Skip to content

CD HORIZON® Spinal System 54840006540Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 54840006540 MAS 6.5X40 CC · Class 2 · Implant

All models of CD HORIZON Spinal System (341)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00613994579621), received between Sep 25, 2017 and Mar 9, 2022 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K091974); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break5
  • Failure to Align1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2022Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT POSTERIOR LUMBAR FUSION (PLF) FOR STENOSIS. IT WAS REPORTED THAT THE SCREW WAS BROKEN. THERE WERE NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT FIRST THREAD ON TULIP HEAD CHIPPED OFF AND NO BROKEN FRAGMENT WAS LEFT INSIDE THE PATIENT. THERE WAS NO REVISION SURGERY/ IN- PATIENT OR PROLONGATION OF EXISTING HOSPITALIZATION/ ADDITIONAL SURGERY/ DELAY IN OVERALL PROCEDURE TIME. THE IMPLANT AND EXPLANT DATE OF THE SCREW IS (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13710903 ↗
  • InjuryFeb 22, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LUMBAR FUSION DUE TO BACK/LEG PAIN. ON AN UNKNOWN DATE, POST-OP, THE IMPLANTED PEDICLE SCREW BROKE. HENCE, A REVISION SURGERY WAS PLANNED TO BE PERFORMED ON (B)(6) 2020; HOWEVER, IT IS UNKNOWN IF THE SURGERY HAS BEEN PERFORMED YET OR NOT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9742707 ↗
  • InjuryFeb 22, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LUMBAR FUSION DUE TO BACK/LEG PAIN. ON AN UNKNOWN DATE, POST-OP, THE IMPLANTED PEDICLE SCREW BROKE. HENCE, A REVISION SURGERY WAS PLANNED TO BE PERFORMED ON (B)(6) 2020; HOWEVER, IT IS UNKNOWN IF THE SURGERY HAS BEEN PERFORMED YET OR NOT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9742705 ↗
  • InjuryJan 17, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SPINAL FUSION AT L5-S1 IN 2015. POST-OP, THE SCREW BROKE. A REVISION SURGERY WAS PERFORMED TO REMOVE THE BROKEN SCREWS AT S1, REFUSION AT L5-S1 AND FIXATION L4-5 ,S1-2. FRAGMENTS OF THE BROKEN PRODUCT REMAINED INSIDE THE PATIENT. PATIENT COMPLICATION REPORTED TO BE UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8256261 ↗
  • InjuryJan 9, 2018Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT TRANSFORAMINAL LUMBAR INTERBODY FUSION. REPORTEDLY POST-OP, S1 SCREW BROKE BELOW TULIP. THE PATIENT UNDERWENT REVISION SURGERY FOR REMOVAL OF HARDWARE AND RE-INSTRUMENTATION. THE SURGEON PERFORMED AN ANTERIOR LUMBAR INTERBODY FUSION AND A REVISION POSTERIOR FUSION. NO FRAGMENTS OF THE DEVICES REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT ACHIEVE SOLID FUSION. THE PATIENT TESTED POSITIVE FOR GRAM NEGATIVE BACILLI.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 7174407 ↗
  • MalfunctionSep 25, 2017Failure to Align; Failure to Align

    IT WAS REPORTED THAT ON (B)(6) 2017, PATIENT UNDERWENT LAMINECTOMY SURGERY AT LEVELS L5-S1. INTRA-OP, SCREW THREAD WAS DAMAGED WHEN SURGEON WAS TRYING TO FIX THE SET SCREW. THIS PROCESS WAS TRIED 3 TIMES. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO FRAGMENT OF THE IMPLANT REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6889487 ↗

FDA clearance and classification

Cleared through Special 510(k) K091974 on Sep 2, 2009 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek USA, Inc. as “Modification To Cd Horizon Spinal System”. The FDA review took 63 days. FDA record ↗

The FDA regulates product code MNH (orthosis, spondylolisthesis spinal fixation) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K091974. One submission often covers many models, so check the product and lot details in each record.

  • Z-2876-2011 · Jun 28, 2011 · Terminated
    CD HORIZON (R) SOLERA (TM) FAS, REF 54410006035, pedicle screw, Rx only, Medtronic Sofamor Danek USA, 1800 Pyramid Place, Memphis, TN 38132 Intended to help provide immobilizatio

Common questions about the CD HORIZON® Spinal System 54840006540

Has the CD HORIZON® Spinal System 54840006540 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00613994579621), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the CD HORIZON® Spinal System 54840006540?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 25, 2017 and Mar 9, 2022. In those reports, the most common specific device problems are “Break” (5) and “Failure to Align” (1); the most common patient problems are “Failure of Implant” (1) and “Unspecified Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CD HORIZON® Spinal System 54840006540?

None of the reports are classified as deaths, 4 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CD HORIZON® Spinal System 54840006540 FDA approved?

It was cleared through Special 510(k) K091974 on Sep 2, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the CD HORIZON® Spinal System 54840006540 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the CD HORIZON® Spinal System 54840006540, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand CD HORIZON® Spinal System, model 54840006540). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page