MOSAIC® 305C223 (UDI 00643169001725)Recalls & reported problems

Medtronic, Inc. · VLV 305C223 MOSAIC AOR CII NNA US TEMP I · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic MOSAIC aortic bioprosthetic heart valve appears in 39 adverse event reports, with no deaths and 36 injuries. Reports mostly describe valve stenosis, regurgitation, or calcification years after implant, often leading to valve-in-valve replacement evaluations. There are no recalls naming this device. Reports dropped from 10 to 2 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
36
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 00643169001725), received between Jun 21, 2017 and Jan 12, 2026 — classified as 36 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (26 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information26
  • Device Stenosis7
  • Gradient Increase6
  • Intravalvular regurgitation6
  • Material Split, Cut or Torn2
  • Adverse Event Without Identified Device or Use Problem1
  • Calcified1
  • Degraded1

Patient problems

  • Aortic Valve Stenosis17
  • Aortic Valve Insufficiency/ Regurgitation13
  • Insufficient Information11
  • No Information5
  • Aortic Regurgitation4
  • Aortic Insufficiency1
  • Calcium Deposits/Calcification1
  • Cusp Tear1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 12, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT 9 YEARS AND 7 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. EVIDENCE OF CALCIUM WAS SEEN IN THE RIGHT CORONARY ARTERY (RCA), LEFT VENTRICULAR OUTFLOW TRACT (LVOT) UNDER THE NON-CORONARY CUSP (NCC) AND PROTRUDING CALCIUM SEEN IN THE ABDOMINAL AORTA AND LEFT FEMORAL ARTERY (LFA). IT WAS STATED THAT ANTERIOR CALCIUM WAS ALSO SEEN IN THE RIGHT FEMORAL ARTERY (RFA) AND ANTEROMEDIAL CALCIUM IN THE LEFT FEMORAL ARTERY (LFA). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Calcium Deposits/Calcification

    MAUDE report 24035009 ↗
  • InjuryDec 11, 2025Intravalvular regurgitation; Intravalvular regurgitation

    IT WAS REPORTED THAT 8 YEARS AND 12 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS MODERATE AORTIC CENTRAL REGURGITATION. IT WASREPORTED THAT AN INTERVENTION IS REQUIRED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 23776618 ↗
  • InjurySep 16, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis

    ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR INTERVENTION AS AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 23060974 ↗
  • InjurySep 4, 2025Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. SUBSEQUENTLY 22 DAYS LATER, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 22963035 ↗
  • InjuryJul 18, 2025Gradient Increase; Insufficient Device Problem Information; Gradient Increase; Insufficient Device Problem Information

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE GRADIENT WAS 30MMHG. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22540053 ↗
  • InjuryApr 28, 2025Calcified; Thickening of Material; Device Stenosis; Calcified; Thickening of Material; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 2 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS CALCIFICATION, AORTIC STENOSIS AND HYPOATTENUATED LEAFLET THICKENING (HALT). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 21918762 ↗
  • MalfunctionApr 24, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21900487 ↗
  • MalfunctionJan 24, 2025Gradient Increase; Device Stenosis; Gradient Increase; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 9 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS AND GRADIENT INCREASE. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 21233794 ↗
  • InjuryJan 23, 2025Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 9 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 21222599 ↗
  • InjuryDec 13, 2024Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    IT WAS REPORTED THAT AN UNKNOWN FAILURE ASSOCIATED WITH THE USE OF A MOSAIC AORTIC VALVE HAD OCCURRED. THE DEVICE REMAINS IN SERVICE. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 20923732 ↗
  • InjuryNov 28, 2024Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE AORTIC INSUFFICIENCY AS SEVERE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REP ORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 20805740 ↗
  • InjuryNov 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, ONE WEEK LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20757874 ↗
  • InjuryApr 1, 2024Device Stenosis; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE THAT THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 MONTH AND 11 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 19014510 ↗
  • InjuryFeb 8, 2024Degraded; Intravalvular regurgitation; Degraded; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 2 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS REPORTED AS AORTIC INSUFFICIENCY (AI). IT WAS STATED THAT PRIOR TO THE EXPLANT PROCEDURE, THE PATIENT COMPLAINED OF SHORTNESS OF BREATH. IT WAS ALSO REPORTED THAT UPON EXPLANT, IT WAS NOTED THAT ONE OF THE VALVE LEAFLET HAD DETERIORATED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Dyspnea

    MAUDE report 18670924 ↗
  • InjuryAug 11, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION WHICH STATED THAT THE SEVERITY OF AORTIC INSUFFICIENCY WAS MODERATE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Cusp Tear

    MAUDE report 17508061 ↗
  • InjuryJun 30, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 5 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 17236696 ↗
  • InjuryMar 25, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 6 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 16616741 ↗
  • InjuryDec 12, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 10 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A NON-MEDTRONIC TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 15959056 ↗
  • InjuryNov 21, 2022Gradient Increase; Gradient Increase

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS AND INCREASED GRADIENT. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 15 DAYS LATER, A TRANSCATHETER VALVE WAS ATTEMPTED TO BE IMPLANTED VALVE-IN-VALVE. IT WAS STATED THAT THE CASE WAS ABORTED DUE TO PATIENT ANATOMY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 15834901 ↗
  • InjuryJul 14, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 11 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 15014969 ↗
  • InjuryMay 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS SEVERE AORTIC STENOSIS AND SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 14323369 ↗
  • InjuryFeb 3, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 9 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE IN THE PULMONARY POSITION, IT WAS EXPLANTED AND REPLACED WITH A 20MM PULMONARY VALVED CONDUIT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 13446859 ↗
  • InjuryOct 25, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 2 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 12689548 ↗
  • InjuryOct 5, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 12576321 ↗
  • InjuryAug 5, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 10 MONTHS POST IMPLANT OF THIS 23MM BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 12281446 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →