PROFILE 3D® 680R26Recalls & reported problems

Medtronic, Inc. · RING 680R26 PROFILE 3D 26L MEXICO · Class 2 · Implant

All models of PROFILE 3D (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Profile 3D annuloplasty ring is an implant used to help repair a heart valve. The 24 reports mostly describe rings being removed and replaced, sometimes with leaking valves mentioned, and patients receiving pacemakers after the implant. There are no deaths and 24 injuries mentioned, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
24
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00613994759252), received between Jan 25, 2017 and Nov 26, 2025 — classified as 24 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information18
  • Adverse Event Without Identified Device or Use Problem3
  • Inadequacy of Device Shape and/or Size3
  • Patient Device Interaction Problem3
  • Intravalvular regurgitation2
  • Perivalvular Leak1

Patient problems

  • Insufficient Information10
  • No Information9
  • Mitral Valve Insufficiency/ Regurgitation5
  • Mitral Regurgitation4
  • Non specific EKG/ECG Changes2
  • Endocarditis1
  • Mitral Valve Stenosis1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT POST IMPLANT OF THIS 26MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 23657784 ↗
  • InjuryNov 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 2 DAYS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING AND 26MM TRICUSPID ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20746930 ↗
  • InjuryNov 8, 2024Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 5 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 20641896 ↗
  • InjuryAug 22, 2024Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS MODERATE TO SEVERE MITRAL REGRUGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 20025674 ↗
  • InjuryJun 17, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 5 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 19549164 ↗
  • InjuryApr 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION 11 YEARS AND 4 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO AN INTENSIVE CARE UNIT (ICU). THE REASON FOR HOSPITALIZATION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19152924 ↗
  • InjuryApr 10, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 2 DAYS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SINUS BRADYCARDIA. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 19081877 ↗
  • InjuryJun 15, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE IMPLANT POSITION OF THE ANNULOPLASTY RING WAS THE MITRAL POSITION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 17138658 ↗
  • InjuryAug 24, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 8 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 31MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 15281931 ↗
  • InjuryApr 19, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 3 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Endocarditis; Insufficient Information; Mitral Valve Stenosis

    MAUDE report 14141104 ↗
  • InjuryApr 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, MODERATE RESIDUAL REGURGITATION WAS PRESENT. THE PATIENT WAS PLACED BACK ON CARDIOPULMONARY BYPASS (CPB), THE RING WAS EXPLANTED, AND THE ENTIRE NATIVE VALVE WAS REPLACED. IT WAS REPORTED THAT THERE WAS NO ALLEGED PRODUCT ISSUE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 11656568 ↗
  • InjuryJan 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 26 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON MEDTRONIC PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 11144235 ↗
  • InjuryJan 5, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 11119996 ↗
  • InjuryJul 8, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 25MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 10247766 ↗
  • InjuryFeb 4, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27 MM BIOPROSTHETIC VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Regurgitation; No Information

    MAUDE report 9665272 ↗
  • InjuryOct 8, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH A 25MM ANNULOPLASTY RING OF A DIFFERENT MODEL. THE PHYSICIAN STATED THAT THE 26MM RING WAS NOT THE RIGHT SIZE FOR THE PATIENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 9167028 ↗
  • InjuryJul 12, 2019Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 24MM MITRAL ANNULOPLASTY RING OF A DIFFERENT MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Regurgitation; No Information

    MAUDE report 8785021 ↗
  • InjuryDec 9, 2018Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH A 24MM RING OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8145249 ↗
  • InjuryNov 9, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY FOLLOWING THE IMPLANT OF THIS MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A BIOPROSTHETIC VALVE. THE REASON FOR EXPLANT IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8056480 ↗
  • InjuryMar 29, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 26MM ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 7382204 ↗
  • InjuryMar 20, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS 7 MONTHS POST IMPLANT OF THIS 26MM MITRAL ANNULOPLASTY RING WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL BIOPROSTHETIC VALVE DUE TO MITRAL REGURGITATON. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.

    Patient problems: Mitral Regurgitation

    MAUDE report 7354712 ↗
  • InjuryJan 18, 2018Perivalvular Leak; Insufficient Device Problem Information; Perivalvular Leak; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS ANNULOPLASTY RING, THE RING WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 7203765 ↗
  • InjuryFeb 2, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION FROM THIS PATIENT'S PHYSICIAN THAT THIS VALVE WAS REPLACED DUE TO RECURRENT MITRAL VALVE REGURGITATION. UPON OPENING THE PATIENT'S LEFT ATRIUM AND VISUALIZING THE VALVE, THE PHYSICIAN NOTED THAT "THE ATRIAL CHORD IN THE P2 POSITION APPEARED TO BE REDUNDANT AND LOOSE. THE REMAINDER OF THE VALVE APPEARED TO BE THICKENED, AND THE ANTERIOR LEAFLET PROLAPSED ABOVE THE LEVEL OF THE ANNULUS." AN ANNULOPLASTY RING WAS IMPLANTED, AND A "WEDGE RESECTION PERFORMED OF THE FLAIL SEGMENT OF THE P2 LEAFLET." THIS RING WAS SUBSEQUENTLY EXPLANTED AND REPLACED WITH A MECHANICAL VALVE DUE TO A "MILD LEAK, WHICH APPEARED TO BE SECONDARY TO THE ANTERIOR LEAFLET PROLAPSE." NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Regurgitation

    MAUDE report 6298314 ↗
  • InjuryJan 25, 2017Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY FOLLOWING IMPLANT OF THIS ANNULOPLASTY RING IN THE MITRAL POSITION, IT WAS EXPLANTED AND REPLACED. NO FAILURE MECHANISM WAS PROVIDED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 6277203 ↗

Other RING, ANNULOPLASTY devices

All RING, ANNULOPLASTY devices →