MOSAIC ULTRA® 305U229 (UDI 00643169001480)Recalls & reported problems

Medtronic, Inc. · VLV 305U229 MOSAIC AOR ULT CII US TEMP I · Class 3 · Implant

All models of MOSAIC ULTRA (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00643169001480), received between Jul 16, 2018 and Sep 14, 2026 — classified as 13 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Device Stenosis3
  • Intravalvular regurgitation3
  • Adverse Event Without Identified Device or Use Problem2
  • Patient Device Interaction Problem2
  • Biocompatibility1
  • Perivalvular Leak1
  • Structural Problem1

Patient problems

  • Aortic Valve Stenosis4
  • Aortic Valve Insufficiency/ Regurgitation3
  • No Information3
  • Aortic Regurgitation2
  • Corneal Pannus2
  • Insufficient Information2
  • Non specific EKG/ECG Changes2
  • Atrial Fibrillation1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Intravalvular regurgitation

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 50762976 ↗
  • InjuryMar 24, 2026Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 11 MONTHS POST IMPLANT OF THIS 29MM MOSAIC AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS (AS) AND SHORTNESS OF BREATH (SOB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Dyspnea

    MAUDE report 24673002 ↗
  • InjuryNov 18, 2025Patient Device Interaction Problem; Intravalvular regurgitation; Patient Device Interaction Problem; Intravalvular regurgitation

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT ON (B)(6) 2025, THE PATIENT SUCCESSFULLY HAD A 29MM TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 23581150 ↗
  • InjuryAug 1, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 7 MONTHS POST IMPLANT OF THIS MOSAIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 22685310 ↗
  • InjuryMar 18, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 21632654 ↗
  • MalfunctionAug 21, 2024Device Stenosis; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 2 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 20020809 ↗
  • InjuryMar 11, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 18873048 ↗
  • InjurySep 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT IMMEDIATELY POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT EXPERIENCED VARIABLE HEART BLOCK WHICH REQUIRED AV SEQUENTIAL PACING AND ONE EPISODE OF ATRIAL FIBRILLATION WHICH WAS TREATED WITH CARDIOVERSION. IT WAS STATED THAT THE PATIENTS RHYTHM BECAME STABLE. THE PRESENCE OF FIRST DEGREE ATRIOVENTRICULAR BLOCK WAS REPORTED AT DISCHARGE. A NON-MEDTRONIC MITRAL ANNULOPLASTY DEVICE WAS ALSO IMPLANTED DURING THE SAME PROCEDURE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Atrial Fibrillation; Non specific EKG/ECG Changes

    MAUDE report 17682502 ↗
  • InjuryJul 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 2 DAYS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 17412355 ↗
  • InjuryApr 8, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 6 MONTHS POST IMPLANT OF THIS 29 MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM AORTIC BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 14056494 ↗
  • InjuryFeb 11, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT THREE YEARS AND THREE MONTHS POST IMPLANT OF THIS 29MM BIOPROSTHETIC AORTIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM NON-MEDTRONIC BIOPROSTHETIC VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8327495 ↗
  • InjuryOct 22, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 2 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. THE REASON FOR THE EXPLANT WAS NOT REPORTED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Endocarditis; No Information

    MAUDE report 7993700 ↗
  • InjuryOct 1, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT ONE YEAR AND FOUR MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A VALVE OF THE SAME SIZE AND MODEL. THE REASON FOR EXPLANT IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation; Aortic Valve Stenosis; Corneal Pannus; Cusp Tear; No Information

    MAUDE report 7924623 ↗
  • InjuryJul 16, 2018Perivalvular Leak; Structural Problem; Biocompatibility; Biocompatibility

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS 9 MONTHS THIS 29 MM AORTIC BIOPROSTHETIC VALVE WAS EXPLANTED DUE TO SEVERE AORTIC REGURGITATION AND PARAVALVULAR LEAK (PVL). THE VALVE WAS SUCCESSFULLY REPLACED WITH A NON MEDTRONIC DEVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation; Corneal Pannus; Host-Tissue Reaction

    MAUDE report 7689469 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →