Medtronic, Inc. · LEAD 6707-15 US MKT · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994793942), received between Oct 10, 2017 and Nov 13, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Fracture1
  • Low impedance1
  • Mechanical Problem1

Patient problems

  • Muscle Stimulation1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2018Fracture; Fracture

    A LEAD ADAPTOR APPARENTLY PREVIOUSLY ASSOCIATED WITH THE RV LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8067353 ↗
  • InjuryOct 10, 2017Mechanical Problem; Low impedance; Appropriate Device Problem Term/Code Not Available; Mechanical Problem; Low impedance; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING DIAPHRAGMATIC STIMULATION. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW IMPEDANCE. THE PACE/SENSE RV COIL AND THE SUPERIOR VENA CAVA (SVC) COIL WERE REVERSED IN THE DEVICE AND THE PACE/SENSE PINS WERE ALSO REVERSED IN THE LEAD ADAPTOR. THE POCKET WAS OPENED AND THE LEAD AND THE ADAPTOR WERE PLACED IN THE PROPER POSITIONS IN RELATION TO THE DEVICE AND RV/SVC COILS. THE ADAPTOR REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Muscle Stimulation

    MAUDE report 6936272 ↗