Sprint Quattro Secure® DXAC/DSP 694758 (UDI 00613994740915)Recalls & reported problems
Medtronic, Inc. · LEAD 694758 SPRINT US 58CM FPA QUAD · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · SPRINT QUATTRO SECURE DXAC/DSP (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Sprint Quattro Secure lead is a wire that connects a permanent pacemaker to the heart. Its 419 reports often describe the lead misreading heartbeats, high electrical resistance, suspected breaks in the wire, and noise. The reports include 3 deaths and 278 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 421
- Deaths reported
- 3
- Injuries reported
- 278
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 421 adverse event reports that list this exact device (DI 00613994740915), received between Dec 9, 2016 and Sep 9, 2026 — classified as 3 deaths, 278 injuries, 140 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (165 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing165
- High impedance136
- Fracture86
- Signal Artifact/Noise79
- Impedance Problem64
- Adverse Event Without Identified Device or Use Problem52
- Under-Sensing46
- High Capture Threshold32
Patient problems
- No Clinical Signs, Symptoms or Conditions186
- No Known Impact Or Consequence To Patient121
- Unspecified Infection47
- Shock from Patient Lead(s)30
- No Consequences Or Impact To Patient20
- Syncope/Fainting6
- Sepsis4
- Endocarditis3
Recent adverse event reports
The 25 most recent of 421 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Over-Sensing
It was reported that the Right Ventricular (RV) lead triggered a Lead Integrity Alert (LIA) for High Rate Non-Sustained episodes and Sensing Integrity Counter (SIC). The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50696417 ↗MalfunctionSep 1, 2026Under-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50545180 ↗MalfunctionAug 27, 2026Signal Artifact/Noise; Over-Sensing
It was reported that the Right Ventricular (RV) lead triggered a Lead Integrity Alert (LIA) for high-rate non-sustained episodes and Sensing Integrity Counter (SIC). The RV lead also exhibited oversensing and non-physiologic intervals/noise. The RV lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50479053 ↗MalfunctionAug 26, 2026Under-Sensing
It was further reported that the patient had a heart transplant and the Implantable Cardioverter Defibrillator (ICD) system was explanted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50462135 ↗MalfunctionJul 10, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT WAS HEARING AN ALERT TONE EVERY FOUR HOURS, INDICATIVE OF A POSSIBLE LEAD INTEGRITY ISSUE. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25785288 ↗InjuryJun 9, 2026Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA, DIZZINESS, AND SYNCOPE AND WAS HOSPITALIZED. THE RV LEAD EXHIBITED LEAD INTEGRITY ALERT (LIA) WARNING FOR NON SUSTAINED EPISODES WITH POSSIBLE NOISE. IT WAS ALSO REPORTED THAT THE RV LEAD EXHIBITED HIGH IMPEDANCE, NO PACING NOTED IN TRUE BIPOLAR OR TIP TO COIL PACING ATTEMPTS DUE TO OVERSENSING AND SUSPECTED FRACTURE. DEVICE SETTINGS WERE RE-PROGRAMMED WITH DETECTIONS OFF. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Dizziness; Syncope/Fainting
MAUDE report 25443325 ↗MalfunctionMay 12, 2026Capturing Problem; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A GRADUAL DECLINE IN PACING IMPEDANCE AND RV AND SUPERIOR VENA CAVA (SVC) DEFIBRILLATION IMPEDANCE. THERE WAS ALSO A GRADUAL INCREASE IN CAPTURE THRESHOLD. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25156275 ↗MalfunctionMay 5, 2026High impedance
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PALPITATIONS AND THE RIGHT VENTRICULAR (RV) LEAD HAD TRIGGERED ALERTS DUE TO HIGH, OUT-OF-RANGE SUPERIOR VENA CAVA (SVC) DEFIBRILLATION IMPEDANCE. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Palpitations
MAUDE report 25081149 ↗MalfunctionApr 30, 2026Under-Sensing; Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT UNDERSENSING DURING A VENTRICULAR ARRHYTHMIA, POST-DETEC TION OF ARRHYTHMIA. THE DEVICE DELIVERED APPROPRIATE THERAPY. THE RV LEAD ALSO TRIGGERED AN ALERT DUE TO LOW, OUT-OF-RANGE PACING IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25044713 ↗InjuryApr 28, 2026Signal Artifact/Noise
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH VOLTAGE (HV) NOISE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019389 ↗MalfunctionMar 6, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING ON STORED EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24534863 ↗MalfunctionFeb 24, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS) POST VENTRICULAR SENSE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24434662 ↗MalfunctionFeb 18, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONIC RV HIGH THRESHOLD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379264 ↗InjuryFeb 18, 2026High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH/UNDEFINED SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE. IT WAS ALSO REPORTED THAT A PATIENT REPRESENTATIVE NOTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) IS ¿FLAWED.¿ THE RV LEAD AND ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379031 ↗InjuryFeb 11, 2026Capturing Problem; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING AND HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315369 ↗InjuryFeb 11, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS REMOVED DUE TO CANDIDEMIA INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Fungal Infection; Medical device site infection; Sepsis; Unspecified Infection
MAUDE report 24314824 ↗InjuryFeb 9, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24293114 ↗InjuryJan 29, 2026Fracture; High impedance; Impedance Problem
IT WAS REPORTED THAT THE EPICARDIAL RIGHT VENTRICULAR (RV) LEAD EXHIBITED A FLUCTUATION IN IMPEDANCES. THE RV DEFIBRILLATION, SUPERIOR VENA CAVA (SVC) COIL, AND RV PACING IMPEDANCES WERE UNDEFINED HIGH OUT OF RANGE. DURING THE SYSTEM CHANGEOUT, THE RV LEAD WAS FOUND TO HAVE A COIL FRACTURE. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24215337 ↗InjuryJan 28, 2026Over-Sensing; Impedance Problem
IT WAS FURTHER REPORTED THAT TACHYARRHYTHMIA THERAPIES WERE TURNED OFF AND THE LEAD WILL BE MONITORED UNTIL THE PATIENT'S NEXT FOLLOW-UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199825 ↗MalfunctionDec 17, 2025Defibrillation/Stimulation Problem; Low impedance; Defibrillation/Stimulation Problem; Low impedance
IT WAS REPORTED THAT FOLLOWING A PHYSICAL ALTERCATION AND A FALL THE PATIENT RECEIVED INAPPROPRIATE SHOCKS FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE PATIENT NOTED DISCOMFORT AND A JERKING FEELING IN THEIR CHEST. UPON REVIEW OF THE EPISODES, IT WAS ALSO NOTED THAT SHORT CIRCUIT PROTECTION (SCP) OCCURRED RESULTING IN LESS THAN PROGRAMMED HIGH VOLTAGE ENERGY DELIVERY FOR ALL SIX SHOCKS PROVIDED. IT COULD NOT BE DETERMINED WHETHER THE ICD OR RIGHT VENTRICULAR (RV) LEAD WAS RESPONSIBLE FOR THE SCP. OUT OF RANGE (OOR) LOW IMPEDANCE WAS ALSO OBSERVED ON THE RV LEAD DEFIBRILLATION AND SUPERIOR VENA CAVA (SVC) COILS. THE RV LEAD REMAINS IN USE AND THE ICD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23828640 ↗MalfunctionDec 17, 2025Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR OUT OF RANGE (OOR) LOW PACING IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23827699 ↗MalfunctionDec 10, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLY INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH-VOLTAGE THERAPY FOR ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) CLASSIFIED AS VENTRICULAR TACHYCARDIA (VT). ADDITIONALLY, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RARE UNDERSENSING NOTED ON A STORED EPISODE. THE ICD AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771559 ↗InjuryDec 4, 2025Signal Artifact/Noise; High impedance; Over-Sensing; Signal Artifact/Noise; High impedance; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED ALERTS FOR UNDEFINED HIGH OUT OF RANGE (OOR) BIPOLAR AND TIP TO COIL PACING IMPEDANCES, NOISE, AND A LEAD INTEGRITY ALERT (LIA) FOR HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC)COUNTS. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23715929 ↗InjuryNov 17, 2025Fracture; High impedance; Impedance Problem; Fracture; High impedance; Impedance Problem
IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED MULTIPLE SPIKES IN PACING IMPEDANCE A YEAR AND A HALF PRIOR TO THE LEAD BEING CAPPED. AN RV LEAD FRACTURE WAS SUSPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23569244 ↗InjuryOct 31, 2025High impedance; Impedance Problem; Unstable Impedance; High impedance; Impedance Problem; Unstable Impedance
IT WAS REPORTED THAT A LEAD IMPEDANCE WARNING TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD DUE TO A SPIKE TO UNDEFINED, HIGH SUPERIOR VENA CAVA (SVC) IMPEDANCE. THE SVC COIL WAS PROGRAMMED OFF, AND THE LEAD REMAINS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23438972 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page