Skip to content

IMMULITE® 2000 Systems E2 Estradiol 10702834Recalls & reported problems

Siemens Healthcare Diagnostics Products Ltd · Class 1

All models of IMMULITE 2000 Systems E2 Estradiol (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00630414947914), received between Feb 26, 2020 and Feb 26, 2020 — classified as 6 malfunctions.

The most frequently reported specific device problem is “High Test Results” (6 reports).

1 FDA recall record name this device's identifier.

2 other recalls cover devices cleared under the same FDA submission (K932926); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Feb 5, 2020Class IITerminated

    Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.

    Z-1514-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • High Test Results6

Patient problems

  • No Known Impact Or Consequence To Patient6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    TWO DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULTS WERE OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULTS WERE CONSIDERED CORRECT AND WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753652 ↗
  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    ONE DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753643 ↗
  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    ONE DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753642 ↗
  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    ONE DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753638 ↗
  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    ONE DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753637 ↗
  • MalfunctionFeb 26, 2020High Test Results; High Test Results

    ONE DISCORDANT, FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED USING IMMULITE 2000 XPI. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING AN ALTERNATE NON-SIEMENS METHOD. THE REPEAT RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED E2 RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9753624 ↗

FDA clearance and classification

Cleared through 510(k) K932926 on Jan 10, 1994 (decision: Substantially Equivalent). Submitted by Diagnostic Products Corp. as “Immulite Estradiol”. The FDA review took 209 days. FDA record ↗

The FDA regulates product code CHP (radioimmunoassay, estradiol) under 21 CFR 862.1260 as a Class I device. 21 CFR 862.1260 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K932926. One submission often covers many models, so check the product and lot details in each record.

  • Z-1243-2016 · Jan 13, 2016 · Terminated
    IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use t
  • Z-1242-2016 · Jan 13, 2016 · Terminated
    IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21, LKE21(D), LKE25 SMN: 10381132, 10702832, 10381142 Product Usage: For in vitro diagnostic use with the IMMULITE and IMMU

Common questions about the IMMULITE® 2000 Systems E2 Estradiol 10702834

Has the IMMULITE® 2000 Systems E2 Estradiol 10702834 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1514-2020 (Class II), began on Feb 5, 2020: Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post-… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the IMMULITE® 2000 Systems E2 Estradiol 10702834?

The FDA's MAUDE database lists 6 adverse event reports for this device, received on Feb 26, 2020. In those reports, the most common specific device problem is “High Test Results” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the IMMULITE® 2000 Systems E2 Estradiol 10702834?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the IMMULITE® 2000 Systems E2 Estradiol 10702834 FDA approved?

It was cleared through 510(k) K932926 on Jan 10, 1994. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the IMMULITE® 2000 Systems E2 Estradiol 10702834, and is it still sold?

The FDA's device database lists Siemens Healthcare Diagnostics Products Ltd as the company responsible for it (brand IMMULITE® 2000 Systems E2 Estradiol, model 10702834). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other RADIOIMMUNOASSAY, ESTRADIOL devices

All RADIOIMMUNOASSAY, ESTRADIOL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page