Revo MRI™ SureScan™ RVDR01 (UDI 00613994614742)Recalls & reported problems
Medtronic, Inc. · IPG RVDR01 REVO MRI SURESCAN US MKT IS-1 · Class 3 · Implant
All models of Revo MRI SureScan (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Revo MRI SureScan implanted pulse generator is a permanent implant. Its 721 reports, including 4 deaths and 445 injuries, mostly describe unexpected resets, batteries running down early, and pacing problems. Recent injury reports mention fainting when the generator reached end of service, and one describes a generator bulging out and needing replacement. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 721
- Deaths reported
- 4
- Injuries reported
- 445
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 721 adverse event reports that list this exact device (DI 00613994614742), received between Feb 9, 2013 and Aug 11, 2026 — classified as 4 deaths, 445 injuries, 272 malfunctions.
The most frequently reported specific device problem is “Reset Problem” (142 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Reset Problem142
- Adverse Event Without Identified Device or Use Problem111
- Premature Discharge of Battery87
- Appropriate Device Problem Term/Code Not Available86
- Pacing Problem82
- Inappropriate or Unexpected Reset69
- Battery Problem51
- Unable to Obtain Readings48
Patient problems
- No Known Impact Or Consequence To Patient265
- No Clinical Signs, Symptoms or Conditions157
- Unspecified Infection64
- No Consequences Or Impact To Patient42
- Bradycardia26
- Dizziness23
- Syncope/Fainting20
- Dyspnea19
Recent adverse event reports
The 25 most recent of 721 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50208838 ↗MalfunctionJul 10, 2026Failure to Interrogate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25778440 ↗MalfunctionMay 29, 2026Pacing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDER SENSING DURING AT/AF STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS AND MISCLASSIFICATION OF EPISODE. IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED SPIKES SEEN WITHIN THE QRS COMPLEX ON TELEMETRY. THE SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25329213 ↗InjuryMar 18, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT HAD A SYNCOPAL EPISODE WHEN THEIR IMPLANTABLE PULSE GENERATOR (IPG) HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED DUE TO NORMAL END OF SERVICE (EOS). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 24625547 ↗InjuryJan 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) WAS "BULGES OUT", THUS REQUIRING IT TO BE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24188177 ↗InjuryNov 26, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS AT END OF SERVICE (EOS). THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 23656006 ↗InjurySep 8, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. THE IMPLANTABLE PULSE GENERATOR (IPG) REACHED END OF SERVICE (EOS);AND WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 22992362 ↗MalfunctionAug 27, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THERE WAS LOSS OF CAPTURE NOTED ON EXTERNAL TELEMETRY, WITH PATIENT HEART RATE INTO THE FORTIES. IT WAS FOUND THAT THE IMPLANTABLE PULSE GENERATOR (IPG) HAD REACHED END OF SERVICE (EOS). IN ADDITION, THE RIGHT ATRIAL (RA) LEAD EXHIBITED A VARIATION AND AN INCREASE IN LEAD IMPEDANCES NOTED ON LEAD TRENDS AND ATRIAL CAPTURE COULD NOT BE CONFIRMED. THE IPG WAS EXPLANTED AND REPLACED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI). THE RA LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 22903361 ↗InjuryAug 18, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22826547 ↗InjuryMay 30, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE IPG HAD REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND WAS EXPLANTED AND REPLACED WITH A CARDIAC RE SYNCHRONIZATION THERAPY PACEMAKER (CRT-P) AS AN UPGRADE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22112648 ↗MalfunctionMar 28, 2025Inappropriate or Unexpected Reset; Reset Problem; Inappropriate or Unexpected Reset; Reset Problem
IT WAS REPORTED THAT DURING A DEVICE CHECK A MESSAGE CAME UP ON THE "POP-UP WINDOW" BUT THE HEALTHCARE PROFESSIONAL HIT CLEAR AND DID NOT READ EVERYTHING IN THE POP-UP WINDOW. IT WAS REPORTED THAT THERE WAS POTENTIAL FOR A DEVICE RESET. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21718056 ↗InjuryJan 30, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED NON-PROPHYLACTICALLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21272146 ↗InjuryJan 23, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SHORTNESS OF BREATH AND LIGHTHEADEDNESS. PATIENT STATED THAT THERE WAS A DELAY WITH THE IMPLANTABLE PULSE GENERATOR (IPG) TO RESPOND WHILE THEY DO EXERCISE. THE PATIENT WAS NOT HAPPY WITH THE DEVICE. THE IPG WAS EXPLANTED AND REPLACED DUE TO REACHING ELECTIVE REPLACEMENT INDICATOR (ERI). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness; Dyspnea
MAUDE report 21216509 ↗InjuryJan 20, 2025Reset Problem; Reset Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A POWER ON RESET (POR). IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUTPUTS. THE IPG AND RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21189959 ↗InjuryJan 20, 2025Reset Problem; Reset Problem
IT WAS REPORTED THAT AN ELECTRICAL RESET OCCURRED ON THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS NOTED THAT THE IPG WAS AT END OF SERVICE (EOS). THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21189888 ↗InjuryJul 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE PARTIALLY REMOVED, WITH THE REMAINING PART BEING CAPPED. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 19748108 ↗InjuryJul 10, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED CHEST PAIN, VERY BAD HEADACHES, LOW BLOOD PRESSURE, EAR PAIN AND THEY FAINTED ON THEIR FLOOR. THE PATIENT NOTED THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NOT WORKING AND WAS DEAD. THE IPG WAS REMOVED AND REPLACED. FOLLOWING THE IMPLANTATION OF THE REPLACEMENT IPG THE PATIENT EXPERIENCED BLACKING OUT REPEATEDLY AND REDNESS AT THE INCISION SITE. THE PATIENT ALSO NOTED THEIR PACING LEAD WAS NOT WORKING FULLY. THE REPLACEMENT IPG AND THE PACING LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Chest Pain; Ear Pain; Headache; Syncope/Fainting
MAUDE report 19711688 ↗InjuryJul 3, 2024Failure to Interrogate; Pacing Problem; Failure to Interrogate; Pacing Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER RATE AND WAS UNABLE TO BE INTERROGATED. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA, LIGHTHEADEDNESS AND FATIGUE. THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19664361 ↗InjuryMay 20, 2024Battery Problem; Appropriate Device Problem Term/Code Not Available; Battery Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEY ENDED UP IN HOSPITAL FOR ATRIAL FIBRILLATION (AF). THE IPG WAS EXPLANTED AND REPLACED. THE PATIENT ALSO EXPERIENCED ATRIAL FIBRILLATION (AF) WITH THEIR NEW IPG. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Atrial Fibrillation
MAUDE report 19350430 ↗MalfunctionMay 8, 2024Reset Problem; Reset Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED BIT FLIP RESULTING IN FULL RESET. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19268351 ↗InjuryMay 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. IT WAS ALSO NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS AT END OF SERVICE (EOS). THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 19268314 ↗InjuryApr 18, 2024Battery Problem; Battery Problem
THE PATIENT REPORTED THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) BATTERY WAS DEAD. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19133856 ↗MalfunctionFeb 1, 2024Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING. IT WAS REPORTED THAT THE OVERSENSING ON THE RA LEAD AND THE RV LEAD APPEARED TO BE NON-PHYSIOLOGIC. IT WAS FURTHER REPORTED THAT POSSIBLE ELECTROMAGNETIC INTERFERENCE WAS NOTED ON THE IMPLANTABLE PULSE GENERATOR (IPG) ON EPISODE ELECTROGRAMS (EGM). THIS COINCIDED WITH THE PATIENT RECEIVING AN ORTHOPEDIC SURGICAL PROCEDURE. THE SUBSEQUENT AVAILABLE ELECTROGRAMS (EGM) SHOW APPROPRIATE SENSING. THE IPG SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18622040 ↗MalfunctionJan 30, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT HAD THEIR HEART RATE IN THIRTIES VIA APPLE WATCH. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE PROGRAMMED LOWER RATE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 18607256 ↗InjuryJan 19, 2024Pacing Problem; Battery Problem; Pacing Problem; Battery Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PALPITATIONS, CHEST DISCOMFORT AND ARRHYTHMIA. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI) ALONG WITH SUSPECTED HIGH AND INAPPROPRIATE PACING. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Arrhythmia; Discomfort
MAUDE report 18541884 ↗
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page