MOSAIC ULTRA® 305U223 (UDI 00613994760623)Recalls & reported problems
Medtronic, Inc. · VALVE 305U223 MOS AOR ULT CINCHII US · Class 3 · Implant
All models of MOSAIC ULTRA (40)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Mosaic Ultra aortic valve is a surgically implanted tissue valve that replaces a diseased aortic heart valve. The 62 reports mostly describe the valve narrowing over time (stenosis) or leaking backward (regurgitation), often leading years later to a catheter-based valve-in-valve replacement;58 reports involved injuries and no deaths. Reports fell sharply, from 13 in the prior 12 months to 1 in the last12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 62
- Deaths reported
- 0
- Injuries reported
- 58
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 00613994760623), received between Jul 24, 2017 and Mar 18, 2026 — classified as 58 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (42 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information42
- Device Stenosis11
- Intravalvular regurgitation7
- Gradient Increase6
- Patient Device Interaction Problem2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Inadequacy of Device Shape and/or Size1
Patient problems
- Aortic Valve Stenosis35
- Aortic Valve Insufficiency/ Regurgitation17
- Insufficient Information17
- No Information8
- Aortic Insufficiency6
- Aortic Regurgitation3
- Atrial Fibrillation2
- Atrial Flutter2
Recent adverse event reports
The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.
InjuryMar 18, 2026Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR EVALUATION WAS VALVE STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 24625502 ↗InjuryJul 18, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 8 MONTHS POST IMPLANT OF THIS MOSAIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22539982 ↗InjuryJun 5, 2025Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS SEVERE AORTIC STENOSIS AND MILD AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 22151408 ↗InjuryJun 4, 2025Device Stenosis; Device Stenosis
IT WAS REPORTED THAT POST IMPLANT OF THIS MOSAIC AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS STENOSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Aortic Valve Stenosis
MAUDE report 22143462 ↗MalfunctionJun 3, 2025Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 22136716 ↗InjuryMay 29, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. SUBSEQUENTLY, IN AN UNKNOWN DATE, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A NON MEDTRONIC TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 22108384 ↗InjuryMay 16, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM MOSAIC AORTIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS ATRIAL ARRHYTHMIAS (ATRIAL FIBRILLATION, ATRIAL FLUTTER & ATRIAL TACHYCARDIA). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Atrial Fibrillation; Atrial Flutter; Tachycardia
MAUDE report 22037540 ↗InjuryMar 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT DID NOT HAVE ANY PRE-EXISTING ARRHYTHMIAS PRIOR TO THE HEART VALVE IMPLANT AND THE PHYSICIAN DOES NOT BELIEVE THE HEART VALVE OR IMPLANT PROCEDURE CAUSED OR CONTRIBUTED TO THE CONDUCTION DISTURBANCE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Atrial Fibrillation; Atrial Flutter; Tachycardia
MAUDE report 21677393 ↗InjuryFeb 21, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS 1 MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 21433595 ↗MalfunctionFeb 4, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A UNKNOWN REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 21304684 ↗InjuryJan 29, 2025Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED WHICH STATED THAT SUBSEQUENTLY, 6 WEEKS LATER, IT WAS REPORTED THAT A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE IN THE AORTIC POSITION. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC STENOSIS (AS) AND MODERATE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 21261850 ↗MalfunctionJan 14, 2025Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 8 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 21149568 ↗InjuryJan 10, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A NON-MEDTRONIC TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 21124027 ↗MalfunctionDec 4, 2024Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC INSUFFICIENCY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 20840617 ↗InjuryAug 22, 2024Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 10 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVA LUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS HIGH GRADIENTS. IT WAS NOTED THAT THERE WAS NO EVIDENCE OF THROMBUS ON THE BIOPROSTHETIC VALVE. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 5 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 20027028 ↗InjuryJul 8, 2024Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 2 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 19687560 ↗InjuryApr 29, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 4 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 19209402 ↗InjuryMar 7, 2024Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE REASON FOR INTERVENTION WAS DUE TO SEVERE AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 18854211 ↗InjuryFeb 6, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 3 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18649882 ↗InjuryNov 9, 2023Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 18099417 ↗InjuryNov 7, 2023Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 5 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 21 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 18087690 ↗InjuryOct 30, 2023Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION 13 YEARS AND 1 MONTH POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 MONTH AND 12 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND MODERATE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 18030693 ↗InjuryAug 11, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 5 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 17511107 ↗InjuryJul 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 17245373 ↗InjuryMay 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 8 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 16860361 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page