MOSAIC® 305C225 (UDI 00613994760715)Recalls & reported problems
Medtronic, Inc. · VALVE 305C225 MOS AOR CINCHII NNA US · Class 3 · Implant
All models of MOSAIC (59)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00613994760715), received between Aug 27, 2018 and Sep 3, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information6
- Device Stenosis3
- Gradient Increase3
- Appropriate Device Problem Term/Code Not Available1
- Inadequacy of Device Shape and/or Size1
- Intravalvular regurgitation1
- Material Split, Cut or Torn1
Patient problems
- Aortic Valve Stenosis6
- Aortic Valve Insufficiency/ Regurgitation4
- Insufficient Information3
- Aortic Insufficiency2
- Aortic Dissection1
- Aortic Regurgitation1
- Cusp Tear1
- Host-Tissue Reaction1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2025Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Device Stenosis; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Device Stenosis
IT WAS REPORTED THAT 1 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM AORTIC BIOPROSTHETIC VALVE OF DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 22952077 ↗InjuryAug 29, 2025Gradient Increase; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 1 MONTH'S POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION, AORTIC STENOSIS, A FLAIL LEAFLET AND A MEAN GRADIENT OF 40MMHG. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 3 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 22926193 ↗InjuryAug 22, 2024Gradient Increase; Appropriate Device Problem Term/Code Not Available; Device Stenosis; Gradient Increase; Appropriate Device Problem Term/Code Not Available; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS WITH A MEAN GRADIENT OF 42MMHG, A JET VELOCITY OF 4.33 M/S AND AN AORTIC VALVE AREA OF .33CM2. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20031464 ↗InjuryOct 12, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25 MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, ONE DAY LATER, A NON-MDT TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 15586107 ↗InjuryJun 27, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 3 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM AORTIC BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS "WIDE OPEN" AORTIC INSUFFICIENCY DUE TO A TEAR IN THE VALVE LEAFLET. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Cusp Tear
MAUDE report 14830491 ↗InjuryMar 10, 2022Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS STENOSIS, MEAN GRADIENT OF 30MMHG AND PEAK GRADIENT OF 45MMHG. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 13720968 ↗InjuryApr 26, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 1 MONTH POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER AORTIC VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 11723412 ↗InjuryFeb 10, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, IT WAS REPLACED VALVE-IN-VALVE WITH A 26MM TRANSCATHETER VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS A COMBINATION OF AORTIC STENOSIS, SEVERE AORTIC INSUFFICIENCY, MODERATE PARAVALVULAR LEAK, AND PANNUS GROWTH. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Insufficiency; Aortic Valve Stenosis; Host-Tissue Reaction; Insufficiency, Valvular
MAUDE report 9688859 ↗InjuryFeb 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT EIGHT YEARS AND THREE MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO STENOSIS AND MILD REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Regurgitation; Aortic Valve Stenosis
MAUDE report 8339256 ↗InjuryAug 27, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT SEVEN YEARS AND FOUR MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH ANOTHER AORTIC BIOPROSTHETIC VALVE OF THE SAME SIZE AND MODEL. THE REASON FOR THE EXPLANT WAS REPORTED AS AORTIC DISSECTION AND SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Dissection; Aortic Insufficiency
MAUDE report 7820597 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page