HANCOCK® II T505C225 (UDI 00643169000353)Recalls & reported problems

Medtronic, Inc. · VLV T505C225 HKII AOR CII US TEMP IND · Class 3 · Implant

All models of HANCOCK II (43)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00643169000353), received between Aug 4, 2017 and Feb 18, 2026 — classified as 16 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (9 reports).

4 other recalls cover devices cleared under the same FDA submission (P980043); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information9
  • Intravalvular regurgitation4
  • Device Stenosis3
  • Gradient Increase2
  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Calcified1
  • Perivalvular Leak1

Patient problems

  • Aortic Valve Stenosis6
  • Aortic Valve Insufficiency/ Regurgitation5
  • No Information4
  • Insufficient Information2
  • Non specific EKG/ECG Changes2
  • Aortic Insufficiency1
  • Aortic Regurgitation1
  • Bradycardia1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 11 MONTHS POST IMPLANT OF THIS HANCOCK II AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS (AS). EVIDENCE OF CALCIFICATION IN THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT) UNDER THE NON-CORONARY CUSP (NCC). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 24378038 ↗
  • InjuryAug 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHTIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS BRADYCARDIA. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Bradycardia

    MAUDE report 22841013 ↗
  • InjuryMay 14, 2025Intravalvular regurgitation; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 22019134 ↗
  • InjuryMay 12, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT POST IMPLANT OF THIS HANCOCK II AORTIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS SICK SINUS SYNDROME (SSS) . NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 22000437 ↗
  • InjuryMay 5, 2025Calcified; Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Calcified; Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT PRIOR TO THE TRANSCATHETER VALVE-IN-VALVE IMPLANT, AN ECHOCARDIOGRAM PERFORMED NOTED SEVERE AORTIC REGURGITATION, MEAN GRADIENT OF 31.40MMHG, SINUS RHYTHM WITH LEFT BUNDLE BRANCH BLOCK AND THE AORTIC VALVE WAS CALCIFIED. IT WAS STATED THAT THE PATIENT HAD SIGNIFICANT SHORTNESS OF BREATH WITH MINIMAL EXERTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Dyspnea; Insufficient Information; Non specific EKG/ECG Changes

    MAUDE report 21960119 ↗
  • MalfunctionJan 28, 2025Gradient Increase; Gradient Increase

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT REPORTED THAT THE "VALVE IS COLLECTING PLAQUE AND NEEDS TO BE REPLACED PREMATURELY". NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED THAT 8 DAYS LATER, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS GRADIENT INCREASE. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21250396 ↗
  • InjuryAug 26, 2024Device Stenosis; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 20071191 ↗
  • InjuryJul 2, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 5 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 19653763 ↗
  • InjuryMay 14, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 5 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC INSUFFICIENCY (AI) OF UNKNOWN SEVERITY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, APPROXIMATELY 4 WEEKS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 19308346 ↗
  • InjuryMay 8, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 16884584 ↗
  • InjuryJun 24, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 12059719 ↗
  • InjuryMar 13, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 11 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC AORTIC VALVE, IT WAS REPLACED VALVE-IN-VALVE WITH A 26MM TRANSCATHETER VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Insufficiency

    MAUDE report 9830010 ↗
  • InjuryJun 4, 2019Perivalvular Leak; Insufficient Device Problem Information; Perivalvular Leak; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A 27 MM VALVE OF THE SAME MODEL. THE REASON FOR THE EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8667319 ↗
  • InjuryJun 26, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT THREE YEARS AND EIGHT MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A VALVE FROM AN UNKNOWN MANUFACTURER. THE REASON FOR EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 7639254 ↗
  • InjuryMay 29, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT THREE YEARS AND SIX MONTHS POST IMPLANT OF THIS 25MM BIOPROSTHETIC VALVE IMPLANTED IN A SEVEN YEAR OLD PEDIATRIC PATIENT, THE VALVE WAS REMOVED AND REPLACED WITH A 27MM VALVE IN THE PULMONARY POSITION. THE REASON FOR REPLACEMENT IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Calcium Deposits/Calcification; Endocarditis; No Information

    MAUDE report 7551291 ↗
  • InjuryMar 8, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING CONSIDERED FOR VALVE-IN-VALVE REPLACEMENT DUE TO UNKNOWN REASONS. NO KNOWN INTERVENTION HAS OCCURRED. SUBSEQUENTLY, AN UNKNOWN DURATION LATER, THE PATIENT UNDERWENT A VALVE-IN-VALVE REPLACEMENT WITH A NON MEDTRONIC DEVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 7323141 ↗
  • InjurySep 26, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE DEVICE WAS REPLACED VALVE-IN-VALVE WITH A TRANSCATHETER VALVE DUE TO SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation

    MAUDE report 6894988 ↗
  • InjuryAug 4, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE DEVICE WAS REPLACED VALVE-IN-VALVE DUE TO AORTIC STENOSIS. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 6768070 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980043. One submission often covers many models, so check the product and lot details in each record.

  • Z-1884-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1882-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T505U227 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1883-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1226-2009 · Jan 26, 2009 · Terminated
    Hancock II Bioprostheses with Cinch, Heart Valve, Model # T505 & T510; Catalog #s T505C2501, T505C2101, T505U25ID , T505C2301, T505C2301, T505C2501, T505C2101, T510C29ID, T510C25ID

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →