epoc Host 2 as follows recallZ-1214-2023 · Siemens Healthcare Diagnostics Inc · initiated Dec 12, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens has issued an Urgent Medical Device Correction POC 23-004.A.US on December 12, 2022 and Urgent Field Safety Notice POC 23-004.A.OUS to customers informing them of the issue and providing workaround for customers to use until the affected sensor configuration/software version expires on June 12, 2023. The next sensor configuration with the fix and its corresponding sw version will be available to customers in April 2023. Letter states reason for recall, health risk and action to take: " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " If you are a distributor, please ensure your customers receive this UMDC letter. " If you perform QC tests at barometric pressures equal to or above 730 mmHg (97 kPa), no additional action is required. " If you perform QC tests at barometric pressures less than 730 mmHg (97 kPa), apply barometric correction factors before comparing the gas readings with the published value assignments. ***Important Note*** The above barometric correction factor should be applied only to epoc System Software sensor configuration 41.1, epoc Host 2 SW Version 3.37.3 and epoc NXS Host SW Version 4.10.6, which expires June 12, 2023. Following the expiration of this sensor configuration and the software version refer to the epoc System Manual Section 9.5.1 for the appropriate Barometric Correction factor to apply." Please review this letter with your Medical Director." Please retain this letter with your laboratory records and forward this letter to those who may have received the affected product. Questions, please contact your Siemens Healthineers Technical Solution Center or your local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-1214-2023
Status
Open, Classified
Initiated by the company
Dec 12, 2022
Posted by the FDA
Mar 3, 2023
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
22,156 units
Product code
CHL
FDA submissions
K061597
Distribution
Nationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kuwait Laos Latvia Lithuania Luxembourg Malaysia Mali Martinique Mexico Myanmar Netherlands New Caledonia New Zealand Norway Papua Nw Guinea Peru Philippines Poland Portugal Qatar Republic Korea Reunion Romania Russian Fed. San Marino Singapore Slovakia Slovenia Solomon Islands South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey U.A.E. Ukraine United Kingdom Uruguay Vietnam

Product description

epoc Host 2 as follows: Product SMN epoc Host 10736387 epoc Host 2, Canada 10736388 epoc Host 2, China 10736389 epoc Host 2, Japan 10736390 epoc Host 2, Refurbished 10736393 epoc Host 2, Canada, Refurbished 10736394 epoc Host 2, Japanese, Refurbished 10736395 epoc Host 2, Chinese, Refurbished 10736433 epoc Host 2, US (MC55X) 11413524 epoc Host 2, ROW (MC55X) 11413528 epoc Host 2, India (MC55X) 11413541 epoc Host 2, China (MC55X) 11413542 epoc Host 2, Japan (MC55X) 11413543 epoc Host 2 (MC55X) - US - Refurbished 11413601 "epoc Host 2 (MC55X) - ROW -Refurbished" 11413602 epoc Host 2 (MC55X) - India - Refurbished 11413603 epoc Host 2 (MC55X) - Japan - Refurbished 11413604

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Host Software Version /Sensor Configuration: 3.37.3/41.1 Product Description Siemens Material Number (SMN) Unique Device Identification (UDI-DI) epoc Host 2 10736387 00809708052898; epoc Host 2, Canada 10736388 00809708069117; epoc Host 2, China 10736389 00809708097219; epoc Host 2, Japan 10736390 00809708075576; epoc Host 2, Refurbished 10736393 00809708053901; epoc Host 2, Canada, Refurbished 10736394 00809708069124; epoc Host 2, Japanese, Refurbished 10736395 00809708075583; epoc Host 2, Chinese, Refurbished 10736433 00809708097226; epoc Host 2, US (MC55X) 11413524 00630414606460; epoc Host 2, ROW (MC55X) 11413528 00630414606378; epoc Host 2, India (MC55X) 11413541 00630414608617; epoc Host 2, China (MC55X) 11413542 00630414608624; epoc Host 2, Japan (MC55X) 11413543 00630414608532; epoc Host 2 (MC55X) - US - Refurbished 11413601 00630414615004; "epoc Host 2 (MC55X) - ROW - Refurbished" 11413602 00630414614885; epoc Host 2 (MC55X) - India - Refurbished 1413603 00630414614762; epoc Host 2 (MC55X) - Japan - Refurbished 11413604 00630414614892.