MOSAIC® 305C225 (UDI 00613994760708)Recalls & reported problems
Medtronic, Inc. · VALVE 305C225 MOSAIC AOR CINCHII US · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Mosaic heart valve is an implanted tissue valve used to replace a narrowed aortic valve. Reports mostly describe the valve narrowing or leaking many years after implant, leading to evaluation or replacement with a new valve placed inside it. There are 24 reports that name its identifier, with 24 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00613994760708), received between Nov 8, 2017 and Feb 9, 2026 — classified as 24 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (14 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information14
- Device Stenosis4
- Intravalvular regurgitation4
- Fluid/Blood Leak2
- Gradient Increase2
- Patient Device Interaction Problem2
- Appropriate Device Problem Term/Code Not Available1
- Calcified1
Patient problems
- Aortic Valve Stenosis11
- Insufficient Information9
- Aortic Valve Insufficiency/ Regurgitation7
- Cardiomyopathy2
- Aortic Regurgitation1
- Bradycardia1
- Dyspnea1
- Host-Tissue Reaction1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
InjuryFeb 9, 2026Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT 14 YEARS AND 8 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 24289895 ↗InjuryFeb 6, 2026Fluid/Blood Leak; Patient Device Interaction Problem
IT WAS REPORTED THAT 14 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS BRADYCARDIA-TACHYCARDIA SYNDROME AND SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Bradycardia; Tachycardia; Valvular Insufficiency/ Regurgitation
MAUDE report 24277845 ↗InjuryJan 30, 2026Fluid/Blood Leak; Intravalvular regurgitation
ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR EVALUATION WAS REPORTED AS SEVERE REGURGITATION. SUBSEQUENTLY ON (B)(6) 2026, THE PATIENT SUCCESSFULLY HAD A 26MM AORTIC TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 24220183 ↗InjuryJun 6, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE AORTIC REGRUGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22162014 ↗InjuryFeb 18, 2025Device Stenosis; Device Stenosis
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 25MM MOSAIC AORTIC VALVE IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. SUBSEQUENTLY, 17 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 21393231 ↗InjuryNov 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 20761753 ↗InjurySep 13, 2024Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVAL UATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 20211353 ↗InjuryAug 26, 2024Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY 28 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND MODERATE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 20065389 ↗InjuryJun 3, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiomyopathy
MAUDE report 19448445 ↗InjuryMay 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 8 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 19325891 ↗InjuryMay 10, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiomyopathy
MAUDE report 19288314 ↗InjuryNov 30, 2023Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY 12 YEARS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE IMPLANTED IN THE PUL MONARY POSITION, A 18MM TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS OUTGROWTH. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18239656 ↗InjuryJun 12, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 3 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE IMPLANTED IN THE PULMONARY LOCATION OF AN 8 YEAR OLD PEDIATRIC PATIENT, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 17113361 ↗InjuryMar 21, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 3 MONTHS AND 14 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 16584649 ↗InjuryJul 22, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A 26MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED TO BE DUE TO SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 15083067 ↗InjuryMar 30, 2022Gradient Increase; Insufficient Device Problem Information; Gradient Increase; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 13954058 ↗InjuryAug 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 2 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A 26MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 12295183 ↗InjuryApr 16, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 11 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 11681406 ↗InjuryFeb 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 11 MONTHS POST IMPLANT OF THIS BIOPROSTHETIC AORTIC VALVE IMPLANTED IN THE PULMONARY POSITION OF A PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information; Non specific EKG/ECG Changes
MAUDE report 11308004 ↗InjuryOct 8, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY 9 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A NON-MEDTRONIC TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO MODERATE REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Regurgitation
MAUDE report 10648451 ↗InjuryNov 5, 2019Calcified; Incomplete Coaptation; Calcified; Incomplete Coaptation
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE IT WAS EXPLANTED AND REPLACED WITH A NON MEDTRONIC BIOPROSTHETIC VALVE DUE TO STENOSIS AND CALCIFICATION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.
Patient problems: Aortic Valve Stenosis; Host-Tissue Reaction
MAUDE report 9278932 ↗InjuryApr 16, 2019Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF A DIFFERENT MODEL DUE TO DYSPNEA AND A MEAN GRADIENT OF 35MMHG. UPON EXPLANT, IT WAS REPORTED THAT THE VALVE APPEARED "PRISTINE". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Dyspnea
MAUDE report 8518725 ↗InjuryMay 22, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT SEVEN YEARS POST IMPLANT OF THIS 25 MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED DUE TO STENOSIS. THE VALVE WAS REPLACED WITH A NON-MEDTRONIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 7534011 ↗InjuryNov 8, 2017Material Too Rigid or Stiff; Material Too Rigid or Stiff
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 7 MONTHS POST IMPLANT OF THIS BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH THE SAME SIZE AND MODEL. THE PHYSICIAN REPORTED THAT THE VALVE WAS STENOSED DUE TO THROMBOSIS ON THE LEAFLET, CAUSING LEAFLET STIFFNESS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Thrombosis
MAUDE report 7010993 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page