Skip to content

MOSAIC® 310C25 (UDI 00643169001657)Recalls & reported problems

Medtronic, Inc. · VLV 310C25 MOSAIC MIT CINCH US TEMP IND · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Mosaic 310C25 tissue mitral valve is implanted surgically to replace a diseased heart valve. The 35 reports mostly describe problems years after implantation: valves being removed and replaced because leaflets tore or the tissue hardened, blood leaking around or through the valve, and heart rhythm problems that led some patients to need a pacemaker. Two deaths and 33 injuries are reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
2
Injuries reported
33
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00643169001657), received between Jan 23, 2017 and Mar 31, 2026 — classified as 2 deaths, 33 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (26 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information26
  • Adverse Event Without Identified Device or Use Problem5
  • Patient Device Interaction Problem3
  • Calcified2
  • Degraded2
  • Intravalvular regurgitation2
  • Material Split, Cut or Torn2
  • Perivalvular Leak2

Patient problems

  • Insufficient Information11
  • No Information11
  • Mitral Valve Insufficiency/ Regurgitation5
  • Non specific EKG/ECG Changes5
  • Mitral Regurgitation3
  • Mitral Valve Stenosis3
  • Complete Heart Block2
  • Endocarditis2

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT 11 YEARS AND 1 MONTH'S POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM MITRAL BIOPROSTHETIC VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24741905 ↗
  • InjuryMar 26, 2026Patient Device Interaction Problem

    IT WAS REPORTED THAT 1 YEAR AND 11 MONTHS POST IMPLANT OF THIS OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS ATRIAL ARRYTHMIAS (ATRIAL TACHYCARDIA (AT), ATRIAL FIBRILLATION (AF) AND ATRIAL FLUTTER (A-FL). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Atrial Fibrillation; Atrial Flutter; Tachycardia

    MAUDE report 24696830 ↗
  • InjuryNov 12, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT 6 YEARS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, A 26MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN -VALVE IN THE SAME POSITION. IT WAS ALSO REPORTED THAT DURING THIS PROCEDURE THE MOSAIC STENT FRAME WAS FRACTURED. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23534944 ↗
  • InjuryJun 23, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 22307602 ↗
  • InjuryMar 11, 2025Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information

    IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM MOSAIC MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A DEVICE OF THE SAME MAKE AND SIZE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21572028 ↗
  • InjuryNov 12, 2024Insufficient Device Problem Information; Material Split, Cut or Torn; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Material Split, Cut or Torn; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS MITRAL STENOSIS, SEVERE MITRAL REGURGITATION AND A TORN LEAFLET. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation; Mitral Valve Stenosis

    MAUDE report 20663709 ↗
  • InjuryAug 21, 2024Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED A PHONE CALL FROM THE PATIENT 9 YEARS AND 1 MONTHS POST IMPLANT OF THIS MITRAL BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS THE VALVE IS "LEAKING" AND NEEDS TO BE REPLACED. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 20009892 ↗
  • DeathApr 29, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 DAY POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE IMPLANTED IN THE TRICUSPID POSITION, THE PATIENT DIED. THE CAUSE OF DEATH WAS RECEIVED AS CARDIOGENIC SHOCK, STAPH AUREUS BACTEREMIA AND SEPSIS. IT IS UNKNOWN WHETHER AN AUTOPSY WAS PERFORMED.

    Patient problems: Cardiogenic Shock; Sepsis

    MAUDE report 19202784 ↗
  • InjuryFeb 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18686158 ↗
  • InjuryNov 10, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 6 DAYS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 18109726 ↗
  • InjuryOct 31, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18038251 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 4 DAYS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 17859366 ↗
  • InjuryMay 22, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF THE SAME EXACT MAKE AND MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16975678 ↗
  • InjuryJan 17, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 29MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS REPORTED AS STENOSIS WITH MILD MITRAL REGURGITATION (MR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation; Mitral Valve Stenosis; Valvular Insufficiency/ Regurgitation

    MAUDE report 16185316 ↗
  • InjuryNov 21, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE IMPLANTED IN THE PULM ONARY POSITION, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 15838335 ↗
  • InjuryJul 15, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE IMPLANTED IN THE TRICUSPID POSITION, IT WAS EXPLANTED AND REPLACED WITH A VALVE OF THE SAME SIZE AND MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Tricuspid Valve Insufficiency/ Regurgitation

    MAUDE report 15028572 ↗
  • InjuryApr 26, 2022Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 10 MONTHS POST IMPLANT OF THIS  25 MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A VALVE OF THE SAME SIZE AND MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 14214342 ↗
  • InjuryMar 2, 2021Calcified; Perivalvular Leak; Insufficient Device Problem Information; Calcified; Perivalvular Leak; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 11 MONTHS POST IMPLANT OF THIS 25 MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 29 MM BIOPROSTHETIC VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation; Valvular Insufficiency/ Regurgitation

    MAUDE report 11398651 ↗
  • DeathNov 20, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 DAY POST IMPLANT OF THIS BIOPROSTHETIC VALVE, THE PATIENT EXPIRED. THE CAUSE OF DEATH WAS NOT RECEIVED. THERE IS NO EVIDENCE TO SUGGEST THAT THE VALVE OR ITS FUNCTION CONTRIBUTED TO THE PATIENT¿S DEATH. IT IS UNKNOWN WHETHER AN AUTOPSY WAS PERFORMED.

    Patient problems: Death

    MAUDE report 10879607 ↗
  • InjuryAug 31, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 8 MONTHS POST IMPLANT OF THIS 25 MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 29 MM NON MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Endocarditis; Mitral Regurgitation; Mitral Valve Stenosis; No Information

    MAUDE report 10475325 ↗
  • InjuryMay 22, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 11 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE IMPLANTED IN THE TRICUSPID POSITION OF A PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 10083394 ↗
  • InjuryMay 7, 2020Insufficient Device Problem Information; Material Split, Cut or Torn; Insufficient Device Problem Information; Material Split, Cut or Torn

    MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A BIOPROSTHETIC VALVE OF THE SAME SIZE AND MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Regurgitation; No Information

    MAUDE report 10032971 ↗
  • InjuryMay 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 11 DAYS POST IMPLANT OF THIS 25MM BIOPROSTHETIC MITRAL VALVE AND THIS 30MM TRICUSPID ANNULOPLASTY BAND, A PERMANENT PACEMAKER WAS IMPLANTED IN THE PATIENT. THE REASON FOR THE PACEMAKER IMPLANT WAS REPORTED AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Complete Heart Block

    MAUDE report 10020523 ↗
  • InjuryAug 30, 2019Perivalvular Leak; Insufficient Device Problem Information; Perivalvular Leak; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 10 MONTHS POST IMPLANT OF THIS BIOPROSTHETIC MITRAL VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A BIOPROSTHETIC VALVE OF A DIFFERENT MODEL.  THE REASON FOR THE EXPLANT WAS NOT REPORTED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8949817 ↗
  • InjuryMar 5, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT TWO YEARS AND SIX MONTHS POST IMPLANT OF THIS MITRAL BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 8391992 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page