Atellica IM eE2 10995562Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Estradiol (eE2) Atellica IM RGT - 500T · Class 1
All models of Atellica IM eE2 (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00630414598864), received between Dec 17, 2021 and Apr 21, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 3, 2022Class IIOpen, Classified
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Z-1021-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2023High Test Results; High Test Results
THE CUSTOMER OBTAINED AN ELEVATED ENHANCED ESTRADIOL (EE2), LOT 090, INITIAL RESULT FOR SERUM SAMPLE TESTED ON AN ATELLICA IM (B)(6). THE RESULT WAS CONSIDERED DISCORDANT COMPARED TO THE LOWER RETEST RESULT OBTAINED ON A DIFFERENT DAY ON A DIFFERENT ATELLICA IM ANALYZER. THE INITIAL PATIENT RESULT HAD BEEN REPORTED, BUT NOT QUESTIONED BY PHYSICIAN(S) AT THE TIME. A CORRECTED REPORT WAS ISSUED AFTER RETESTING. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT EE2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16792070 ↗MalfunctionApr 18, 2023High Test Results; High Test Results
THE CUSTOMER OBTAINED ELEVATED ENHANCED ESTRADIOL (EE2), LOT 090, INITIAL RESULTS FOR SERUM SAMPLES TESTED ON AN ATELLICA IM 1600 ANALYZER (S/N (B)(4)). THE SAMPLES WERE FROM DIFFERENT PATIENTS. THE RESULTS WERE CONSIDERED DISCORDANT COMPARED TO THE LOWER RETEST RESULTS OBTAINED ON DIFFERENT DAYS ON DIFFERENT ATELLICA IM ANALYZERS WITH LOT 092. THE INITIAL PATIENT RESULTS HAD BEEN REPORTED, BUT NOT QUESTIONED BY PHYSICIAN(S) AT THE TIME. CORRECTED REPORTS WERE ISSUED AFTER RETESTING. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT EE2 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765551 ↗MalfunctionApr 18, 2023High Test Results; High Test Results
THE CUSTOMER OBTAINED AN ELEVATED ENHANCED ESTRADIOL (EE2), LOT 090, INITIAL RESULT FOR A SERUM SAMPLE TESTED ON AN ATELLICA IM 1600 ANALYZER (S/N (B)(4)). THE RESULT WAS CONSIDERED DISCORDANT COMPARED TO THE LOWER RETEST RESULT OBTAINED ON A DIFFERENT DAY ON A DIFFERENT ATELLICA IM ANALYZER WITH LOT 092. THE INITIAL PATIENT RESULT WAS REPORTED, BUT NOT QUESTIONED BY THE PHYSICIAN AT THE TIME. A CORRECTED REPORT WAS ISSUED AFTER RETESTING. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT EE2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765485 ↗MalfunctionDec 17, 2021High Test Results; High Test Results
A CUSTOMER OBSERVED A DISCORDANT, FALSELY-ELEVATED ATELLICA IM ENAHNCED ESTRADIOL (EE2) RESULT FOR ONE PATIENT. REPEAT TESTING PRODUCED A LOWER RESULT WHICH WAS ACCEPTED AS CORRECT. THERE ARE NO REPORTS THAT TREATMENT WAS ALTERED OR PRESCRIBED, OR OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY-ELEVATED ATELLICA IM ENHANCED ESTRADIOL (EE2) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13026288 ↗
Other RADIOIMMUNOASSAY, ESTRADIOL devices
- Atellica IM eE2 10995561Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, ESTRADIOL1 recall61 reports
- Elecsys Estradiol III 6656021190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, ESTRADIOL101 reports
- VIDAS® Estradiol II 30431-01Biomerieux SA · RADIOIMMUNOASSAY, ESTRADIOL2 recalls0 reports
- IMMULITE® 2000 Systems E2 Estradiol 10702833Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, ESTRADIOL1 recall34 reports
- ADVIA Centaur® eE2 10491445Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, ESTRADIOL1 recall14 reports
- Elecsys Estradiol III 07027249190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, ESTRADIOL61 reports
- ADVIA Centaur® eE2 10490889Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, ESTRADIOL1 recall8 reports
- IMMULITE® 2000 Systems E2 Estradiol 10702834Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, ESTRADIOL1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page