FREESTYLE® 995MS29 (UDI 00643169002074)Recalls & reported problems

Medtronic, Inc. · VLV 995MS29 FREESTYLE MS US TEMP IND · Class 3 · Implant

All models of FREESTYLE (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00643169002074), received between Sep 11, 2017 and Mar 20, 2026 — classified as 14 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information10
  • Intravalvular regurgitation3
  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Stenosis1
  • Incomplete Coaptation1
  • Material Split, Cut or Torn1

Patient problems

  • Insufficient Information6
  • Aortic Valve Insufficiency/ Regurgitation4
  • Aortic Regurgitation2
  • No Information2
  • Aortic Insufficiency1
  • Aortic Valve Stenosis1
  • Chest Pain1
  • Complete Heart Block1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 20, 2026Patient Device Interaction Problem; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR EVALUATION WAS MODERATE AORTIC INSUFFICIENCY. SUBSEQUENTLY ON (B)(6) 2026 THE PATIENT SUCCESSFULLY HAD A NON-MEDTRONIC TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Chest Pain; Dyspnea; Heart Failure/Congestive Heart Failure

    MAUDE report 24647136 ↗
  • InjuryDec 4, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23717715 ↗
  • InjurySep 11, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT 11 YEARS AND 3 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A NEW 27MM AORTIC BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23023211 ↗
  • InjuryJun 24, 2025Insufficient Device Problem Information; Material Split, Cut or Torn; Insufficient Device Problem Information; Material Split, Cut or Torn

    ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR REPLACEMENT WAS DUE TO A CORONARY CUSP TEAR. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22318123 ↗
  • InjuryMar 7, 2025Intravalvular regurgitation; Intravalvular regurgitation

    IT WAS REPORTED THAT A PATIENT WITH A VALVE 995MS29 FREESTYLE MS 29MM IMPLANT EXPERIENCED SEVERE AORTIC INSUFFICIENCY. A COMPUTED TOMOGRAPHY SCAN WAS CONDUCTED TO EVALUATE THE POSSIBILITY OF A TRANSCATHETER AORTIC VALVE REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 21548061 ↗
  • InjuryJan 13, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 1 MONTH POST IMPLANT OF THIS 29 MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27 MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21140147 ↗
  • MalfunctionDec 5, 2024Patient Device Interaction Problem; Intravalvular regurgitation; Patient Device Interaction Problem; Intravalvular regurgitation

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION WHICH STATED THAT SUBSEQUENTLY, 5 DAYS LATER, A TRANSCATHETER VALVE OF A DIFFERENT MANUFAC TURER WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MILD TO MODERATE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Thrombosis/Thrombus

    MAUDE report 20852474 ↗
  • InjuryJun 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 29MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19447262 ↗
  • InjuryDec 15, 2022Incomplete Coaptation; Incomplete Coaptation

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 8 MONTHS POST IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, A 34MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY DUE TO FLAIL LEAFLET. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 15988233 ↗
  • InjuryMar 1, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 29MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 13635993 ↗
  • InjuryFeb 26, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT SIX YEARS AND ONE MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation

    MAUDE report 8371621 ↗
  • InjuryNov 19, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY ONE MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED DUE TO COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Complete Heart Block; Non specific EKG/ECG Changes

    MAUDE report 8086009 ↗
  • InjuryNov 16, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT THREE YEARS AND SEVEN MONTHS POST IMPLANT OF THIS BIOPROSTHETIC AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO SEVERE INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Insufficiency; Regurgitation

    MAUDE report 8077961 ↗
  • InjuryNov 28, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 10 MONTHS POST IMPLANT OF THIS BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A VALVE OF SAME SIZE AND MODEL. THE REASON FOR THE EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Cusp Tear; No Information

    MAUDE report 7063050 ↗
  • InjurySep 11, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS 9 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE DEVICE WAS REPLACED VALVE-IN-VALVE WITH A MEDTRONIC TRANSCATHETER VALVE FOR UNKNOWN REASONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation; No Information

    MAUDE report 6855976 ↗

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page