Capsure® EPI 4965-15 (UDI 00613994501721)Recalls & reported problems

Medtronic, Inc. · LEAD 4965-15 US UNI EPI STEROID 15CM · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Capsure EPI pacemaker lead is a wire that connects a pacemaker to the heart. Of its 128 reports, 90 are injuries, 37 are malfunctions and 1 describes a death; many describe electrical readings out of range, and others describe the lead fracturing or failing to pace the heart. Reports fell to 3 in the last 12 months from 22 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
129
Deaths reported
1
Injuries reported
91
Recalls naming it
0

Summary

The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 00613994501721), received between Mar 16, 2017 and Jul 30, 2026 — classified as 1 death, 91 injuries, 37 malfunctions.

The most frequently reported specific device problem is “Fracture” (50 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture50
  • Low impedance35
  • High impedance34
  • High Capture Threshold23
  • Failure to Capture21
  • Over-Sensing17
  • Break10
  • Under-Sensing10

Patient problems

  • No Clinical Signs, Symptoms or Conditions74
  • No Known Impact Or Consequence To Patient34
  • Unspecified Infection6
  • No Consequences Or Impact To Patient5
  • Bradycardia4
  • Device Overstimulation of Tissue2
  • Syncope2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 30, 2026Fracture

    It was noted that the RA lead did not exhibit a fracture.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50083254 ↗
  • MalfunctionJan 21, 2026Over-Sensing

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A FALSE HIGH THRESHOLD MEASUREMENT AND RVCM NOT RUNNING. IT WAS FURTHER NOTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OCCASIONAL FAR FIELD R-WAVE (FFRW) OVERSENSING. THE IPG AND RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24131907 ↗
  • MalfunctionNov 11, 2025Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23520432 ↗
  • InjuryNov 10, 2025Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem

    THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23510579 ↗
  • InjuryAug 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A LEAD FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22858387 ↗
  • MalfunctionAug 13, 2025Low impedance; Impedance Problem; Low impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED OUT OF RANGE UNIPOLAR IMPEDANCE. BOTH THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22791686 ↗
  • MalfunctionAug 13, 2025Low impedance; Impedance Problem; Low impedance; Impedance Problem

    IT WAS FURTHER REPORTED THAT THE RA AND THE RV LEAD IMPEDANCES WERE LOW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22791673 ↗
  • InjuryAug 5, 2025Failure to Capture; Fracture; High impedance; Pacing Problem; Under-Sensing; High Capture Threshold; Failure to Capture; Fracture; High impedance; Pacing Problem; Under-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED SUSPECTED FRACTURE, HIGH UNDEFINED IMPEDANCE, HIGH THRESHOLD, NO CAPTURE AT HIGH OUTPUT, INAPPROPRIATE PACING AND NOT SENSING. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22714345 ↗
  • InjuryJul 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS NOTED TO BE FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22662654 ↗
  • InjuryJul 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A LOW VOLTAGE FRACTURE. IT WAS NOTED THE RV LEAD EXHIBITED A UNIPOLAR LEAD IMPEDANCE WARNING ALSO. THE RV LEAD WAS PARTIALLY EXTRACTED, CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22565061 ↗
  • InjuryJul 15, 2025Failure to Capture; Fracture; High impedance; Failure to Capture; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDEFINED HIGH IMPEDANCE, NO CAPTURE AND THE LEAD WAS NOTED TO BE FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22507468 ↗
  • MalfunctionJul 10, 2025Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OUT OF RANGE LOW IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22466743 ↗
  • InjuryJun 12, 2025Fracture; High impedance; Capturing Problem; Fracture; High impedance; Capturing Problem

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A CAPTURE ISSUE, UNDEFINED HIGH IMPEDANCE AND A LOW VOLTAGE LEAD FRACTURE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A CAPTURE ISSUE, OVERSENSING, UNDEFINED HIGH IMPEDANCE AND A LOW VOLTAGE LEAD FRACTURE. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22192512 ↗
  • InjuryJun 12, 2025Fracture; High impedance; Over-Sensing; Capturing Problem; Fracture; High impedance; Over-Sensing; Capturing Problem

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A CAPTURE ISSUE, UNDEFINED HIGH IMPEDANCE AND A LOW VOLTAGE LEAD FRACTURE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A CAPTURE ISSUE, OVERSENSING, UNDEFINED HIGH IMPEDANCE AND A LOW VOLTAGE LEAD FRACTURE. THE LEADS WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22192510 ↗
  • InjuryJun 4, 2025Failure to Capture; Fracture; High impedance; High Capture Threshold; Failure to Capture; Fracture; High impedance; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FRACTURED AND EXHIBITED SPIKED AND HIGH IMPEDANCE AND NO CAPTURE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FRACTURED AN EXHIBITED HIGH IMPEDANCE, HIGH THRESHOLDS AND NO CAPTURE. IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION. THE IPG SYSTEM WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22141158 ↗
  • InjuryJun 4, 2025Failure to Capture; Fracture; High impedance; Impedance Problem; Failure to Capture; Fracture; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FRACTURED AND EXHIBITED SPIKED AND HIGH IMPEDANCE AND NO CAPTURE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FRACTURED AN EXHIBITED HIGH IMPEDANCE, HIGH THRESHOLDS AND NO CAPTURE. IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION. THE IPG SYSTEM WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22141126 ↗
  • InjuryMay 30, 2025Intermittent Capture; High impedance; Over-Sensing; Intermittent Capture; High impedance; Over-Sensing

    IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED.

    Patient problems: Heart Block; Unspecified Infection; Wound Dehiscence

    MAUDE report 22113423 ↗
  • InjuryMay 28, 2025Intermittent Capture; Failure to Capture; Device Dislodged or Dislocated; High Capture Threshold; Intermittent Capture; Failure to Capture; Device Dislodged or Dislocated; High Capture Threshold

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. IT WAS ALSO REPORTED THAT THE RV LEAD EXHIBITED INTERMITTENT RV CAPTURE, LOSS OF RV CAPTURE AND ELEVATED CAPTURE THRESHOLDS. THE RV LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Asystole; Syncope/Fainting

    MAUDE report 22096463 ↗
  • InjuryMay 22, 2025Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INSULATION BREACH. THE RV LEAD WAS PARTIALLY REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22068365 ↗
  • MalfunctionApr 3, 2025Break; Break

    THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21762785 ↗
  • MalfunctionApr 3, 2025Break; Break

    THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21762758 ↗
  • InjuryMar 31, 2025Intermittent Capture; Failure to Capture; Under-Sensing; Low impedance; Intermittent Capture; Failure to Capture; Under-Sensing; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDEFINED LOW IMPEDANCE, UNDERSENSING, INTERMITTENT AND LOSS OF CAPTURE. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21730933 ↗
  • InjuryMar 25, 2025Failure to Capture; Low impedance; Failure to Capture; Low impedance

    IT WAS REPORTED POST IMPLANT THAT THERE WAS A RIGHT VENTRICULAR (RV) LEAD IMPEDANCE ISSUE WITH A READING OF 3401. IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A LOW IMPEDANCE WARING. IT WAS ALSO REPORTED THAT THERE WAS A POCKET REVISION ON THE DAY OF IMPLANT DUE TO LOSS OF CAPTURE. THE RV LEAD AND THE RA LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687480 ↗
  • InjuryMar 25, 2025Failure to Capture; High impedance; Failure to Capture; High impedance

    IT WAS REPORTED POST IMPLANT THAT THERE WAS A RIGHT VENTRICULAR (RV) LEAD IMPEDANCE ISSUE WITH A READING OF 3401. IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A LOW IMPEDANCE WARING. IT WAS ALSO REPORTED THAT THERE WAS A POCKET REVISION ON THE DAY OF IMPLANT DUE TO LOSS OF CAPTURE. THE RV LEAD AND THE RA LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687469 ↗
  • MalfunctionFeb 26, 2025Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POSSIBLE CAPTURE ISSUE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE. BOTH THE RA AND RV LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21468707 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →