IMMULITE®/IMMULITE® 1000 Systems BRM BR-MA (CA15-3) 10380948Recalls & reported problems

Siemens Healthcare Diagnostics Products Ltd · CA 15-3 - IMMULITE - RGT - 100 Tests · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00630414963655), received between Nov 17, 2017 and Nov 17, 2017 — classified as 1 malfunction.

2 other recalls cover devices cleared under the same FDA submission (K013984); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SIEMENS TECHNICAL OPERATIONS INTERNAL INVESTIGATION HAS IDENTIFIED THAT DISCORDANT FALSELY LOW RESULTS WERE OBTAINED ON TWO INVESTIGATIONAL SAMPLES ON AN IMMULITE/ IMMULITE 1000 INSTRUMENT WHEN USING THE BR-MA (CA15-3) ASSAY. THE TWO INVESTIGATIONAL SAMPLES WERE SPIKED WITH BIOTIN AT MULTIPLE CONCENTRATIONS AND RUN FOR INVESTIGATIONAL PURPOSES ON AN IMMULITE/IMMULITE 1000 INSTRUMENT. THE RESULTS OF THE TWO SAMPLES SPIKED WITH 500 NG/ML OF BIOTIN WERE FALSELY LOW WITH A BIAS OF -71% AND -51% RESPECTIVELY COMPARED TO RESULTS OF THE TWO SAMPLES WITHOUT BIOTIN. THERE WERE NO PATIENT SAMPLES IMPACTED AS THIS WAS FOR INVESTIGATIONAL PURPOSES ONLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7041476 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K013984. One submission often covers many models, so check the product and lot details in each record.

  • Z-1691-2018 · Jan 2, 2018 · Terminated
    IMMULITE 2000/IMMULITE 2000 XPi BR-MA (CA15-3)
  • Z-1677-2018 · Jan 2, 2018 · Terminated
    IMMULITE /IMMULITE 1000 BR-MA (CA15-3)

Other SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices

All SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page