSING AND A HIGH NUMBER OF SHORT VENTRICLE TO VENTRICLE (V-V) INTERVALS. THE RA LEAD AND THE RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25649246 ↗InjuryJun 16, 2026Failure to Capture
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING, INHIBITION OF PACING AND ONE BEAT OF FAILURE TO CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS KNOWN NOT TO HAVE CAPTURE AND WAS PROGRAMMED OFF. IT WAS ALSO NOTED THAT THE RV LEAD IMPEDANCE TRENDS SHOW SLIGHT VARIABILITY IN BIPOLAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25517728 ↗MalfunctionJun 11, 2026Failure to Capture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES AND TERMINATION OF AT/AF DETECTED EVENTS IN PROGRESS. THE RIGHT ATRIAL (RA) LEAD ALSO EXHIBITED NO CAPTURE. THE LEFT VENTRICULAR (LV) LEAD EXHIBTED NO CAPTURE. THE LEADS REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25470705 ↗MalfunctionJun 3, 2026Low impedance; High Capture Threshold
IT WAS REPORTED APPROXIMATELY THREE MONTHS POST IMPLANT THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND FAR FIELD R-WAVE (FFRW) OVERSENSING WHILE THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND LOW IMPEDANCE. THE RA AND RV LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25372962 ↗MalfunctionJun 3, 2026Over-Sensing; High Capture Threshold
IT WAS REPORTED APPROXIMATELY THREE MONTHS POST IMPLANT THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS AND FAR FIELD R-WAVE (FFRW) OVERSENSING WHILE THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND LOW IMPEDANCE. THE RA AND RV LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25372961 ↗InjuryMay 25, 2026Failure to Capture
IT WAS REPORTED THAT ONE DAY POST IMPLANT THE PATIENT EXPERIENCED BLEEDING. IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) CAPTURE MANAGEMENT WAS ABORTED DUE TO HIGH CAPTURE THRESHOLDS. LOSS OF CAPTURE WAS OBSERVED ON THE RV AND LEFT VENTRICULAR (LV) LEAD. UNDERSENSING WAS NOTED ON THE RV LEAD WHICH RESULTED IN INAPPROPRIATE PACING. THE PATIENT WAS TAKEN TO THE OPERATING ROOM AND BOTH LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 25268925 ↗InjuryMay 25, 2026Failure to Capture; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT ONE DAY POST IMPLANT THE PATIENT EXPERIENCED BLEEDING. IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) CAPTURE MANAGEMENT WAS ABORTED DUE TO HIGH CAPTURE THRESHOLDS. LOSS OF CAPTURE WAS OBSERVED ON THE RV AND LEFT VENTRICULAR (LV) LEAD. UNDERSENSING WAS NOTED ON THE RV LEAD WHICH RESULTED IN INAPPROPRIATE PACING. THE PATIENT WAS TAKEN TO THE OPERATING ROOM AND BOTH LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268906 ↗InjuryMay 21, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). REPROGRAMMING WAS PERFORMED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25246626 ↗MalfunctionMay 18, 2026High Capture Threshold
IT WAS REPORTED THAT THERE HIGH THRESHOLD ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25201995 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026