Skip to content

Astral 150 27003RRecalls & reported problems

Resmed Pty Ltd · Astral 150 - Refurbished · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Sep 13, 2023Open, Classified

    If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all

    Z-0111-2024 · Root cause: Device Design · Recall details · FDA record ↗

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page