Skip to content

Atellica CH uALB_2 11097610Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Microalbumin (µALB_2) Atellica CH - RGT 840T · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00630414596310), received between Nov 30, 2023 and Nov 24, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2024Class IIOpen, Classified

    Atellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

    Z-3169-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Low Test Results3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THAT AN ERRONEOUSLY DEPRESSED MICROALBUMIN_2 (ALB_2) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CI 1900 ANALYZER. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS RETESTED MULTIPLE TIMES WITH AND WITHOUT DILUTIONS ON THE ORIGINAL ANALYZER AS PART OF TROUBLESHOOTING. THEN, THE SAMPLE WAS REPEATED WITH 1:100 MANUAL DILUTION ON THE ORIGINAL ANALYZER AND A HIGHER RESULT WAS OBTAINED. THE HIGHER REPEAT RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED ALB_2 RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23630238 ↗
  • MalfunctionDec 27, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED ERRONEOUSLY DEPRESSED MICROALBUMIN_2 (ALB_2) RESULTS ON TWO (2) PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN AND WERE NOT QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE HIGHER REPROCESSED RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED MICROALBUMIN_2 (ALB_2) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21025015 ↗
  • MalfunctionNov 30, 2023Low Test Results; Low Test Results

    A DISCORDANT FALSELY DEPRESSED MICROALBUMIN 2 (ALB 2) PATIENT SAMPLE RESULT WAS OBTAINED ON AN ATELLICA ®CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAME SAMPLE WAS REPROCESSED ON AN ALTERNATE ATELLICA ® CH 930 ANALYZER. THE HIGHER REPROCESSED RESULT OBTAINED WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED MICROALBUMIN 2 (ALB 2) PATIENT RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18236122 ↗

Other INDICATOR METHOD, PROTEIN OR ALBUMIN (URINARY, NON-QUANT.) devices

All INDICATOR METHOD, PROTEIN OR ALBUMIN (URINARY, NON-QUANT.) devices →

Sources and updates