ADVIA Centaur HCV 11203483Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · ADVIA Centaur HCV 200T Kit · Class 2

All models of ADVIA Centaur HCV (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414278759), received between Apr 4, 2024 and Jan 3, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2025False Negative Result; False Negative Result

    THE CUSTOMER REPORTS OBSERVATION OF A NONREACTIVE (NEGATIVE) ADVIA CENTAUR HCV (AHCV) RESULT WHICH WAS DISCORDANT RELATIVE TO REPEAT TESTING. AHCV REAGENT LOT#: 473 WAS IN USE AT THE TIME. AN INITIAL REACTIVE (POSITIVE) RESULT WAS OBTAINED FOR THE PATIENT IN QUESTION. THE SAME SAMPLE WAS RETESTED ON THE SAME ANALYZER AND OBTAINED A NONREACTIVE (NEGATIVE) RESULT. BOTH RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE NONREACTIVE (NEGATIVE) RESULT WAS CONSIDERED DISCORDANT BY THE CUSTOMER. THEREFORE, A FOLLOW-UP TESTING WAS PERFORMED FOR THE SAME SAMPLE IN DUPLICATE WHICH PRODUCED REACTIVE (POSITIVE) RESULTS. THE REACTIVE (POSITIVE) RESULTS WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE IS NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21065826 ↗
  • MalfunctionOct 25, 2024False Negative Result; False Negative Result

    THE CUSTOMER REPORTS AN OBSERVATION OF A NON-REACTIVE (NEGATIVE) ADVIA CENTAUR HCV (AHCV) RESULT WHICH WAS CONSIDERED DISCORDANT RELATIVE TO CLINICAL EXPECTATION. AHCV REAGENT LOT# 466 WAS IN USE AT THE TIME. AN INITIAL REACTIVE (POSITIVE) RESULT WAS OBTAINED FOR THE PATIENT IN QUESTION. THE INITIAL REACTIVE RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). PER ADVIA CENTAUR HCV (AHCV) INSTRUCTIONS FOR USE (IFU), THE CUSTOMER PERFORMED A FOLLOW-UP TESTING WITH THE SAME SAMPLE, WHICH PRODUCED A NONREACTIVE (NEGATIVE) RESULT. THIS RESULT WAS CONSIDERED DISCORDANT BY THE CUSTOMER. THE SAME SAMPLE WAS THEN REPEATED IN DUPLICATE WHICH PRODUCED REACTIVE (POSITIVE) RESULTS SIMILAR TO THE INITIAL RESULT. THE REACTIVE RESULTS WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S) AS IT MATCHED THE CLINICAL EXPECTATION FOR THE AFFECTED PATIENT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20531637 ↗
  • MalfunctionApr 4, 2024False Negative Result; False Negative Result

    THE CUSTOMER REPORTS OBSERVATION OF NONREACTIVE ADVIA CENTAUR HCV (AHCV) RESULTS WHICH WERE DISCORDANT RELATIVE TO PATIENT HISTORY AND ALTERNATE-METHOD TESTING. AHCV REAGENT LOT 459 WAS IN USE AT THE TIME. TWO PATIENTS PRODUCED NONREACTIVE (NEGATIVE) RESULTS IN INITIAL TESTING. THE PHYSICIAN QUESTIONED THESE RESULTS BASED ON PATIENT HISTORY. THE SAMPLES WERE RE-TESTED IN DUPLICATE THE FOLLOWING DAY, AND BOTH PRODUCED REACTIVE (POSITIVE) RESULTS WHICH WERE ACCEPTED AS CORRECT. CORRECTIVE REPORTS WERE FILED FOR BOTH PATIENTS; THE PHYSICIAN STATES THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19040282 ↗