PROFILE 3D® 680R30Recalls & reported problems

Medtronic, Inc. · RING 680R30 PROFILE 3D 26L MEXICO · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Profile 3D 680R30 is a flexible ring used to repair the mitral valve, the heart's main valve between the left chambers. The 42 reports mostly describe the ring being taken out and replaced during or after surgery, along with ongoing valve leakage or narrowing, thickened and calcified leaflets, and heart block or rhythm problems in some patients; 39 reports involve injuries and 1 death. Most reports give little detail about the device itself.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
1
Injuries reported
38
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00613994759238), received between Mar 6, 2017 and Nov 14, 2025 — classified as 1 death, 38 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information28
  • Adverse Event Without Identified Device or Use Problem10
  • Patient Device Interaction Problem5
  • Intravalvular regurgitation3
  • Calcified1
  • Device Operates Differently Than Expected1
  • Device Stenosis1

Patient problems

  • Insufficient Information14
  • No Information12
  • Mitral Valve Insufficiency/ Regurgitation9
  • Mitral Regurgitation8
  • Heart Block3
  • Non specific EKG/ECG Changes3
  • Mitral Valve Stenosis2
  • Aortic Dissection1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE IMPLANT OF THIS 30MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN NEW HEART RING IN THE MITRAL POSITION. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 23554313 ↗
  • InjuryOct 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS AN UNSUCCESSFUL REPAIR. IT WAS REPORTED THAT THE ANNULOPLASTY PRODUCT MALFUNCTIONED AND WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23427597 ↗
  • InjuryAug 26, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEARS AND 6 MONTHS POST IMPLANT OF THIS 30MM PROFILE 3D HEART RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 22894097 ↗
  • InjuryJun 16, 2025Calcified; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Calcified; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS SEVERE MITRAL STENOSIS AND TRACE MITRAL REGURGITATION. IT WAS ALSO REPORTED THAT THE MITRAL LEAFLETS WERE SEVERELY THICKENED AND CALCIFIED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation; Mitral Valve Stenosis

    MAUDE report 22209899 ↗
  • InjuryFeb 12, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 5 DAYS POST IMPLANT OF THESE PROFILE 3D AND CONTOUR 3D ANNULOPLASTY RINGS, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 21363123 ↗
  • MalfunctionNov 14, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 5 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS REPORTED THAT A FOLLOW-UP ECHOCARDIOGRAM (ECHO) DISCOVERED MODERATE MITRAL REGURGITATION. IT WAS ALSO STATED THAT THE PATIENT WAS EXPERIENCING DYSPNEA. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Dyspnea; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 20688153 ↗
  • InjuryOct 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 2 DAYS POST IMPLANT OF THIS 30MM TRICUSPID ANNULOPLASTY RING AND 30MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS 3RD DEGREE ATRIOVENTRICULAR BLOCK/COMPLETE HEART BLOCK AND ATRIAL ARRHYTHMIAS INDICATED TO BE ATRIAL TACHYCARDIA, ATRIAL FIBRILLATION (AFIB) AND ATRIAL FLUTTER. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Atrial Fibrillation; Atrial Flutter; Heart Block; Tachycardia

    MAUDE report 20449258 ↗
  • DeathOct 9, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION FROM A PATIENT REPRESENTATIVE THAT ON THE SAME DAY AS THE IMPLANT DATE OF THIS 30MM ANNULOPLASTY RING, THE PATIENT DIED. THE CAUSE OF DEATH WAS NOT RECEIVED. THEREFORE, RELATEDNESS OF THE DEATH TO THE 30MM ANNULOPLASTY RING COULD NOT BE EXCLUDED. IT IS UNKNOWN WHETHER AN AUTOPSY WAS PERFORMED.

    Patient problems: Insufficient Information

    MAUDE report 20406880 ↗
  • InjurySep 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 31MM MITRAL MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20147651 ↗
  • InjuryMay 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19352834 ↗
  • InjuryMar 14, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 5 DAYS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS REPORTED THAT THE PATIENT HAD RIGHT BUNDLE BRANCH BLOCK (RBBB). ANTICOAGULANT, ASPIRIN AND ANTIPLATELET WERE GIVEN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 18908770 ↗
  • InjuryMar 14, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING AND THIS 32MM TRICUSPID ANNULOPLASTY RING, IT WAS REPORTED THAT THE PATIENT DEVELOPED THIRD-DEGREE ATRIOVENTRICULAR (AV) BLOCK. IT WAS ALSO REPORTED THAT TEMPORARY PACING WAS PROVIDED, THE AV BLOCK WAS TRANSIENT AND NO PERMANENT PACEMAKER WAS NEEDED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 18902451 ↗
  • InjuryJan 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 1 WEEK POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Cardiomyopathy

    MAUDE report 18428717 ↗
  • InjuryDec 27, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT ON THE SAME DAY POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, THIRD DEGREE ATRIO-VENTRICULAR BLOCK (AVB) WAS OBSERVED ON A ELECTROCARDIOGRAM (ECG). IT WAS STATED THAT THIS WAS POSSIBLY DUE TO PERIOPERATIVE DAMAGE OF THE CONDUCTION SYSTEM. 1 WEEK LATER, A PERMANENT PACEMAKER WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 18396894 ↗
  • InjuryMay 16, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE 30MM MITRAL ANNULOPLASTY RING WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC BIO PROSTHETIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 16937475 ↗
  • InjuryMay 8, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 11 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS MITRAL REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 16885067 ↗
  • InjuryMar 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 2 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE MITRAL REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 16470723 ↗
  • InjuryOct 14, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A N ON-MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. IT WAS REPORTED THAT THE PATIENT'S NATIVE MITRAL VALVE REQUIRED REPLACEMENT DUE TO ENDOCARDITIS AND A REPAIR WAS ATTEMPTED BUT ULTIMATELY REPLACED WITH A NON-MEDTRONIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 15611496 ↗
  • InjuryFeb 1, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 8 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 13415372 ↗
  • InjuryJan 21, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED. THE REASON FOR THE EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 13315486 ↗
  • InjuryOct 18, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 7 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 12647526 ↗
  • InjuryAug 31, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM BIOPROSTHETIC VALVE. NO MALFUNCTION OF THE ANNULOPLASTY RING WAS REPORTED. THE PHYSICIAN DECIDED TO REPLACE THE VALVE INSTEAD OF REPAIRING IT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 12393572 ↗
  • InjuryApr 26, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 31MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 11725410 ↗
  • InjuryNov 2, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 10 MONTHS POST IMPLANT OF THIS 30 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 10771171 ↗
  • InjuryAug 12, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED. THE REASON FOR THE REPLACEMENT AND REPLACEMENT PRODUCT WERE NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 10395272 ↗

Other RING, ANNULOPLASTY devices

All RING, ANNULOPLASTY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page