MOSAIC® 305C221 (UDI 00643169001763)Recalls & reported problems
Medtronic, Inc. · VLV 305C221 MOSAIC AOR CII US TEMP IND · Class 3 · Implant
Quick summary
This Medtronic Mosaic305C221 is a tissue aortic heart valve used to replace a diseased aortic valve. Its 50 reports mostly describe the valve narrowing or stiffening over time, leaking, and calcification, often years after implant, sometimes leading to a replacement valve or a pacemaker;46 reports involved injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 45
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 00643169001763), received between Dec 5, 2016 and Apr 29, 2026 — classified as 45 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (32 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information32
- Gradient Increase10
- Device Stenosis9
- Inadequacy of Device Shape and/or Size3
- Patient Device Interaction Problem3
- Adverse Event Without Identified Device or Use Problem2
- Intravalvular regurgitation2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Aortic Valve Stenosis25
- No Information14
- Insufficient Information12
- Aortic Valve Insufficiency/ Regurgitation7
- Aortic Regurgitation3
- Aortic Insufficiency2
- No Known Impact Or Consequence To Patient2
- Cardiomyopathy1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Appropriate Device Problem Term/Code Not Available
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 9 MONTHS POST IMPLANT OF THIS 21 MM MOSAIC AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS HEART FAILURE/HEMODYNAMIC INSTABILITY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiovascular Dysfunction/Insufficiency
MAUDE report 25031224 ↗InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem; Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT 10 YEARS AND 9 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM AORTIC BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 24696709 ↗InjuryFeb 11, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 4 DAYS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY, LOW EJECTION FRACTION (EF) AND COMPLETE HEART BLOCK (CHB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiomyopathy; Heart Block
MAUDE report 24313787 ↗InjuryJan 21, 2026Insufficient Device Problem Information; Patient Device Interaction Problem; Device Stenosis
ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR EVALUATION WAS REPORTED AS VALVE STENOSIS, HEMODYNAMIC INSTABILITY AND CONGESTIVE HEART FAILURE (CHF). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Heart Failure/Congestive Heart Failure; Insufficient Information; Stenosis
MAUDE report 24129864 ↗InjuryJan 6, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 4 YEARS AND 8 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS ATRIOVENTRICULAR BLOCK (2ND DEGREE, MOBITZ II). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 23972259 ↗MalfunctionJan 2, 2026Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION FROM THE PATIENT THAT 3 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THEIR CAR DIOLOGIST STATED THEIR DEVICE IS NO LONGER WORKING. THE MODE OF FAILURE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23945967 ↗InjuryDec 29, 2025Calcified; Calcified
IT WAS REPORTED THAT 11 YEARS AND 11 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS VALVE CALCIFICATION. SUBSEQUENTLY ON (B)(6) 2025, THE PATIENT SUCCESSFULLY HAD A 23MM TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23904592 ↗InjuryMay 15, 2025Intravalvular regurgitation; Intravalvular regurgitation
IT WAS REPORTED THAT POST IMPLANT OF THIS MOSAIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY (AI). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22025681 ↗InjuryMar 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT AN 21MM AORTIC VALVE, IMPLANTED IN THE AORTIC POSITION ON (B)(6) 2015, WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF A DIFFERENT MODEL ON (B)(6) 2025. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 21527617 ↗InjuryFeb 25, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT THE REASON FOR INTERVENTION WAS DUE TO AORTIC STENOSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 21457392 ↗MalfunctionJan 9, 2025Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 4 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 21112757 ↗MalfunctionNov 7, 2024Device Stenosis; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE PATIENT IS BEING EVALUATED FOR A POTENTIAL FULL REPLACEMENT OF THE BIOPROSTHETIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 20634620 ↗InjurySep 6, 2024Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVA LUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 5 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 20155583 ↗InjuryJun 12, 2024Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 7 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 3 DAYS LATER, THE PATIENT HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE OF A DIFFERENT MANUFACTURER. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 19521657 ↗InjuryFeb 15, 2024Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 8 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 18710559 ↗InjuryFeb 14, 2024Gradient Increase; Device Stenosis; Gradient Increase; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 1 MONTH AND 22 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE -IN-VALVE. AN ADDITIONAL REASON FOR INTERVENTION WAS REPORTED AS A HIGH GRADIENT OF 111 MMHG DUE TO THE SEVERE AS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 18707625 ↗InjuryMay 17, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 1 MONTH POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND MODERATE AORTIC INSUFFICIENCY (AI). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 16950471 ↗InjuryMay 2, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 5 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 16855078 ↗InjuryMar 10, 2023Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS INCREASED GRADIENTS AND AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 16520269 ↗InjuryFeb 27, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 9 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, APPROXIMATELY 6 WEEKS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE AORTIC INSUFFICIENCY AND MILD AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 16443672 ↗InjuryNov 12, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 12806630 ↗InjuryJul 13, 2021Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 4 MONTHS POST IMPLANT OF THIS 21MM BIOPROSTHETIC AORTIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM BIOPROSTHETIC ROOT REPLACEMENT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS INCREASED GRADIENTS WITH AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 12159726 ↗InjuryJun 17, 2021Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO HIGH GRADIENTS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 12016753 ↗InjuryMar 17, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 3 MONTHS POST IMPLANT OF THIS 21MM AORTIC BIOPROSTHETIC VALVE, A 23MM TRANSCATHETER AORTIC VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 11505596 ↗InjuryDec 15, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 11 MONTHS POST IMPLANT OF THIS 21MM BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 11008110 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports