ADVIA Centaur® eE2 10491445Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Estradiol - CENTAUR - RGT - 500 tests · Class 1

All models of ADVIA Centaur eE2 (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00630414574646), received between Jul 16, 2018 and Jul 4, 2025 — classified as 14 malfunctions.

The most frequently reported specific device problem is “High Test Results” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 3, 2022Class IIOpen, Classified

    Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

    Z-1024-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Apparent Adverse Event10
  • High Test Results4

Patient problems

  • No Known Impact Or Consequence To Patient10
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 4, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED OBSERVATION OF ELEVATED ADVIA CENTAUR ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO REPEAT TESTING ON AN ALTERNATE METHOD WHEN USING LOT 104. AN INITIAL ELEVATED RESULT WAS OBTAINED FOR ONE (1) PATIENT SAMPLE. THE INITIAL ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS RETESTED ON THE ORIGINAL ANALYZER WHICH PRODUCED A SIMILAR ELEVATED RESULT. FOR TROUBLESHOOTING PURPOSES, THE SAME SAMPLE WAS TREATED WITH PEG (POLYETHYLENE GLYCOL 6000) AND RETESTED ON THE ORIGINAL ANALYZER WHICH PRODUCED A LOWER RESULT. THE SAMPLE WAS THEN TESTED ON AN ALTERNATE METHOD WHICH OBTAINED A LOWER RESULT. THIS LOWER RESULT WAS CONSIDERED CORRECT SINCE IT MATCHED THE CLINICAL PICTURE OF THE PATIENT AND WAS ISSUED ON THE CORRECTED REPORT TO THE PHYSICIAN. THERE ARE NO ALLEGATIONS OF PATIENT HARM, CHANGES IN TREATMENT, OR DELAYS OF DIAGNOSIS OR TREATMENT RESULTING FROM THE OBSERVED RESULT DISCORDANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22402685 ↗
  • MalfunctionMar 1, 2023High Test Results; High Test Results

    THE CUSTOMER REPORTS OBSERVATION OF ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULTS WHICH WERE DISCORDANT RELATIVE TO ALTERNATE-METHOD TESTING. PRODUCT LOT USED AND DETAILS OF TESTING WERE NOT PROVIDED. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16461620 ↗
  • MalfunctionJul 15, 2021High Test Results; High Test Results

    DISCORDANT, FALSELY ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) NEAT (UNDILUTED) AND ONBOARD (AUTO) DILUTION RESULTS WERE OBTAINED ON A PATIENT SAMPLE. THE ELEVATED RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE CUSTOMER PERFORMED REPEAT TESTING ON THREE ALTERNATE ESTRADIOL TEST METHODS WITH THE SAME PATIENT SAMPLE (NEAT), ALL RESULTING LOWER. THE LOWER ALTERNATE ESTRADIOL TEST RESULTS AGREED WITH THE CLINICAL PICTURE OF THIS PATIENT. THE LOWER ESTRADIOL RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12176466 ↗
  • MalfunctionSep 22, 2020High Test Results; High Test Results

    A DISCORDANT, FALSELY ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN AND QUESTIONED. A LOWER RESULT WAS EXPECTED BY THE PHYSICIAN DUE TO ADMINISTERED MEDICATION. THE CUSTOMER SENT THE SAMPLE TO AN ALTERNATE LABORATORY WITH AN ALTERNATE ESTRADIOL TEST METHOD, RESULTING LOWER. THE LOWER ALTERNATE ESTRADIOL TEST RESULT WAS ACCEPTED BY THE PHYSICIAN. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED, PRESCRIBED, OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10570805 ↗
  • MalfunctionMay 24, 2019No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE PHYSICIAN QUESTIONED THE RESULT. THE PHYSICIAN SUGGESTED THAT THE SAMPLE BE SENT TO AN ALTERNATE SITE AND TESTED WITH LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY ((LC-MS). THE LC-MS TESTING YIELDED THE EXPECTED RESULT. THE PATIENT HAD A PREVIOUS HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT THAT WAS NOT QUESTIONED. PATIENT IS TAKING FASLODEX. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8639896 ↗
  • MalfunctionMay 24, 2019No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE PHYSICAN QUESTIONED THE RESULT. THE PHYSICIAN SUGGESTED THAT THE SAMPLE BE SENT TO AN ALTERNATE SITE AND TESTED WITH LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY ((LC-MS). THE LC-MS TESTING YIELDED THE EXPECTED RESULT. THE PATIENT HAD A PREVIOUS HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT THAT WAS NOT QUESTIONED. PATIENT IS TAKING FASLODEX. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8639895 ↗
  • MalfunctionSep 13, 2018No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER OBSERVED ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS FOR THREE SAMPLES FROM THE SAME PATIENT COMPARED TO CLINICAL EXPECTATIONS AND AN ALTERNATE METHOD. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7872351 ↗
  • MalfunctionSep 13, 2018No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER OBSERVED ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS FOR THREE SAMPLES FROM THE SAME PATIENT COMPARED TO CLINICAL EXPECTATIONS AND AN ALTERNATE METHOD. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7872342 ↗
  • MalfunctionSep 13, 2018No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER OBSERVED ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS FOR THREE SAMPLES FROM THE SAME PATIENT COMPARED TO CLINICAL EXPECTATIONS AND AN ALTERNATE METHOD. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7872338 ↗
  • MalfunctionSep 13, 2018No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER OBSERVED ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS FOR THREE SAMPLES FROM THE SAME PATIENT COMPARED TO CLINICAL EXPECTATIONS AND AN ALTERNATE METHOD. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ELEVATED ADVIA CENTAUR XP ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7872337 ↗
  • MalfunctionAug 14, 2018No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE PHYSICIAN QUESTIONED THE RESULT AFTER A REDRAW SAMPLE WAS TESTED IN ANOTHER LABORATORY WITH THE ELECTROCHEMILUMINESCENCE METHOD AND THE RESULT WAS LOW. AFTER THE RESULT WAS QUESTIONED, THE LABORATORY REPEATED THE SAMPLE AGAIN AND THE RESULT WAS ELEVATED. THE PATIENT SAMPLE WAS SENT TO A LABORATORY OF SUPPORT AND TESTED. THE RESULT WAS LOW. THE PATIENT IS MENOPAUSAL. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7781970 ↗
  • MalfunctionJul 16, 2018No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PATIENT. THE PATIENT PROVIDED A PREVIOUS LOWER RESULT FROM AN ALTERNATE METHOD. THE LABORATORY REQUESTED A NEW COLLECTION. THE NEW SAMPLE WAS TESTED ON THE ADVIA CENTAUR XP EE2 ASSAY AND THE RESULT WAS HIGH. THE SAMPLE WAS THEN SENT TO THE LABORATORY OF SUPPORT AND ELECTROCHEMILUMINESCENCE TESTING WAS PERFORMED. THE RESULT WAS LOWER. THE PATIENT SAMPLE WAS TESTED AFTER RECENTRIFUGATION, NEAT AND DILUTED, DURING TECHNICAL VISITS, THE RESUTLS WERE HIGH. THE PATIENT IS IN THE POSTMENOPAUSAL PHASE AND DOES NOT TAKE HORMONAL SUPPLEMENTATION. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7691847 ↗
  • MalfunctionJul 16, 2018No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PATIENT. THE PATIENT PROVIDED A PREVIOUS LOWER RESULT FROM AN ALTERNATE METHOD. THE LABORATORY REQUESTED A NEW COLLECTION. THE NEW SAMPLE WAS TESTED ON THE ADVIA CENTAUR XP EE2 ASSAY AND THE RESULT WAS HIGH. THE SAMPLE WAS THEN SENT TO THE LABORATORY OF SUPPORT AND ELECTROCHEMILUMINESCENCE TESTING WAS PERFORMED. THE RESULT WAS LOWER. THE PATIENT SAMPLE WAS TESTED AFTER RECENTRIFUGATION, NEAT AND DILUTED, DURING TECHNICAL VISITS, THE RESULTS WERE HIGH. THE PATIENT IS IN THE POSTMENOPAUSAL PHASE AND DOES NOT TAKE HORMONAL SUPPLEMENTATION. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7691841 ↗
  • MalfunctionJul 16, 2018No Apparent Adverse Event; No Apparent Adverse Event

    A FALSE HIGH ADVIA CENTAUR XP ENHANCED ESTRADIOL (EE2) RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PATIENT. THE PATIENT PROVIDED A PREVIOUS LOWER RESULT FROM AN ALTERNATE METHOD. THE LABORATORY REQUESTED A NEW COLLECTION. THE NEW SAMPLE WAS TESTED ON THE ADVIA CENTAUR XP EE2 ASSAY AND THE RESULT WAS HIGH. THE SAMPLE WAS THEN SENT TO THE LABORATORY OF SUPPORT AND ELECTROCHEMILUMINESCENCE TESTING WAS PERFORMED. THE RESULT WAS LOWER. THE PATIENT SAMPLE WAS TESTED AFTER RECENTRIFUGATION, NEAT AND DILUTED, DURING TECHNICAL VISITS, THE RESULTS WERE HIGH. THE PATIENT IS IN THE POSTMENOPAUSAL PHASE AND DOES NOT TAKE HORMONAL SUPPLEMENTATION. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THESE DISCORDANT ENHANCED ESTRADIOL RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7691838 ↗

FDA clearance and classification

The FDA regulates product code CHP (radioimmunoassay, estradiol) under 21 CFR 862.1260 as a Class I device. 21 CFR 862.1260 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ADVIA Centaur® eE2 10491445

Has the ADVIA Centaur® eE2 10491445 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1024-2022 (Class II), began on Mar 3, 2022: Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ADVIA Centaur® eE2 10491445?

The FDA's MAUDE database lists 14 adverse event reports for this device, received between Jul 16, 2018 and Jul 4, 2025. In those reports, the most common specific device problem is “High Test Results” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ADVIA Centaur® eE2 10491445?

None of the reports are classified as deaths, 0 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ADVIA Centaur® eE2 10491445 FDA approved?

Devices of this type (product code CHP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Siemens Healthcare Diagnostics Inc.

Who makes the ADVIA Centaur® eE2 10491445, and is it still sold?

The FDA's device database lists Siemens Healthcare Diagnostics Inc. as the company responsible for it (brand ADVIA Centaur® eE2, model 10491445). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.