CD HORIZON® Spinal System 1475501055Recalls & reported problems

Medtronic Sofamor Danek, Inc. · ROD 1475501055 4.75 CCM NS CURV 55MM · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994792723), received between Apr 24, 2023 and Apr 24, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

1 other recall cover devices cleared under the same FDA submission (K113174); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 24, 2023Migration; Migration

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH SCREW AND ROD USED IN L4-S1 TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF). IT WAS REPORTED THAT ROD ALMOST FULLY RELEASED FROM THE LEFT S1 SCREW HEAD WHILE CAP REMAINED ON. THE SCREWHEAD LOOKED LIKE FAULTY. THERE WAS SCREW HEAD MOVEMENT AND ROD MIGRATED OUT OF SCREW HEAD. PRE-OPERATIVE DIAGNOSIS FOR THE PATIENT WAS LEG PAIN. S1 SCREW AND ROD ON LEFT SIDE WERE EXPLANTED AS AN ADDITIONAL SURGERY ON (B)(6) AND WERE REPLACED. PATIENT WAS EXPERIENCING LEG PAIN. THERE WERE NO FURTHER COMPLICATIONS THAT WERE REPORTED.

    Patient problems: Pain

    MAUDE report 16795428 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113174. One submission often covers many models, so check the product and lot details in each record.

  • Z-2603-2018 · Jul 9, 2018 · Terminated
    Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.