PROFILE 3D® 680R34Recalls & reported problems
Medtronic, Inc. · RING 680R34 PROFILE 3D 26L MEXICO · Class 2 · Implant
All models of PROFILE 3D (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00613994759214), received between May 18, 2018 and Apr 21, 2026 — classified as 16 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information13
- Adverse Event Without Identified Device or Use Problem3
- Inadequacy of Device Shape and/or Size2
- Patient Device Interaction Problem1
Patient problems
- Insufficient Information8
- No Information5
- Mitral Regurgitation2
- Mitral Valve Insufficiency/ Regurgitation2
- Heart Block1
- Mitral Valve Prolapse1
- No Code Available1
- Non specific EKG/ECG Changes1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryApr 21, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 1 YEAR AND 4 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 24938302 ↗InjuryFeb 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 3 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 33MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 21401987 ↗InjuryOct 8, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 11 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, ENDOCARDITIS WAS SUSPECTED BASED ON A ECHOCARDIOGRAM PERFORMED. IT WAS REPORTED THAT 3 DAYS FOLLOWING THE ECHOCARDIOGRAM, THE DEVICE WAS EXPLANTED AND REPLACED WITH AN UNKNOWN VALVE PRODUCT FROM A DIFFERENT MANUFACTURER. IT WAS NOTED DURING THE REPLACEMENT PROCEDURE THAT THERE WERE "ENDOCARDITIC DEPOSITS" ON THE RING. ANTIBIOTIC MEDICATION WAS PRESCRIBED. THE ADVERSE EVENT WAS ASSESSED AS PROBABLY RELATED TO THE DEVICE BUT NOT RELATED TO THE IMPLANT PROCEDURE. IT WAS REPORTED THAT A MICROBIOLOGY CULTURE FOUND NO GROWTH OF BACTERIA AND YEAST FUNGI. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 20397050 ↗InjuryMar 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 2 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 25MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18856648 ↗InjurySep 12, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 8 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15401957 ↗InjuryJul 19, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
MEDTRONIC RECEIVED INFORMATION THAT 8 DAYS POST IMPLANT OF THIS 34MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 34MM RING OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS REPORTED TO BE DUE TO THE SADDLE SHAPE SITTING TOO HIGH AND NOT BEING A GOOD FIT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15052162 ↗InjuryOct 18, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 7 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 12651460 ↗InjuryJul 28, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 5 MONTHS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 12242053 ↗InjuryMay 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT 5 DAYS POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED IN THE PATIENT. THE REASON FOR THE PERMANENT PACEMAKER IMPLANT WAS REPORTED AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Heart Block
MAUDE report 11821941 ↗InjuryApr 13, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 34 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 11657024 ↗InjurySep 30, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 34 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29 MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 10606110 ↗InjuryApr 13, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Regurgitation; No Information
MAUDE report 9954875 ↗InjuryMar 13, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 34MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 31MM NON-MEDTRONIC BIOPROSTHETIC VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 9829728 ↗InjuryMar 11, 2019Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS MITRAL ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Prolapse; No Information
MAUDE report 8409858 ↗InjuryJul 23, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT FIVE YEARS AND TWO MONTHS POST IMPLANT OF THIS ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH A BIOPROSTHETIC VALVE. THE REASON FOR REPLACEMENT WAS DUE TO RUPTURED NEOCHORDS (THAT HAD ALSO BEEN IMPLANTED FIVE YEARS AND TWO MONTHS PRIOR, TO REPLACE THE NATIVE CHORDAE TENDONAE), AND A FLAILING LEAFLET THAT ULTIMATELY CAUSED SEVERE REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Regurgitation; No Code Available; Regurgitation; Rupture
MAUDE report 7710268 ↗InjuryMay 18, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT NINE MONTHS POST IMPLANT OF THIS 34MM ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH A 29MM ANNULOPLASTY RING. THE REASON FOR EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 7526629 ↗
Other RING, ANNULOPLASTY devices
- CG FUTURE™ 638BL32Medtronic, Inc. · RING, ANNULOPLASTY1 death71 reports
- PROFILE 3D® 680R28Medtronic, Inc. · RING, ANNULOPLASTY1 death67 reports
- CG FUTURE™ 638BL30Medtronic, Inc. · RING, ANNULOPLASTY1 death65 reports
- CG FUTURE™ 638RL32Medtronic, Inc. · RING, ANNULOPLASTY59 reports
- Tri-Ad™ 900SFC228Medtronic, Inc. · RING, ANNULOPLASTY4 deaths56 reports
- CG FUTURE™ 638BL34Medtronic, Inc. · RING, ANNULOPLASTY51 reports
- Contour 3D® 690R30Medtronic, Inc. · RING, ANNULOPLASTY1 death50 reports
- CG FUTURE™ 638BL28Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page