CD HORIZON® Spinal System 1555000500Recalls & reported problems

Medtronic Sofamor Danek, Inc. · ROD 1555000500 5.5 CCM NS STRHT 500MM · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00613994990037), received between Apr 17, 2018 and Jan 5, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Break” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K113174); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7

Patient problems

  • Discomfort2
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 5, 2024Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING GROWTH ROD EXTE NSION / REVISION FOR SCOLIOSIS. IT WAS REPORTED THAT ROD TO BE REPLACED DUE TO BREAKING INSITU. DURING PROCEDURE TO REPLACE ROD METALLOSIS WAS NOTED IN THE GROWTH ROD CONNECTOR SO IT WAS REPLACED WITH A NEW GROWTH CONNECTOR AND TITANIUM ROD. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18456631 ↗
  • InjuryDec 20, 2023Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED IN A PATIENT POST A SCOLIOSIS CORRECTION PROCEDURE. IT WAS REPORTED THAT RODS BROKE TWICE IN THE PATIENT. THE RODS WERE EXPLANTED. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18361964 ↗
  • InjuryJan 9, 2020Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT THORACIC AND LUMBAR ARTHRODESIS. ON AN UNKNOWN DATE, POST-OP, THE ROD BROKE. HENCE, A REVISION SURGERY WAS PERFORMED, IN WHICH A CONNECTOR WAS FIXED TO THE ROD; AND ANOTHER ROD WAS ALSO ADDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9571697 ↗
  • InjuryNov 18, 2019Break; Break

    PATIENT GOT THE REVISION OR ON (B)(6) 2018 DUE TO AN EASING OF THE SACRUM ONE SCREWS ON BOTH SIDES. DURING THIS REVISION THE PATIENT GOT THICKER SCREWS IN THE SACRUM ONE. IN (B)(6) 2019 THE PATIENT GOT THE FOLLOWING BESIDES DISEASES: PULMONARY EDEMA, EPILEPSY ATTACK, INFARCT. NOW THE PATIENT GET TO THE HOSPITAL OF STRANGE NOISES DURING MOVEMENTS. ON (B)(6) 2019 THE PATIENT IS GOING TO GET THE REVISION DUE TO THE BROKEN RODS. THE SURGEON IS GOING TO LEAVE USE SIDE TO SIDE CONNECTORS THE BUILD A PARALLEL BRIDGE WITH A SHORTER ROD. THE SURGEON DON`T WANT THE OPEN THE PATIENT AGAIN COMPLETELY FROM SACRUM TO THE TOP SCREWS TO IMPLANT A NEW ROD DUE TO THE MANY BESIDES DISEASES OF THE PATIENT. THE SURGEON IMPLANTED ROD, CONNECTORS AND SET SCREWS. THE PHYSICIAN MADE THE REVISION OF THE BROKEN WINGS ON (B)(6). INSTEAD OF EXPLANTING THE BROKEN RODS DR. (B)(6) BUILD AN DIVERSION WITH SIDE-TO-SIDE CONNECTORS ON THE MEDIAL SIDE OF THE EXISTING CONSTRUCT.

    Patient problems: Discomfort

    MAUDE report 9334057 ↗
  • InjuryNov 18, 2019Break; Break

    PATIENT GOT THE REVISION OR ON (B)(6) 2018 DUE TO AN EASING OF THE SACRUM ONE SCREWS ON BOTH SIDES. DURING THIS REVISION THE PATIENT GOT THICKER SCREWS IN THE SACRUM ONE. IN AUGUST 2019 THE PATIENT GOT THE FOLLOWING BESIDES DISEASES: PULMONARY EDEMA, EPILEPSY ATTACK, INFARCT. NOW THE PATIENT GET TO THE HOSPITAL OF STRANGE NOISES DURING MOVEMENTS. ON (B)(6) 2019 THE PATIENT IS GOING TO GET THE REVISION DUE TO THE BROKEN RODS. THE SURGEON IS GOING TO LEAVE USE SIDE TO SIDE CONNECTORS THE BUILD A PARALLEL BRIDGE WITH A SHORTER ROD. THE SURGEON DON`T WANT THE OPEN THE PATIENT AGAIN COMPLETELY FROM SACRUM TO THE TOP SCREWS TO IMPLANT A NEW ROD DUE TO THE MANY BESIDES DISEASES OF THE PATIENT. THE SURGEON IMPLANTED ROD, CONNECTORS AND SET SCREWS. THE PHYSICIAN MADE THE REVISION OF THE BROKEN WINGS ON 14TH OF NOVEMBER. INSTEAD OF EXPLANTING THE BROKEN RODS DR. HARTUNG BUILD AN DIVERSION WITH SIDE-TO-SIDE CONNECTORS ON THE MEDIAL SIDE OF THE EXISTING CONSTRUCT

    Patient problems: Discomfort

    MAUDE report 9333912 ↗
  • InjuryJun 28, 2019Break; Break

    IT WAS REPORTED THAT IN (B)(6) 2017, THE PATIENT UNDERWENT DEGENERATIVE SCOLIOSIS SURGERY. POST-OP, THE IMPLANTED ROD BROKE. HENCE, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2019, IN WHICH THE BROKEN ROD WAS COMPLETELY REMOVED. NO FRAGMENT OF THE BROKEN ROD REMAINED INSIDE THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8744937 ↗
  • InjuryApr 17, 2018Break; Break

    IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH SCOLIOSIS AND BACK PAIN AND UNDERWENT SCOLIOSIS CORRECTION FROM T10 TO L5 AND LAMINECTOMY T12-L4. POST-OP, ROD BREAKAGE WAS OBSERVED AFTER PATIENT'S LATERAL BENDING MOVEMENT. PAIN WAS OBSERVED AS THE RESULT OF THE EVENT. BROKEN ROD REPLACEMENT WAS PERFORMED DUE TO THE EVENT. NO FRAGMENTS OF THE BROKEN PART REMAINED IN THE PATIENT.

    Patient problems: Pain

    MAUDE report 7438223 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113174. One submission often covers many models, so check the product and lot details in each record.

  • Z-2603-2018 · Jul 9, 2018 · Terminated
    Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →