CLINITEK Novus® Automated Urine Chemistry Analyzer 10697937Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · The CLINITEK Novus® Automated Urine Chemistry analyzer is a fully automated urinalysis instrument for clinical laboratory use. The CLINITEK Novus Analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus cassettes, as well as determine urine specific gravity and urine clarity. The CLINITEK Novus cassettes are intended for the semi-quantitative measurement of the following parameters in urine: albumin, bilirubin, blood (occult), creatinine, glucose, ketone (acetoacetic acid), leu · Class 1

All models of CLINITEK Novus Automated Urine Chemistry Analyzer (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00630414981147), received between Mar 28, 2018 and Nov 20, 2024 — classified as 20 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 8 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result17
  • Low Test Results1
  • Non Reproducible Results1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT TWO PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20738173 ↗
  • MalfunctionNov 20, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT TWO PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20738017 ↗
  • MalfunctionNov 20, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT ONE PATIENT PRODUCED A FALSE NEGATIVE RED BLOOD CELL (RBC) RESULT ON THE NOVUS. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20737692 ↗
  • MalfunctionNov 15, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT FOUR PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS. COMPARISON TESTING WAS PERFORMED ON BOTH THEIR LABORATORY ANALYZER AND UAS800 (SEDIMENT TEST) WHICH YIELDED POSITIVE RESULTS. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20698427 ↗
  • MalfunctionOct 31, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT TWO PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20584871 ↗
  • MalfunctionOct 31, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT FOUR PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20584867 ↗
  • MalfunctionOct 31, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT FOUR PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20584850 ↗
  • MalfunctionOct 31, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT ONE PATIENT PRODUCED A FALSE NEGATIVE RED BLOOD CELL (RBC) RESULT ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON THE UAS800 (SEDIMENT TEST) WHICH YIELDED A POSITIVE RESULT. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20584834 ↗
  • MalfunctionOct 23, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT FOUR PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON BOTH THEIR LABORATORY ANALYZER AND UAS800 (SEDIMENT TEST) WHICH YIELDED POSITIVE RESULTS. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20519064 ↗
  • MalfunctionOct 23, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT ONE PATIENT PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON BOTH THEIR LABORATORY ANALYZER AND UAS800 (SEDIMENT TEST) WHICH YIELDED POSITIVE RESULTS. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20518918 ↗
  • MalfunctionOct 23, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED THAT THREE PATIENTS PRODUCED FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING THE NOVUS CASSETTES 450 LOT. COMPARISON TESTING WAS PERFORMED ON BOTH THEIR LABORATORY ANALYZER AND UAS800 (SEDIMENT TEST) WHICH YIELDED POSITIVE RESULTS. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514435 ↗
  • MalfunctionOct 10, 2024False Negative Result; False Negative Result

    THE CUSTOMER ALLEGED TWO FALSE NEGATIVE RED BLOOD CELL (RBC) RESULTS ON THE NOVUS WHILE USING TWO DIFFERENT LOTS OF THE NOVUS CASSETTES. COMPARISON TESTING WAS PERFORMED ON BOTH THEIR LABORATORY ANALYZER AND UAS800 (SEDIMENT TEST) WHICH YIELDED POSITIVE RESULTS. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20416155 ↗
  • MalfunctionJul 23, 2024Low Test Results; Low Test Results

    THE CUSTOMER ALLEGED AN ISSUE WITH THE CLINITEK NOVUS VERSES THE ATELLICA UAS 800 AFTER INSTALLATION WHERE THE NOVUS WAS GIVING LESS LEUKOCYTES (LEU) THAN WHITE BLOOD CELLS (WBC). THE CASSETTE WAS CALIBRATED AGAIN BUT KEEPS GIVING LESS LEU THAN IT SHOULD. THE MICROSCOPIC PICTURES SHOW THAT THE ATELLICA UAS 800 IS MEASURING THE PARTICLE WELL, BUT IT DOES NOT IN ANY WAY MEET THE MEASUREMENTS IN THE NOVUS. THE ESCALATION WAS OPENED AGAINST THE CLINITEK NOVUS, BUT THE CUSTOMER DOES NOT KNOW WHICH ANALYZER'S RESULTS ARE EXPECTED TO BE TRUE. THERE WAS NO REPORT OF HARM DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19809967 ↗
  • MalfunctionDec 2, 2022False Negative Result; False Negative Result

    THE CUSTOMER REPORTED THAT THEIR CLINITEK NOVUS INSTRUMENT GAVE A FALSE NEGATIVE LEUKOCYTE RESULT COMPARED TO RETESTING THE SAME PAITENT SAMPLE USING A VISUAL READ MULTISTIX 10SG URINE STRIP . THE CUSTOMER ALSO REPORTED THE APPERANCE OF THIS PAITENT SAMPLE WAS NOTABLY CLOUDY. THERE WAS NO REPORT OF INJURY AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15901292 ↗
  • MalfunctionDec 1, 2022False Negative Result; False Negative Result

    THE CUSTOMER REPORTED THAT THEIR CLINITEK NOVUS INSTRUMENT GAVE A FALSE NEGATIVE LEUKOCYTE RESULT COMPARED TO RETESTING THE SAME PATIENT SAMPLE USING A VISUAL READ MULTISTIX 10SG URINE STRIP. THERE WAS NO REPORT OF INJURY AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15899522 ↗
  • MalfunctionSep 2, 2022False Negative Result; False Negative Result

    THE CUSTOMER REPORTED THAT THEIR CLINITEK NOVUS INSTRUMENT GAVE A FALSE NEGATIVE LEUKOCYTE RESULT COMPRED TO RETESTING OF THE SAME SAMPLE ON A SEDIMENT ANALYZER AND ANOTHER CHEMISTRY ANALYZER. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15352403 ↗
  • MalfunctionOct 2, 2020Smoking; Smoking

    THE CUSTOMER PLUGGED THE PRINTER INTO A NEW SAFETY OUTLET AND THE PRINTER STARTED SMOKING, NO SPARKS OR FLAMES. CUSTOMER UNPLUGGED THE PRINTER. THE CUSTOMER INSPECTED THE PRINTER AND SAID A TONER CARTRIDGE BURST INSIDE THE PRINTER. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10619132 ↗
  • MalfunctionMay 14, 2019Non Reproducible Results; Non Reproducible Results

    THE CUSTOMER REPORTED FALSELY DEPRESSED LEUKOCYTES ON THE CLINITEK NOVUS WHEN COMPARED TO THE MICROSCOPIC EXAMINATION OF THE URINE SEDIMENT. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8609857 ↗
  • MalfunctionJan 9, 2019False Negative Result; False Negative Result

    THE CUSTOMER REPORTED FALSE NEGATIVE LEUKOCYTE RESULTS ON THE NOVUS WHEN COMPARED TO THE MICROSCOPIC SEDIMENT READING. THERE IS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8232105 ↗
  • MalfunctionMar 28, 2018False Negative Result; False Negative Result

    THE CUSTOMER REPORTED DISCREPANT BLOOD RESULTS IN URINE BETWEEN THE CLINITEK NOVUS AND A MICROSCOPIC EXAMINATION OF THE SEDIMENT. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7376928 ↗

Other INDICATOR METHOD, PROTEIN OR ALBUMIN (URINARY, NON-QUANT.) devices

All INDICATOR METHOD, PROTEIN OR ALBUMIN (URINARY, NON-QUANT.) devices →