Skip to content

Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP).… recallZ-2542-2025 · Siemens Healthcare Diagnostics, Inc. · initiated Aug 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Healthcare Diagnostics notified consignees on about 08/11/2025 via letter. Customers were instructed to perform QC on each well, track additional reagent consumption (number of tests) to report to Siemens Healthineers for future reimbursement/credit, review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, and retain this letter with your laboratory records and forward this letter to those who may have received this product. Additionally, consignees were instructed to complete and return the Field Correction Effectiveness Check Form.

Recall details

Recall number
Z-2542-2025
Status
Open, Classified
Initiated by the company
Aug 11, 2025
Posted by the FDA
Sep 8, 2025
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
11,811 units
Product code
CEK
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Macedonia, Malaysia, Maldives, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Uruguay, Vatikancity, Vietnam, Zaire.

Product description

Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.

Code information (models, lots, serial numbers)

Material Number: 11097543. Test Code: UCFP. UDI-DI: 00630414279206. All lot numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page